Performance evaluation of the LumiraDx quantitative microfluidic point-of-care CRP test

被引:0
|
作者
Ellis, Jayne [1 ]
Harnett, James [2 ]
Cameron, Gregor [2 ]
Moss, Phil [3 ]
Gray, Alasdair [4 ]
机构
[1] JEMMDx Ltd, Bedford MK40 2QW, England
[2] Univ Coll London Hosp NHS Fdn Trust, London, England
[3] St George Hosp, ED Clin Res Unit, London, England
[4] Royal Infirm Edinburgh NHS Trust, Emergency Med Res Grp EMERGE, Edinburgh, Scotland
关键词
C-reactive protein; CRP; Lower respiratory tract infection; Point-of-care test; Primary care; Secondary care; C-REACTIVE PROTEIN;
D O I
10.1016/j.plabm.2023.e00349
中图分类号
R446 [实验室诊断]; R-33 [实验医学、医学实验];
学科分类号
1001 ;
摘要
C-reactive protein (CRP) is an established acute-phase marker for infection, inflammation and tissue injury, used to guide clinical decision-making in primary and secondary care. This study compared the analytical performance of the quantitative microfluidic point-of-care LumiraDx CRP Test to a laboratory-based reference method (Siemens RCRP Flex assay on the Dimension (R) Xpand (R)) and evaluated equivalence of sample matrices (blood versus plasma) in point-of-care settings using samples from patients presenting with symptoms of infection or inflammation. The LumiraDx CRP Test demonstrated close agreement with the lab reference test (range, 5.1 to 245.2 mg/L, r = 0.992, slope = 0.998, intercept = -0.476; n = 205) and notable agreement between fingerstick and venous blood and plasma (r = 0.974-0.983; n = 44). Paired replicate precision had mean coefficients of variation of 6.4 % (plasma), 6.6 % (capillary direct) and 8.1 % (venous blood); overall error rates were 2.9 %. The quantitative LumiraDx CRP Test showed robust analytical performance across sample matrices and close agreement compared to the laboratory reference method when used at the point of care.
引用
收藏
页数:6
相关论文
共 50 条
  • [1] Performance evaluation of the LumiraDx quantitative microfluidic point-of-care CRP test in pediatric emergency
    Masri, W.
    Djeddou, R.
    Cung, A.
    Vignaud, O.
    Bendaoud, M.
    CLINICA CHIMICA ACTA, 2024, 558
  • [2] Performance Evaluation of the Quantitative Point-of-Care LumiraDx d-Dimer Test
    Ellis, Jayne E.
    Johnston, Thomas W.
    Craig, David
    Scribner, Anita
    Simon, William
    Kirstein, Judith
    CARDIOLOGY AND THERAPY, 2021, 10 (02) : 547 - 559
  • [3] Performance Evaluation of the Quantitative Point-of-Care LumiraDx d-Dimer Test
    Jayne E. Ellis
    Thomas W. Johnston
    David Craig
    Anita Scribner
    William Simon
    Judith Kirstein
    Cardiology and Therapy, 2021, 10 : 547 - 559
  • [4] Evaluation of the point-of-care Afinion™ CRP test
    Van Aelst, S.
    Peeters, L.
    Cardinaels, H.
    Verjans, M-P.
    Raes, M.
    Mewis, A.
    ACTA CLINICA BELGICA, 2015, 70 : S9 - S10
  • [5] Performance of the LumiraDx Microfluidic Immunofluorescence Point-of-Care SARS-CoV-2 Antigen Test in Asymptomatic Adults and Children
    Drain, Paul
    Sulaiman, Raed
    Hoppers, Melanie
    Lindner, Nigel M.
    Lawson, Vicki
    Ellis, Jayne E.
    AMERICAN JOURNAL OF CLINICAL PATHOLOGY, 2022, 157 (04) : 602 - 607
  • [6] Evaluation of the Performance of a Point-of-Care Test for Chlamydia and Gonorrhea
    Van der Pol, Barbara
    Taylor, Stephanie N.
    Mena, Leandro
    Lebed, Joel
    McNeil, Candice Joy
    Crane, LaShonda
    Ermel, Aaron
    Sukhija-Cohen, Adam
    Gaydos, Charlotte A.
    JAMA NETWORK OPEN, 2020, 3 (05)
  • [7] EVALUATION OF A CANINE C-REACTIVE PROTEIN (CRP) POINT-OF-CARE TEST
    Plickert, H.
    Kohn, B.
    Brunnberg, L.
    Einspanier, R.
    JOURNAL OF VETERINARY INTERNAL MEDICINE, 2010, 24 (03) : 777 - 778
  • [8] Performance of the LumiraDx Platform INR Test in an Anticoagulation Clinic Point-of-Care Setting Compared With an Established Laboratory Reference Method
    Tait, Robert Campbell
    Hung, Annielle
    Gardner, Roy S.
    CLINICAL AND APPLIED THROMBOSIS-HEMOSTASIS, 2019, 25
  • [9] Analytical accuracy and userfriendliness of the Afinion point-of-care CRP test
    Verbakel, Jan Y.
    Aertgeerts, Bert
    Lemiengre, Marieke
    De Sutter, An
    Bullens, Dominique M. A.
    Buntinx, Frank
    JOURNAL OF CLINICAL PATHOLOGY, 2014, 67 (01) : 83 - +
  • [10] A multicenter evaluation of a point of care CRP Test
    Scharnhorst, Volkher
    Noordzij, Peter G.
    Lutz, Andrea
    Graser, Uwe
    Puntener, Daniel
    Alquezar-Arbe, Aitor
    CLINICAL BIOCHEMISTRY, 2019, 71 : 38 - 45