Investigation of Optimal Light Source Wavelength for Cuffless Blood Pressure Estimation Using a Single Photoplethysmography Sensor

被引:3
|
作者
Toda, Sogo [1 ]
Matsumura, Kenta [2 ]
机构
[1] Natl Inst Technol, Ishikawa Coll, Tsubata 9290392, Japan
[2] Univ Toyama, Fac Med, Toyama 9300194, Japan
关键词
blood pressure; heart rate; modified normalized pulse volume; optical wavelength; photoplethysmography; PULSE TRANSIT-TIME; HEART-RATE; OPTICAL-PROPERTIES; PROGNOSTIC VALUE; HYPERTENSION; VARIABILITY; RESPONSES; FINGER; VOLUME; RANGE;
D O I
10.3390/s23073689
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
Routine blood pressure measurement is important for the early detection of various diseases. Recently, cuffless blood pressure estimation methods that do not require cuff pressurization have attracted attention. In this study, we investigated the effect of the light source wavelength on the accuracy of blood pressure estimation using only two physiological indices that can be calculated with photoplethysmography alone, namely, heart rate and modified normalized pulse volume. Using a newly developed photoplethysmography sensor that can simultaneously measure photoplethysmograms at four wavelengths, we evaluated its estimation accuracy for systolic blood pressure, diastolic blood pressure, and mean arterial pressure against a standard cuff sphygmomanometer. Mental stress tasks were used to alter the blood pressure of 14 participants, and multiple linear regression analysis showed the best light sources to be near-infrared for systolic blood pressure and blue for both diastolic blood pressure and mean arterial pressure. The importance of the light source wavelength for the photoplethysmogram in cuffless blood pressure estimation was clarified.
引用
收藏
页数:17
相关论文
共 50 条
  • [1] Inverse-model-based cuffless blood pressure estimation using a single photoplethysmography sensor
    Suzuki, Arata
    PROCEEDINGS OF THE INSTITUTION OF MECHANICAL ENGINEERS PART H-JOURNAL OF ENGINEERING IN MEDICINE, 2015, 229 (07) : 499 - 505
  • [2] Cuffless Blood Pressure Estimation Using Single Channel Photoplethysmography: A Two-Step Method
    Khalid, Syed Ghufran
    Liu, Haipeng
    Zia, Tahir
    Zhang, Jufen
    Chen, Fei
    Zheng, Dingchang
    IEEE ACCESS, 2020, 8 : 58146 - 58154
  • [3] Noninvasive Cuffless Estimation of Blood Pressure using Photoplethysmography without Electrocardiograph Measurement
    Samria, Rohan
    Jain, Ridhi
    Jha, Ankita
    Saini, Sandeep
    Chowdhury, Shubhajit Roy
    2014 IEEE REGION 10 SYMPOSIUM, 2014, : 254 - 257
  • [4] Cuffless Blood Pressure Estimation using only Photoplethysmography based on Cardiovascular parameters
    Fukushima, Hayato
    Kawanaka, Haruki
    Bhuiyan, Md. Shoaib
    Oguri, Koji
    2013 35TH ANNUAL INTERNATIONAL CONFERENCE OF THE IEEE ENGINEERING IN MEDICINE AND BIOLOGY SOCIETY (EMBC), 2013, : 2132 - 2135
  • [5] Cuffless Single-Site Photoplethysmography for Blood Pressure Monitoring
    Hosanee, Manish
    Chan, Gabriel
    Welykholowa, Kaylie
    Cooper, Rachel
    Kyriacou, Panayiotis A.
    Zheng, Dingchang
    Allen, John
    Abbott, Derek
    Menon, Carlo
    Lovell, Nigel H.
    Howard, Newton
    Chan, Wee-Shian
    Lim, Kenneth
    Fletcher, Richard
    Ward, Rabab
    Elgendi, Mohamed
    JOURNAL OF CLINICAL MEDICINE, 2020, 9 (03)
  • [6] Cuffless Blood Pressure Estimation Based on Monte Carlo Simulation Using Photoplethysmography Signals
    Haque, Chowdhury Azimul
    Kwon, Tae-Ho
    Kim, Ki-Doo
    SENSORS, 2022, 22 (03)
  • [7] Features Extraction for Cuffless Blood Pressure Estimation by Autoencoder from Photoplethysmography
    Shimazaki, Shota
    Bhuiyan, Shoaib
    Kawanaka, Haruki
    Oguri, Koji
    2018 40TH ANNUAL INTERNATIONAL CONFERENCE OF THE IEEE ENGINEERING IN MEDICINE AND BIOLOGY SOCIETY (EMBC), 2018, : 2857 - 2860
  • [8] Live Demonstration: A Novel Cuffless Photoplethysmography Sensor for Continuous Blood Pressure measurement
    Kao, Yung-Hua
    Chao, Paul C. -P.
    Hung, Yueh
    Wey, Chin-Long
    2017 IEEE SENSORS, 2017, : 470 - 470
  • [9] Cuffless blood pressure estimation from photoplethysmography using deep convolutional neural network and transfer learning
    Koparir, Hueseyin Murat
    Arslan, Ozkan
    BIOMEDICAL SIGNAL PROCESSING AND CONTROL, 2024, 93
  • [10] Photoplethysmography-based cuffless blood pressure estimation: an image encoding and fusion approach
    Liu, Yinsong
    Yu, Junsheng
    Mou, Hanlin
    PHYSIOLOGICAL MEASUREMENT, 2023, 44 (12)