A Universal Device to Convert a Continuous Flow Assist Device to a Pulsatile Flow Device to Simulate Normal Blood Flow and Pressure Patterns

被引:0
|
作者
Gerrah, Rabin [1 ]
机构
[1] Stanford Univ, Dept Cardiothorac Surg, CVRB Falk Bldg,300 Pasteur Dr, Stanford, CA 94305 USA
关键词
assist devices; pulsatility; cardiovascular; hemodynamics;
D O I
10.1177/15533506221140534
中图分类号
R61 [外科手术学];
学科分类号
摘要
BackgroundContinuous follow assist devices (CFAD) are the most commonly used mechanical circulatory support devices. Compared to Pulsatile flow assist devices (PFAD), CFADs deliver a non-physiologic type of flow, which might contribute to complications related to lack of pulsatility in these devices. Moreover, lack of pulsatility complicates the clinical management of these patients who often present with good perfusion but with no palpable pulse and none or a negligible pulse pressure on blood pressure measurement.Methods and resultsPresented here is a concept of a universal converter device that can be added inline other CFADs to convert the flow from continuous to pulsatile, simulating a normal flow and pressure pattern. After initial implantation and stabilization with a CFAD, adding this converter might potentially provide the benefits of pulsatile physiologic flow. The device is made of 2 components connected in parallel, working in tandem in user determined cycles. The continuous flow through a specifically positioned openings create a smooth conversion to a pulsatile flow. This device can convert a continuous flow to a physiologic pulsatile flow to achieve a native-like flow pattern and potentially prevent some CFAD complications.ConclusionThis paper presents the concept of pulsatility generation and simulation for other assist devices. Such a device can be a universal add-on or a supplemental option for CFADs.
引用
收藏
页码:471 / 476
页数:6
相关论文
共 50 条
  • [31] Pulsatile versus Continuous Flow Ventricular Assist Device as a Bridge to Transplantation for the Pediatric Failing Fontan
    Seymour, E.
    Tunuguntla, H.
    Sosa, S.
    Hope, K.
    Spinner, J.
    Ankola, A.
    Adachi, I.
    Puri, K.
    JOURNAL OF HEART AND LUNG TRANSPLANTATION, 2024, 43 (04): : S628 - S628
  • [32] Comparison of a pulsatile and a continuous flow left ventricular assist device in high-risk PCI
    Samol, Alexander
    Wiemer, Marcus
    Kaese, Sven
    INTERNATIONAL JOURNAL OF CARDIOLOGY, 2022, 360 : 7 - 12
  • [33] Accuracy of Doppler blood pressure measurement in continuous-flow left ventricular assist device patients
    Li, Song
    Beckman, Jennifer A.
    Welch, Nathan G.
    Bjelkengren, Jason
    Masri, Sofia Carolina
    Minami, Elina
    Stempien-Otero, April
    Levy, Wayne C.
    O'Brien, Kevin D.
    Lin, Shin
    Farris, Stephen D.
    Chen, Richard K.
    Wood, Gregory
    Koomalsingh, Kevin
    Kirkpatrick, James
    McCabe, James
    Leary, Peter J.
    Chassagne, Fanette
    Chivukula, Venkat Keshav
    Aliseda, Alberto
    Mahr, Claudius
    ESC HEART FAILURE, 2019, 6 (04): : 793 - 798
  • [34] Pulse Oximeter Derived Blood Pressure Measurement in Patients With a Continuous Flow Left Ventricular Assist Device
    Hellman, Yaron
    Malik, Adnan S.
    Lane, Kathleen A.
    Shen, Changyu
    Wang, I-Wen
    Wozniak, Thomas C.
    Hashmi, Zubair A.
    Munson, Sarah D.
    Pickrell, Jeanette
    Caccamo, Marco A.
    Gradus-Pizlo, Irmina
    Hadi, Azam
    ARTIFICIAL ORGANS, 2017, 41 (05) : 424 - 430
  • [35] In Vivo Validation of Pulsatile Flow and Differential Pressure Estimation Models in a Left Ventricular Assist Device
    Zhang, X. T.
    AlOmari, A. H.
    Savkin, A. V.
    Ayre, P. J.
    Lim, E.
    Salamonsen, R. F.
    Rosenfeldt, F. L.
    Lovell, N. H.
    2010 ANNUAL INTERNATIONAL CONFERENCE OF THE IEEE ENGINEERING IN MEDICINE AND BIOLOGY SOCIETY (EMBC), 2010, : 2517 - 2520
  • [36] The new criterion of pulsatile blood flow estimation based on extracorporeal heart assist device POLVAD
    不详
    INTERNATIONAL JOURNAL OF ARTIFICIAL ORGANS, 2005, 28 (09): : 885 - 885
  • [37] Design of a continuous flow centrifugal pediatric ventricular assist device
    Throckmorton, AL
    Wood, HG
    Day, SW
    Song, X
    Click, PC
    Allaire, PE
    Olsen, DB
    INTERNATIONAL JOURNAL OF ARTIFICIAL ORGANS, 2003, 26 (11): : 1015 - 1031
  • [38] Effect of Body Position on Continuous Flow Left Ventricular Assist Device Flow Dynamics
    Muthiah, K.
    Gupta, S.
    Robson, D.
    Walker, R.
    Macdonald, P. S.
    Jansz, P.
    Hayward, C. S.
    JOURNAL OF HEART AND LUNG TRANSPLANTATION, 2013, 32 (04): : S233 - S233
  • [39] RELATION BETWEEN LEFT VENTRICULAR ASSIST DEVICE FLOW AND RECOVERY FROM ORGAN DYSFUNCTION AFTER CONTINUOUS FLOW VENTRICULAR ASSIST DEVICE IMPLANTATION
    Enokidani, Y.
    Uji, M.
    Uchiyama, A.