Optimised quality control method for determination of the radiochemical purity of [99mTc]Tc-mebrofenin and [99mTc]Tc-etifenin in a clinical context

被引:1
|
作者
Pirot, Charlotte [1 ,2 ,4 ]
Peyronnet, Damien [1 ,2 ]
Vigne, Jonathan [1 ,2 ,3 ]
机构
[1] UNICAEN, Dept Nucl Med, Caen, France
[2] UNICAEN, Dept Pharm, Caen, France
[3] Normandie Univ, Inst Blood & Brain Caen Normandie, Ctr Cyceron, UNICAEN,INSERM U1237,PhIND, F-14000 Caen, France
[4] UNICAEN, Dept Nucl Med, F-14032 Caen, France
关键词
quality control; practice guideline; nuclear medicine; pharmacy service; hospital; pharmaceutical preparations;
D O I
10.1136/ejhpharm-2022-003512
中图分类号
R9 [药学];
学科分类号
1007 ;
摘要
ObjectivesIn the context of a supply disruption of mebrofenin (Cholediam)-based kits for radiolabelling with technetium-99m [Tc-99m], the medicine agencies allowed the importation of a back-up radiopharmaceutical diagnostic agent, etifenin (Techida), to ensure continuous management of patients with hepatobiliary disorders in nuclear medicine departments. There are still issues regarding the measurement of radiochemical purity (RCP) with these kits based on the European Pharmacopoeia and the Summary of Product Characteristics (SPC). This study aims to identify and to optimise, in a clinical context, the most suitable thin layer chromatography (TLC) method for the determination of the RCP in terms of speed of response and reliability for [Tc-99m]Tc-mebrofenin and [Tc-99m]Tc-etifenin. Methods[Tc-99m]Tc-etifenin (n=4) and [Tc-99m]Tc-mebrofenin (n=5) were individually controlled using six different TLC methods and one high-performance liquid chromatography (HPLC) method for impurity identification ([Tc-99m](TcO2)(n) and Na[Tc-99m]TcO4), RCP (%) and duration of analysis (min). Two TLC methods were selected according to the recommendations of the Pharmacopoeia and SPC, two others were exactly the same but with a heating step, and the other two corresponded to a mix between the methods of the SPC and the Pharmacopoeia that were chosen to optimise RCP determination parameters. ResultsRadio-HPLC analysis allowed effective separation of [Tc-99m]Tc-etifenin and [Tc-99m]Tc-mebrofenin with a retention time of 8.05 +/- 0.02 min and 8.94 +/- 0.07 min, respectively, from Na[Tc-99m]TcO4 (retention time 2.76 +/- 0.03 min). HPLC showed an absence of Na[Tc-99m]TcO4 for [Tc-99m]Tc-mebrofenin and 0.2% for [Tc-99m]Tc-etifenin. Among the TLC methods, we identified the most suitable method which ensures the most compliant RCP (98.3 +/- 0.9%) in a time of 31.5 +/- 1.1 min. Also, it allowed a time saving of 15 min compared with the methods proposed by the Pharmacopoeia and the SPCs. ConclusionWe propose a TLC method that accelerates quality control by an average of 15 min while guaranteeing a reliable RCP.
引用
收藏
页码:376 / 380
页数:5
相关论文
共 50 条
  • [21] The use of 99mTc as a tracer in the determination of 99Tc by ICP-mass spectrometry
    Más, JL
    Garcia-León, M
    Bolívar, JP
    Sánchez-Angulo, C
    JOURNAL OF ANALYTICAL ATOMIC SPECTROMETRY, 2000, 15 (10) : 1369 - 1373
  • [22] ανβ3 integrin-selective RGDechi peptides labelled with [99mTc][Tc(N)(PNP)]- and [99mTc][Tc-HYNIC]- systems: an in vitro comparison
    Bolzati, C.
    Gobbi, C.
    Salvarese, N.
