Accuracy and impact on quality of life of real-time continuous glucose monitoring in children with hyperinsulinaemic hypoglycaemia

被引:3
|
作者
Sivasubramanian, Madhini [1 ,2 ,3 ]
Avari, Parizad [4 ]
Gilbert, Clare [1 ]
Doodson, Louise [1 ]
Morgan, Kate [1 ]
Oliver, Nick [4 ]
Shah, Pratik [1 ,2 ,5 ]
机构
[1] Great Ormond St Hosp Children NHS Fdn Trust, Dept Paediat Endocrinol, London, England
[2] UCL, Inst Child Hlth, London WC1N, England
[3] Univ Sunderland London, Fac Hlth Sci & Wellbeing, London, England
[4] Imperial Coll London, Dept Metab Digest & Reprod, London, England
[5] Royal London Childrens Hosp, Dept Paediat Endocrinol, Barts Hlth NHS Trust, London, England
来源
基金
芬兰科学院;
关键词
hyperinsulinism; continuous glucose monitoring; hypoglycaemia; neurodevelopment; time below range; INSULIN PUMP THERAPY; AWARENESS; ADULTS;
D O I
10.3389/fendo.2023.1265076
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
Objective: Continuous glucose monitoring (CGM) is the standard of care for glucose monitoring in children with diabetes, however there are limited data reporting their use in hyperinsulinaemic hypoglycaemia (HH). Here, we evaluate CGM accuracy and its impact on quality of life in children with HH.Methods: Real-time CGM (Dexcom G5 and G6) was used in children with HH aged 0-16years. Data from self-monitoring capillary blood glucose (CBG) and CGM were collected over a period of up to 28days and analysed. Quality of life was assessed by the PedsQL4.0 general module and PedsQL2.0 family impact module, completed by children and their parents/carers before and after CGM insertion. Analysis of accuracy metrics included mean absolute relative difference (MARD) and proportion of CGM values within 15, 20, and 30% or 15, 20, and 30 mg/dL of reference glucose values >100 mg/dL or <= 100 mg/dL, respectively (% 15/15, % 20/20, % 30/30). Clinical reliability was assessed with Clarke error grid (CEG) analyses.Results: Prospective longitudinal study with data analysed from 40 children. The overall MARD between reference glucose and paired CGM values (n=4,928) was 13.0% (Dexcom G5 12.8%, Dexcom G6 13.1%). The proportion of readings meeting %15/15 and %20/20 were 77.3% and 86.4%, respectively, with CEG analysis demonstrating 97.4% of all values in zones A and B. Within the hypoglycaemia range (<70 mg/dL), the median ARD was 11.4% with a sensitivity and specificity of 64.2% and 91.3%, respectively. Overall PedsQL child report at baseline and endpoint were 57.6 (50.5 - 75.8) and 87.0 (82.9 - 91.2), and for parents were 60.3 (44.8 - 66.0) and 85.3 (83.7 - 91.3), respectively (both p<0.001).Conclusion: Use of CGM for children with HH is feasible, with clinically acceptable accuracy, particularly in the hypoglycaemic range. Quality of life measures demonstrate significant improvement after CGM use. These data are important to explore use of CGM in disease indications, including neonatal and paediatric diabetes, cystic fibrosis and glycogen storage disorders.
引用
收藏
页数:9
相关论文
共 50 条
  • [31] SUSTAINED REDUCTIONS IN SEVERE HYPOGLYCAEMIA WITH CONTINUOUS GLUCOSE MONITORING: REAL LIFE CLINICAL EXPERIENCE
    Slattery, D.
    Iftikhar, M.
    Brackenridge, A.
    Hopkins, D.
    Amiel, S.
    Choudhary, P.
    DIABETES TECHNOLOGY & THERAPEUTICS, 2018, 20 : A105 - A106
  • [32] Continuous glucose monitoring for children with hypoglycaemia: Evidence in 2023
    Worth, Chris
    Hoskyns, Lucy
    Salomon-Estebanez, Maria
    Nutter, Paul W.
    Harper, Simon
    Derks, Terry G. J.
    Beardsall, Kathy
    Banerjee, Indraneel
    FRONTIERS IN ENDOCRINOLOGY, 2023, 14
  • [34] Continuous glucose monitoring for hypoglycaemia in children: Perspectives in 2020
    Worth, Chris
    Dunne, Mark
    Ghosh, Arunabha
    Harper, Simon
    Banerjee, Indraneel
    PEDIATRIC DIABETES, 2020, 21 (05) : 697 - 706
  • [35] Real-Time Continuous Glucose Monitoring in Adolescents and Young Adults With Type 2 Diabetes Can Improve Quality of Life
    Chesser, Hannah
    Srinivasan, Shylaja
    Puckett, Cassidy
    Gitelman, Stephen E.
    Wong, Jenise C.
    JOURNAL OF DIABETES SCIENCE AND TECHNOLOGY, 2024, 18 (04): : 911 - 919
  • [36] THE IMPACT OF SWITCHING FROM FLASH GLUCOSE MONITORING TO REAL-TIME CGM ON HYPOGLYCAEMIA IN ADULTS WITH TYPE 1 DIABETES AND IMPAIRED AWARENESS OF HYPOGLYCAEMIA
    Reddy, M.
    Jugnee, N.
    Anantharaja, S.
    Oliver, N.
    DIABETES TECHNOLOGY & THERAPEUTICS, 2018, 20 : A25 - A25
  • [37] New perspectives on real-time continuous glucose monitoring Comment
    Stechova, Katerina
    LANCET CHILD & ADOLESCENT HEALTH, 2021, 5 (04): : 235 - 236
  • [38] Real-time continuous glucose monitoring: proven clinical indiciations
    Gomez Peralta, Fernando
    Abreu Padin, Cristina
    AVANCES EN DIABETOLOGIA, 2011, 27 (06): : 198 - 203
  • [39] DIABETES Under Control Real-Time Continuous Glucose Monitoring
    Ramchandani, Neesha
    Saadon, Yael
    Jornsay, Donna
    AMERICAN JOURNAL OF NURSING, 2010, 110 (04) : 60 - 63
  • [40] Is reimbursement for alerts and real-time continuous glucose monitoring needed?
    Norgaard, Kirsten
    Schmidt, Signe
    LANCET, 2021, 397 (10291): : 2230 - 2232