Subcutaneous air entrapment after subcutaneous implantable cardioverter defibrillator implantation evaluated by computed tomography

被引:0
|
作者
Taguchi, Yuka [1 ]
Ishikawa, Toshiyuki [1 ]
Matsumoto, Katsumi [1 ]
Narikawa, Masatoshi [1 ]
Okazaki, Yoshinori [1 ]
Miyagawa, Shuichi [1 ]
Horigome, Akira [1 ]
Hosoda, Junya [1 ]
机构
[1] Yokohama City Univ, Grad Sch Med, Dept Cardiol, 3-9 Fukuura,Kanazawa Ku, Yokohama, Kanagawa, Japan
来源
关键词
complications; ICD; inappropriate shock; oversensing; sensing malfunction; SHOCKS;
D O I
10.1111/pace.14962
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
BackgroundInappropriate shock (IAS) caused by subcutaneous air entrapment (AE) in an early period after subcutaneous implantable cardioverter defibrillator (S-ICD) implantation has been reported, however, no detailed data on air volume are available. We evaluated the subcutaneous air volume after implantation and its absorption rate one week after implantation.MethodsPatients who underwent S-ICD implantation in our hospital received chest CT scans immediately after implantation and followed up 1 week later. The total subcutaneous air volume, air around the generator, the distal electrode, and the proximal electrode within 3 cm were calculated using a three-dimensional workstation. Fat areas at the level of the lower edge of the generator were also analyzed.ResultFifteen patients received CT immediately after implantation. The mean age was 45.6 +/- 17.9 (66.7% of men), and the mean body mass index was 24.3 +/- 3.3. The three-incision technique was applied in seven patients and two-incision technique was in the latter eight patients. The mean total subcutaneous air volume was 18.54 +/- 7.50 mL. Air volume around the generator, the distal electrode, and the proximal electrode were 11.05 +/- 5.12, 0.72 +/- 0.72, and 0.88 +/- 0.87 mL, respectively. Twelve patients received a follow-up CT 1 week later. The mean total subcutaneous air was 0.25 +/- 0.45 mL, showing a 98.7% absorption rate.ConclusionAlthough subcutaneous air was observed in all patients after S-ICD implantation, most of the air was absorbed within 1 week, suggesting a low occurrence of AE-related IAS after a week postoperation.
引用
收藏
页码:496 / 502
页数:7
相关论文
共 50 条
  • [41] Subcutaneous Implantable Cardioverter-Defibrillator Shocks After Left Ventricular Assist Device Implantation
    Pfeffer, Tobias J.
    Koenig, Thorben
    Duncker, David
    Michalski, Roman
    Hohmann, Stephan
    Oswald, Hanno
    Schmitto, Jan D.
    Veltmann, Christian
    CIRCULATION-ARRHYTHMIA AND ELECTROPHYSIOLOGY, 2016, 9 (11):
  • [42] Intermuscular Subcutaneous Implantable Cardioverter-Defibrillator Implantation Without Defibrillation Testing
    Mistry, Amar
    Pooranachandran, Vivetha
    Dhutia, Harshil
    Chelliah, Rajesh
    Pathmanathan, Ravi K.
    JACC-CLINICAL ELECTROPHYSIOLOGY, 2021, 7 (01) : 124 - 126
  • [43] Novel Strategy of Subcutaneous Implantable Cardioverter Defibrillator Implantation Under Regional Anesthesia
    Braver, Omri
    Semyonov, Michael
    Reina, Yair
    Konstantino, Yuval
    Haim, Moti
    Winter, Joachim
    JOURNAL OF CARDIOTHORACIC AND VASCULAR ANESTHESIA, 2019, 33 (09) : 2513 - 2516
  • [44] Liposomal bupivacaine during subcutaneous implantable cardioverter defibrillator implantation for pain management
    Lee, Justin Z.
    Glasgow, Amy E.
    Habermann, Elizabeth B.
    Scott, Luis
    Shen, Win-Kuang
    Kusumoto, Fred M.
    McLeod, Christopher J.
    Goel, Vasudha
    Kaginele, Pranita
    Cha, Yong-Mei
    Friedman, Paul A.
    Asirvatham, Samuel J.
    Mulpuru, Siva K.
    PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY, 2021, 44 (03): : 513 - 518
  • [45] Two-incision technique for implantation of the subcutaneous implantable cardioverter-defibrillator
    Knops, Reinoud E.
    Nordkamp, Louise R. A. Olde
    de Groot, Joris R.
    Wilde, Arthur A. M.
    HEART RHYTHM, 2013, 10 (08) : 1240 - 1243
  • [46] Growth and the subcutaneous implantable cardioverter defibrillator in a small child
    Spentzou, Georgia
    McLean, Andrew
    McLeod, Karen
    CARDIOLOGY IN THE YOUNG, 2019, 29 (12) : 1524 - 1524
  • [47] The Subcutaneous Implantable Cardioverter-Defibrillator: A Patient Perspective
    Nesti, Martina
    Russo, Vincenzo
    Palama, Zefferino
    Panchetti, Luca
    Garibaldi, Silvia
    Startari, Umberto
    Mirizzi, Gianluca
    Piacenti, Marcello
    Rossi, Andrea
    Sciarra, Luigi
    JOURNAL OF CLINICAL MEDICINE, 2023, 12 (20)
  • [48] Subcutaneous implantable cardioverter–defibrillator: Current status and perspectives
    Niehues P.
    Frommeyer G.
    Reinke F.
    Eckardt L.
    Herzschrittmachertherapie + Elektrophysiologie, 2018, 29 (4) : 349 - 354
  • [49] Development of an Entirely Subcutaneous Implantable Cardioverter-Defibrillator
    Lupo, Pier Paolo
    Pelissero, Gabriele
    Ali, Hussam
    Sanghera, Rick
    Cappato, Riccardo
    PROGRESS IN CARDIOVASCULAR DISEASES, 2012, 54 (06) : 493 - 497
  • [50] Subcutaneous implantable cardioverter defibrillator position determines success
    Amin, A.
    Gold, M. R.
    Burke, M. C.
    Knight, B. P.
    Kalbfleisch, S. J.
    Okabe, T.
    Duffy, E.
    Stahl, W. K.
    Husby, M.
    Tyler, J.
    Houmsse, M.
    Daoud, E. G.
    Hummel, J. D.
    Augostini, R. S.
    Weiss, R.
    EUROPEAN HEART JOURNAL, 2018, 39 : 621 - 621