Subcutaneous air entrapment after subcutaneous implantable cardioverter defibrillator implantation evaluated by computed tomography

被引:0
|
作者
Taguchi, Yuka [1 ]
Ishikawa, Toshiyuki [1 ]
Matsumoto, Katsumi [1 ]
Narikawa, Masatoshi [1 ]
Okazaki, Yoshinori [1 ]
Miyagawa, Shuichi [1 ]
Horigome, Akira [1 ]
Hosoda, Junya [1 ]
机构
[1] Yokohama City Univ, Grad Sch Med, Dept Cardiol, 3-9 Fukuura,Kanazawa Ku, Yokohama, Kanagawa, Japan
来源
关键词
complications; ICD; inappropriate shock; oversensing; sensing malfunction; SHOCKS;
D O I
10.1111/pace.14962
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
BackgroundInappropriate shock (IAS) caused by subcutaneous air entrapment (AE) in an early period after subcutaneous implantable cardioverter defibrillator (S-ICD) implantation has been reported, however, no detailed data on air volume are available. We evaluated the subcutaneous air volume after implantation and its absorption rate one week after implantation.MethodsPatients who underwent S-ICD implantation in our hospital received chest CT scans immediately after implantation and followed up 1 week later. The total subcutaneous air volume, air around the generator, the distal electrode, and the proximal electrode within 3 cm were calculated using a three-dimensional workstation. Fat areas at the level of the lower edge of the generator were also analyzed.ResultFifteen patients received CT immediately after implantation. The mean age was 45.6 +/- 17.9 (66.7% of men), and the mean body mass index was 24.3 +/- 3.3. The three-incision technique was applied in seven patients and two-incision technique was in the latter eight patients. The mean total subcutaneous air volume was 18.54 +/- 7.50 mL. Air volume around the generator, the distal electrode, and the proximal electrode were 11.05 +/- 5.12, 0.72 +/- 0.72, and 0.88 +/- 0.87 mL, respectively. Twelve patients received a follow-up CT 1 week later. The mean total subcutaneous air was 0.25 +/- 0.45 mL, showing a 98.7% absorption rate.ConclusionAlthough subcutaneous air was observed in all patients after S-ICD implantation, most of the air was absorbed within 1 week, suggesting a low occurrence of AE-related IAS after a week postoperation.
引用
收藏
页码:496 / 502
页数:7
相关论文
共 50 条
  • [1] Submuscular is Superior to Subcutaneous Implantation for Subcutaneous Implantable Cardioverter Defibrillator
    Winter, Joachim
    Clahsen, Harald
    Shin, Dong-In
    O'Connor, Stephen
    CIRCULATION, 2013, 128 (22)
  • [2] Air Entrapment Causing Inappropriate Shock From a Subcutaneous Implantable Cardioverter Defibrillator
    Yang, Ying Chi
    Thein Tun Aung
    Bailin, Steven J.
    Rhodes, Troy E.
    CARDIOLOGY RESEARCH, 2019, 10 (02) : 128 - 130
  • [3] SUBCUTANEOUS IMPLANTABLE CARDIOVERTER DEFIBRILLATOR IMPLANTATION IN PATIENTS AFTER DEVICE EXTRACTION
    Gupta, A.
    Subzposh, F.
    Saltzman, H.
    Kutalek, S. P.
    CARDIOLOGY, 2014, 128 : 336 - 336
  • [4] Subserratus implantation of the subcutaneous implantable cardioverter-defibrillator
    Smietana, Jeffrey
    Frankel, David S.
    Serletti, Joseph M.
    Arkles, Jeffrey
    Pothineni, Naga Venkata K.
    Marchlinski, Francis E.
    Schaller, Robert D.
    HEART RHYTHM, 2021, 18 (10) : 1799 - 1804
  • [5] Inappropriate Shock Due to Air Entrapment in the Pocket of a Subcutaneous Implantable Cardioverter-defibrillator
    Linhart, Markus
    Maria Tolosana, Jose
    Chipa, Fredy
    Trucco, Emilce
    Mont, Lluis
    REVISTA ESPANOLA DE CARDIOLOGIA, 2019, 72 (01): : 79 - 81
  • [6] Subcutaneous Implantable Cardioverter Defibrillator
    Rowley, Christopher P.
    Gold, Michael R.
    CIRCULATION-ARRHYTHMIA AND ELECTROPHYSIOLOGY, 2012, 5 (03): : 587 - 593
  • [7] The Subcutaneous Implantable Cardioverter Defibrillator
    Steffel, Jan
    EUROPEAN HEART JOURNAL, 2017, 38 (04) : 226 - 228
  • [8] The Subcutaneous Implantable Cardioverter Defibrillator
    Mangels, Daniel
    Frishman, William H.
    CARDIOLOGY IN REVIEW, 2016, 24 (05) : 248 - 255
  • [9] Substernal and subcutaneous electrocardiograms during subcutaneous implantable cardioverter-defibrillator implantation
    Yuji Ishida
    Taihei Itoh
    Shingo Sasaki
    Shota Washima
    Yuichi Toyama
    Hirofumi Tomita
    Journal of Interventional Cardiac Electrophysiology, 2022, 64 : 235 - 236
  • [10] Substernal and subcutaneous electrocardiograms during subcutaneous implantable cardioverter-defibrillator implantation
    Ishida, Yuji
    Itoh, Taihei
    Sasaki, Shingo
    Washima, Shota
    Toyama, Yuichi
    Tomita, Hirofumi
    JOURNAL OF INTERVENTIONAL CARDIAC ELECTROPHYSIOLOGY, 2022, 64 (01) : 235 - 236