MRI-compatible non-invasive continuous blood pressure measurement using fiber optics

被引:0
|
作者
Harja, Juha [1 ]
Myllyla, Teemu S. [1 ]
Sorvoja, Hannu S. S. [1 ]
Myllyla, Risto A. [1 ]
Elseoud, Ahmed A. [2 ]
Nikkinen, Juha [2 ]
Kiviniemi, Vesa [2 ]
Tervonen, Osmo [2 ]
机构
[1] Univ Oulu, Optoelectron & Measurement Tech Lab, Oulu, Finland
[2] Oulu Univ Hosp, Dept Diagnost Radiol, Oulu, Finland
关键词
Non-invasive blood pressure measurement; fibre optical accelerometer; magnetic resonance imaging; pulse wave velocity; MR compatibility;
D O I
10.1117/12.855056
中图分类号
O43 [光学];
学科分类号
070207 ; 0803 ;
摘要
This report focuses on designing and implementing a non-invasive blood pressure measuring device capable of being used during magnetic resonance imaging. This device is based on measuring pulse wave velocity in arterial blood and using the obtained result to estimate diastolic blood pressure. Pulse transit times are measured by two fibre optical accelerometers placed over chest and carotid artery. The fabricated accelerometer contains two static fibres and a cantilever beam. The free end of the beam is angled at 90 degrees to act as a reflecting surface. Optical fibres are used for both illuminating the surface and receiving the reflected light. Acceleration applied to the sensor causes deflection of the beam, whereupon the amount of reflected light changes. The sensor output voltage is proportional to the intensity of the reflected light. Tests conducted on the electronics and sensors inside an MRI room during scanning proved that the device is MR conditional. No artifacts or distortions were detected.
引用
收藏
页数:7
相关论文
共 50 条
  • [1] Feasibility of continuous non-invasive blood pressure measurement in infants
    Schraa, O
    Andriessen, P
    Smit, B
    van den Bosch-Ruis, W
    van Koppen, M
    Settels, JJ
    [J]. JOURNAL OF HYPERTENSION, 2004, 22 : S280 - S280
  • [2] The System of Indicators of the Continuous Non-Invasive Blood Pressure Measurement
    Timokhov, Gennady, V
    Semenova, Evgeniia A.
    Pivovarov, Vladimir V.
    [J]. PROCEEDINGS OF THE 2021 IEEE CONFERENCE OF RUSSIAN YOUNG RESEARCHERS IN ELECTRICAL AND ELECTRONIC ENGINEERING (ELCONRUS), 2021, : 1860 - 1864
  • [3] Non-Invasive Blood Flow Speed Measurement Using Optics
    Zhang, Alex Ce
    Lo, Yu-Hwa
    [J]. SENSORS, 2022, 22 (03)
  • [4] Continuous Non-Invasive Blood Pressure Monitoring Using Photoplethysmography
    Nye, Ross
    Zhang, Zhe
    Fang, Qiang
    [J]. 2015 INTERNATIONAL SYMPOSIUM ON BIOELECTRONICS AND BIOINFORMATICS (ISBB), 2015, : 176 - 179
  • [5] Design and testing of an MRI-compatible cycle ergometer for non-invasive cardiac assessments during exercise
    Silmara Gusso
    Carlo Salvador
    Paul Hofman
    Wayne Cutfield
    James C Baldi
    Andrew Taberner
    Poul Nielsen
    [J]. BioMedical Engineering OnLine, 11
  • [6] Design and Implementation of Non-invasive Continuous Blood Pressure Measurement and Monitoring System Using Photoplethysmography
    Islam, Md. Aminul
    Ahmad, Mohiuddin
    [J]. 2018 10TH INTERNATIONAL CONFERENCE ON ELECTRICAL AND COMPUTER ENGINEERING (ICECE), 2018, : 173 - 176
  • [7] Continuous Non-Invasive Blood Pressure Measurement Using Pulse Transit Time in Children and Adolescents
    Zagrada, Elena
    Krickau, Tobias
    Rascher, Wolfgang
    [J]. KLINISCHE PADIATRIE, 2019, 231 (02): : 67 - 73
  • [8] Design and testing of an MRI-compatible cycle ergometer for non-invasive cardiac assessments during exercise
    Gusso, Silmara
    Salvador, Carlo
    Hofman, Paul
    Cutfield, Wayne
    Baldi, James C.
    Taberner, Andrew
    Nielsen, Poul
    [J]. BIOMEDICAL ENGINEERING ONLINE, 2012, 11
  • [9] MRI-COMPATIBLE GRASPING FORCE SENSOR WITH AN INCLINED DOUBLE PARALLEL STRUCTURE USING FIBER OPTICS
    Arata, Jumpei
    Terakawa, Shogo
    Fujimoto, Hideo
    Sulzer, James
    Gassert, Roger
    [J]. PROCEEDINGS OF THE ASME/ISCIE INTERNATIONAL SYMPOSIUM ON FLEXIBLE AUTOMATION, ISFA 2012, 2013, : 77 - 82
  • [10] Non-invasive blood pressure measurement displays
    Scott, D. H. T.