Continuous Non-Invasive Blood Pressure Measurement Using 60 GHz-Radar-A Feasibility Study

被引:5
|
作者
Vysotskaya, Nastassia [1 ,2 ]
Will, Christoph [1 ]
Servadei, Lorenzo [3 ]
Maul, Noah [2 ]
Mandl, Christian [1 ]
Nau, Merlin [2 ]
Harnisch, Jens [1 ]
Maier, Andreas [2 ]
机构
[1] Infineon Technol AG, Campeon 1-15, D-85579 Neubiberg, Germany
[2] Friedrich Alexander Univ Erlangen Nuremberg FAU, Dept Comp Sci Pattern Recognit 5, Martensstr 3, D-91058 Erlangen, Germany
[3] Tech Univ Munich, Dept Elect & Comp Engn, Arcisstr 21, D-80333 Munich, Germany
关键词
FMCW radar; vital sensing; continuous blood pressure monitoring; signal processing; wearable device; PULSE-WAVE ANALYSIS; MEASURING DEVICES; TRANSIT-TIME; CALIBRATION; SOCIETY; PPG;
D O I
10.3390/s23084111
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
Blood pressure monitoring is of paramount importance in the assessment of a human's cardiovascular health. The state-of-the-art method remains the usage of an upper-arm cuff sphygmomanometer. However, this device suffers from severe limitations-it only provides a static blood pressure value pair, is incapable of capturing blood pressure variations over time, is inaccurate, and causes discomfort upon use. This work presents a radar-based approach that utilizes the movement of the skin due to artery pulsation to extract pressure waves. From those waves, a set of 21 features was collected and used-together with the calibration parameters of age, gender, height, and weight-as input for a neural network-based regression model. After collecting data from 55 subjects from radar and a blood pressure reference device, we trained 126 networks to analyze the developed approach's predictive power. As a result, a very shallow network with just two hidden layers produced a systolic error of 9.2 +/- 8.3mmHg (mean error +/- standard deviation) and a diastolic error of 7.7 +/- 5.7mmHg. While the trained model did not reach the requirements of the AAMI and BHS blood pressure measuring standards, optimizing network performance was not the goal of the proposed work. Still, the approach has displayed great potential in capturing blood pressure variation with the proposed features. The presented approach therefore shows great potential to be incorporated into wearable devices for continuous blood pressure monitoring for home use or screening applications, after improving this approach even further.
引用
收藏
页数:36
相关论文
共 50 条
  • [1] Feasibility of continuous non-invasive blood pressure measurement in infants
    Schraa, O
    Andriessen, P
    Smit, B
    van den Bosch-Ruis, W
    van Koppen, M
    Settels, JJ
    [J]. JOURNAL OF HYPERTENSION, 2004, 22 : S280 - S280
  • [2] Feasibility study of continuous non-invasive blood pressure measurement in critically ill children
    C Hofhuizen
    J Lemson
    L van de Sant
    O Schraa
    J Settels
    J van der Hoeven
    [J]. Critical Care, 10 (Suppl 1):
  • [3] The System of Indicators of the Continuous Non-Invasive Blood Pressure Measurement
    Timokhov, Gennady, V
    Semenova, Evgeniia A.
    Pivovarov, Vladimir V.
    [J]. PROCEEDINGS OF THE 2021 IEEE CONFERENCE OF RUSSIAN YOUNG RESEARCHERS IN ELECTRICAL AND ELECTRONIC ENGINEERING (ELCONRUS), 2021, : 1860 - 1864
  • [4] Continuous Non-Invasive Blood Pressure Monitoring Using Photoplethysmography
    Nye, Ross
    Zhang, Zhe
    Fang, Qiang
    [J]. 2015 INTERNATIONAL SYMPOSIUM ON BIOELECTRONICS AND BIOINFORMATICS (ISBB), 2015, : 176 - 179
  • [5] Feasibility of non-invasive measurement of central blood pressure and arterial stiffness in shock
    Markakis, Konstantinos
    Pagonas, Nikolaos
    Georgianou, Eleni
    Zgoura, Panagiota
    Rohn, Benjamin J.
    Bertram, Sebastian
    Seidel, Maximilian
    Bettag, Sebastian
    Trappe, Hans-Joachim
    Babel, Nina
    Westhoff, Timm H.
    Seibert, Felix S.
    [J]. EUROPEAN JOURNAL OF CLINICAL INVESTIGATION, 2021, 51 (09)
  • [6] Design and Implementation of Non-invasive Continuous Blood Pressure Measurement and Monitoring System Using Photoplethysmography
    Islam, Md. Aminul
    Ahmad, Mohiuddin
    [J]. 2018 10TH INTERNATIONAL CONFERENCE ON ELECTRICAL AND COMPUTER ENGINEERING (ICECE), 2018, : 173 - 176
  • [7] MRI-compatible non-invasive continuous blood pressure measurement using fiber optics
    Harja, Juha
    Myllyla, Teemu S.
    Sorvoja, Hannu S. S.
    Myllyla, Risto A.
    Elseoud, Ahmed A.
    Nikkinen, Juha
    Kiviniemi, Vesa
    Tervonen, Osmo
    [J]. SARATOV FALL MEETING 2009: INTERNATIONAL SCHOOL FOR JUNIOR SCIENTISTS AND STUDENTS ON OPTICS, LASER PHYSICS, AND BIOPHOTONICS, 2010, 7547
  • [8] Continuous Non-Invasive Blood Pressure Measurement Using Pulse Transit Time in Children and Adolescents
    Zagrada, Elena
    Krickau, Tobias
    Rascher, Wolfgang
    [J]. KLINISCHE PADIATRIE, 2019, 231 (02): : 67 - 73
  • [9] Continuous non-invasive estimates of cerebral blood flow using electrocardiography signals: a feasibility study
    van Bohemen, Samuel J. J.
    Rogers, Jeffrey M. M.
    Boughton, Philip C. C.
    Clarke, Jillian L. L.
    Valderrama, Joaquin T. T.
    Kyme, Andre Z. Z.
    [J]. BIOMEDICAL ENGINEERING LETTERS, 2023, 13 (02) : 185 - 195
  • [10] Continuous non-invasive estimates of cerebral blood flow using electrocardiography signals: a feasibility study
    Samuel J. van Bohemen
    Jeffrey M. Rogers
    Philip C. Boughton
    Jillian L. Clarke
    Joaquin T. Valderrama
    Andre Z. Kyme
    [J]. Biomedical Engineering Letters, 2023, 13 : 185 - 195