Evaluation of the Sysmex UF-1000i® urine flow cytometer in the diagnostic work-up of suspected urinary tract infection in a Dutch general hospital

被引:38
|
作者
van der Zwet, Wil C. [1 ]
Hessels, Jan [2 ]
Canbolat, Fatih [1 ]
Deckers, Martine M. L. [2 ]
机构
[1] Deventer Hosp, Dept Med Microbiol & Infect Control, NL-7400 GC Deventer, Netherlands
[2] Deventer Hosp, Dept Clin Chem, NL-7400 GC Deventer, Netherlands
关键词
flow cytometry; Sysmex UF-1000i; urinary tract infection; AUTOMATED URINALYSIS; SEDIMENT EXAMINATION; TEST STRIPS; MICROSCOPY; PERFORMANCE; ANALYZER;
D O I
10.1515/CCLM.2010.342
中图分类号
R446 [实验室诊断]; R-33 [实验医学、医学实验];
学科分类号
1001 ;
摘要
Background: Automation and standardization of sediment analysis of urine samples by flow cytometry might serve as an alternative to labor-intensive laboratory methods, such as microscopic examination and culture. The Sysmex UF-1000i is a urine flow cytometer that uses two separate channels for counting blood cells and bacteria. Methods: In this study, 358 urine samples were analyzed with the Sysmex UF-1000i in parallel with manual microscopy, Gram stain and bacterial culture, the latter considered the gold standard. Results: Reproducibility for detection of white and red blood cells and bacteria was good, while detection of yeast proved unreliable. Depending on the definition of urinary tract infection (UTI) used, the negative predictive value and the percentage of false-negative results were 100% and 0% [UTI >= 10(5) colony forming units (CFU)/mL] and 99% and 1.3%, (UTI >= 10(4) CFU/mL), respectively. Pre-screening with the Sysmex UF-1000i would have resulted in a reduction of bacterial culture by 42%. Carry over of bacteria between consecutive samples due to the use of fixed sample needle was observed, but did not result in false-positive interpretation of Sysmex UF-1000i results. Because of the occurrence of carry over, samples that have been analyzed with the Sysmex UF-1000i cannot be used for subsequent urine culture. Conclusions: In conclusion, the Sysmex UF-1000i offers the possibility for screening high numbers of urine samples in a fast and standardized way, resulting in a reduction in workload and speeding the diagnostic process. It is not recommended for use in complicated patient populations, such as neutropenic patients and patients in whom yeast infection is suspected. Clin Chem Lab Med 2010; 48:1765-71.
引用
收藏
页码:1765 / 1771
页数:7
相关论文
共 33 条
  • [1] Screening for Urinary Tract Infection with the Sysmex UF-1000i Urine Flow Cytometer
    Broeren, Maarten A. C.
    Bahceci, Semiha
    Vader, Huib L.
    Arents, Niek L. A.
    JOURNAL OF CLINICAL MICROBIOLOGY, 2011, 49 (03) : 1025 - 1029
  • [2] Evaluation of the Sysmex UF-1000i flow cytometer for screening of urinary tract infection
    de Frutos-Serna, Monica
    Luz Asensio-Calle, Maria
    Maria Haro-Perez, Ana
    Maria Blazquez-de Castro, Ana
    Nieves Gutierrez-Zufiaurre, Maria
    Iglesias-Garcia, Jesus
    ENFERMEDADES INFECCIOSAS Y MICROBIOLOGIA CLINICA, 2014, 32 (03): : 147 - 151
  • [3] Cutoff values for bacteria and leukocytes for urine flow cytometer Sysmex UF-1000i in urinary tract infections
    Manoni, Fabio
    Fornasiero, Lucia
    Ercolin, Mauro
    Tinello, Agostino
    Ferrian, Melissa
    Hoffer, Paolo
    Valverde, Sara
    Gessoni, Gianluca
    DIAGNOSTIC MICROBIOLOGY AND INFECTIOUS DISEASE, 2009, 65 (02) : 103 - 107
  • [4] Evaluation of the Sysmex UF-1000i for the Diagnosis of Urinary Tract Infection
    Wang, Jie
    Zhang, Ying
    Xu, DongWen
    Shao, WeiJun
    Lu, Yuan
    AMERICAN JOURNAL OF CLINICAL PATHOLOGY, 2010, 133 (04) : 577 - 582
  • [5] Sysmex UF-1000i flow cytometer to screen urinary tract infections: the URISCAM multicentre study
    Herraez, O.
    Asencio, M. A.
    Carranza, R.
    Jarabo, M. M.
    Huertas, M.
    Redondo, O.
    Arias-Arias, A.
    Jimenez-Alvarez, S.
    Solis, S.
    Zamarron, P.
    Illescas, M. S.
    Galan, M. A.
    LETTERS IN APPLIED MICROBIOLOGY, 2018, 66 (03) : 175 - 181
  • [6] Evaluation of the Sysmex UF-1000i® automated system for the diagnosis of urinary tract infection
    Munoz-Algarra, Maria
    Martinez-Ruiz, Rocio
    Orden-Martinez, Beatriz
    ENFERMEDADES INFECCIOSAS Y MICROBIOLOGIA CLINICA, 2013, 31 (01): : 29 - 31
  • [7] Evaluation of the Sysmex UF1000i flow cytometer for ruling out bacterial urinary tract infection
    De Rosa, Rita
    Grosso, Shamanta
    Bruschetta, Graziano
    Avolio, Manuela
    Stano, Paola
    Modolo, Maria Luisa
    Camporese, Alessandro
    CLINICA CHIMICA ACTA, 2010, 411 (15-16) : 1137 - 1142
  • [8] Evaluation of the Sysmex UF-1000i Urine Analyzer as a Screening Test to Reduce the Need for Urine Cultures for Urinary Tract Infection
    Hu, Xiaozhou
    Zhang, Jie
    Zhang, Xiaowei
    LABMEDICINE, 2010, 41 (06): : 349 - 352
  • [9] Validation and Search of the Ideal Cut-Off of the Sysmex UF-1000i® Flow Cytometer for the Diagnosis of Urinary Tract Infection in a Tertiary Hospital in Spain
    Millan-Lou, Maria I.
    Garcia-Lechuz, Juan M.
    Ruiz-Andres, Maria A.
    Lopez, Concepcion
    Aldea, Maria J.
    Revillo, Maria J.
    Rezusta, Antonio
    FRONTIERS IN MEDICINE, 2018, 5
  • [10] The Diagnostic Performance of Urine Flow Cytometer UF1000i for Urinary Tract Infections
    Sun, S. J.
    Zuo, L. L.
    Liu, P. P.
    Wang, X. M.
    He, M. L.
    Wu, S. Y.
    CLINICAL LABORATORY, 2018, 64 (09) : 1395 - 1401