Subcutaneous air entrapment after subcutaneous implantable cardioverter defibrillator implantation evaluated by computed tomography

被引:0
|
作者
Taguchi, Yuka [1 ]
Ishikawa, Toshiyuki [1 ]
Matsumoto, Katsumi [1 ]
Narikawa, Masatoshi [1 ]
Okazaki, Yoshinori [1 ]
Miyagawa, Shuichi [1 ]
Horigome, Akira [1 ]
Hosoda, Junya [1 ]
机构
[1] Yokohama City Univ, Grad Sch Med, Dept Cardiol, 3-9 Fukuura,Kanazawa Ku, Yokohama, Kanagawa, Japan
来源
关键词
complications; ICD; inappropriate shock; oversensing; sensing malfunction; SHOCKS;
D O I
10.1111/pace.14962
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
BackgroundInappropriate shock (IAS) caused by subcutaneous air entrapment (AE) in an early period after subcutaneous implantable cardioverter defibrillator (S-ICD) implantation has been reported, however, no detailed data on air volume are available. We evaluated the subcutaneous air volume after implantation and its absorption rate one week after implantation.MethodsPatients who underwent S-ICD implantation in our hospital received chest CT scans immediately after implantation and followed up 1 week later. The total subcutaneous air volume, air around the generator, the distal electrode, and the proximal electrode within 3 cm were calculated using a three-dimensional workstation. Fat areas at the level of the lower edge of the generator were also analyzed.ResultFifteen patients received CT immediately after implantation. The mean age was 45.6 +/- 17.9 (66.7% of men), and the mean body mass index was 24.3 +/- 3.3. The three-incision technique was applied in seven patients and two-incision technique was in the latter eight patients. The mean total subcutaneous air volume was 18.54 +/- 7.50 mL. Air volume around the generator, the distal electrode, and the proximal electrode were 11.05 +/- 5.12, 0.72 +/- 0.72, and 0.88 +/- 0.87 mL, respectively. Twelve patients received a follow-up CT 1 week later. The mean total subcutaneous air was 0.25 +/- 0.45 mL, showing a 98.7% absorption rate.ConclusionAlthough subcutaneous air was observed in all patients after S-ICD implantation, most of the air was absorbed within 1 week, suggesting a low occurrence of AE-related IAS after a week postoperation.
引用
收藏
页码:496 / 502
页数:7
相关论文
共 50 条
  • [21] Outcomes of subcutaneous implantable cardioverter-defibrillator implantation in patients on hemodialysis
    Eduard Koman
    Ashwani Gupta
    Faiz Subzposh
    Heath Saltzman
    Steven P Kutalek
    Journal of Interventional Cardiac Electrophysiology, 2016, 45 : 219 - 223
  • [22] Implantation of the subcutaneous implantable cardioverter-defibrillator with truncal plane blocks
    Miller, Marc A.
    Bhatt, Himani V.
    Weiner, Menachem
    Brouwer, Tom F.
    Mittnacht, Alexander J.
    Shariat, Ali
    Jeng, Christina L.
    Eden, Caroline
    Lin, Hung-Mo
    Salter, Benjamin
    Dukkipati, Srinivas R.
    Reddy, Vivek Y.
    HEART RHYTHM, 2018, 15 (07) : 1108 - 1111
  • [23] CONSISTENT SUBCUTANEOUS PREPECTORAL IMPLANTATION OF A NEW IMPLANTABLE CARDIOVERTER-DEFIBRILLATOR
    STANTON, MS
    HAYES, DL
    MUNGER, TM
    TRUSTY, JM
    ESPINOSA, RE
    SHEN, WK
    OSBORN, MJ
    PACKER, DL
    HAMMILL, SC
    MAYO CLINIC PROCEEDINGS, 1994, 69 (04) : 309 - 314
  • [24] New surgical approach to implantation of the subcutaneous implantable cardioverter-defibrillator
    Kennergren, Charles E. H.
    EUROPACE, 2017, 19 (12): : 1907 - 1908
  • [25] Subcutaneous Implantable Cardioverter-Defibrillator Technology
    Gupta, Anurag
    Al-Ahmad, Amin
    Wang, Paul J.
    HEART FAILURE CLINICS, 2011, 7 (02) : 287 - +
  • [26] Subcutaneous Implantable Cardioverter Defibrillator: Early Experience
    Oliva Souza, Fernando Sergio
    Sepulvida, Vanessa
    Valerio Almeida, Marcos Cesar
    Weiss, Samuel Campagiotto
    Borges Rodrigues, Lucas Henrique
    Gori Lima, Pedro Augusto
    Mattos, Davi Bongiolo
    BRAZILIAN JOURNAL OF CARDIOVASCULAR SURGERY, 2017, 32 (06) : 498 - 502
  • [27] The subcutaneous implantable cardioverter defibrillator in 2019 and beyond
    van Dijk, Vincent F.
    Boersma, Lucas V. A.
    TRENDS IN CARDIOVASCULAR MEDICINE, 2020, 30 (06) : 378 - 384
  • [28] Reduction in electrocardiographic lateral precordial voltage after subcutaneous implantable cardioverter-defibrillator implantation
    Liang, Jackson J.
    Maeda, Shingo
    Supple, Gregory E.
    ACTA CARDIOLOGICA, 2017, 72 (02) : 221 - 222
  • [29] An unusual indication for subcutaneous implantable cardioverter/defibrillator
    De Schryver, Nicolas
    Van Ruyssevelt, Christian
    Capulzini, Lucio
    ACTA CARDIOLOGICA, 2017, 72 (04) : 487 - 488
  • [30] An Entirely Subcutaneous Implantable Cardioverter-Defibrillator
    Bardy, Gust H.
    Smith, Warren M.
    Hood, Margaret A.
    Crozier, Ian G.
    Melton, Iain C.
    Jordaens, Luc
    Theuns, Dominic
    Park, Robert E.
    Wright, David J.
    Connelly, Derek T.
    Fynn, Simon P.
    Murgatroyd, Francis D.
    Sperzel, Johannes
    Neuzner, Joerg
    Spitzer, Stefan G.
    Ardashev, Andrey V.
    Oduro, Amo
    Boersma, Lucas
    Maass, Alexander H.
    Van Gelder, Isabelle C.
    Wilde, Arthur A.
    van Dessel, Pascal F.
    Knops, Reinoud E.
    Barr, Craig S.
    Lupo, Pierpaolo
    Cappato, Riccardo
    Grace, Andrew A.
    NEW ENGLAND JOURNAL OF MEDICINE, 2010, 363 (01): : 36 - 44