    Giorgio, A.
    Carpanese, D.
    Rossi, V.
    Melendez-Alafort, L.
    Rosato, A.
    Saviano, M.
    Del Gatto, A.
    Zaccaro, L.
    EUROPEAN JOURNAL OF NUCLEAR MEDICINE AND MOLECULAR IMAGING, 2023, 50 (SUPPL 1) : S435 - S435
  • [23] Sentinel lymph node detection in oral cancer: a head to head comparison between [99mTc]Tc-Tilmanocept and [99mTc]Tc-Nanocoll
    Mahieu, R.
    den Toom, I. J.
    van Rooij, R.
    van Es, R. J. J.
    Hobbelink, M. G.
    Krijger, G. C.
    Tijink, B. M.
    de Keizer, B.
    de Bree, R.
    EUROPEAN JOURNAL OF NUCLEAR MEDICINE AND MOLECULAR IMAGING, 2020, 47 (SUPPL 1) : S234 - S234
  • [24] A new method of [99mTc]-ciprofloxacin preparation and quality control
    Rodriguez-Puig, Daniel
    Piera, Carlos
    Fuster, David
    Soriano, Alex
    Sierra, Jose Maria
    Rubi, Sebastia
    Suades, Joan
    JOURNAL OF LABELLED COMPOUNDS & RADIOPHARMACEUTICALS, 2006, 49 (13): : 1171 - 1176
  • [25] CALCULATION OF 99Mo, 99MTc AND 99Tc IN THE GENERATOR AND IN THE ELUATE.
    Thieme, K.
    International journal of radiation applications and instrumentation. Part A, Applied radiation and isotopes, 1988, 39 (03): : 267 - 268
  • [26] Sentinel lymph node detection in oral cancer: a within-patient comparison between [99mTc]Tc-tilmanocept and [99mTc]Tc-nanocolloid
    Inne J. den Toom
    Rutger Mahieu
    Rob van Rooij
    Robert J. J. van Es
    Monique G. G. Hobbelink
    Gerard C. Krijger
    Bernard M. Tijink
    Bart de Keizer
    Remco de Bree
    European Journal of Nuclear Medicine and Molecular Imaging, 2021, 48 : 851 - 858
  • [27] Bladder hernia in the inguinal canal mimicking pubic metastasis on [99mTC] Tc-MDP and [99mTc]Tc-HYNIC-PSMA whole body scans
    Roshanravan, Vahid
    Soltani, Salman
    Sadeghi, Ramin
    Rezayat, Alireza Akhavan
    Esmatinia, Mohammad
    Aghaee, Atena
    IRANIAN JOURNAL OF NUCLEAR MEDICINE, 2022, 30 (02): : 140 - 143
  • [28] Sentinel lymph node detection in oral cancer: a within-patient comparison between [99mTc]Tc-tilmanocept and [99mTc]Tc-nanocolloid
    den Toom, Inne J.
    Mahieu, Rutger
    van Rooij, Rob
    van Es, Robert J. J.
    Hobbelink, Monique G. G.
    Krijger, Gerard C.
    Tijink, Bernard M.
    de Keizer, Bart
    de Bree, Remco
    EUROPEAN JOURNAL OF NUCLEAR MEDICINE AND MOLECULAR IMAGING, 2021, 48 (03) : 851 - 858
  • [29] Chemical characterization and biodistribution, of 99Tc and 99mTc glutathione in rats.
    Baba, K
    Moretti, JL
    Ercan, MT
    EUROPEAN JOURNAL OF NUCLEAR MEDICINE, 1999, 26 (09): : 1194 - 1194
  • [30] On the use of 99Mo/99mTc generators in the analysis of low levels of 99Tc in environmental samples by radiochemical methods
    Dowdall, M
    Selnæs, OG
    Gwynn, JP
    Lind, B
    WATER AIR AND SOIL POLLUTION, 2004, 156 (1-4): : 287 - 297