Performance evaluation of the LumiraDx quantitative microfluidic point-of-care CRP test

被引:0
|
作者
Ellis, Jayne [1 ]
Harnett, James [2 ]
Cameron, Gregor [2 ]
Moss, Phil [3 ]
Gray, Alasdair [4 ]
机构
[1] JEMMDx Ltd, Bedford MK40 2QW, England
[2] Univ Coll London Hosp NHS Fdn Trust, London, England
[3] St George Hosp, ED Clin Res Unit, London, England
[4] Royal Infirm Edinburgh NHS Trust, Emergency Med Res Grp EMERGE, Edinburgh, Scotland
关键词
C-reactive protein; CRP; Lower respiratory tract infection; Point-of-care test; Primary care; Secondary care; C-REACTIVE PROTEIN;
D O I
10.1016/j.plabm.2023.e00349
中图分类号
R446 [实验室诊断]; R-33 [实验医学、医学实验];
学科分类号
1001 ;
摘要
C-reactive protein (CRP) is an established acute-phase marker for infection, inflammation and tissue injury, used to guide clinical decision-making in primary and secondary care. This study compared the analytical performance of the quantitative microfluidic point-of-care LumiraDx CRP Test to a laboratory-based reference method (Siemens RCRP Flex assay on the Dimension (R) Xpand (R)) and evaluated equivalence of sample matrices (blood versus plasma) in point-of-care settings using samples from patients presenting with symptoms of infection or inflammation. The LumiraDx CRP Test demonstrated close agreement with the lab reference test (range, 5.1 to 245.2 mg/L, r = 0.992, slope = 0.998, intercept = -0.476; n = 205) and notable agreement between fingerstick and venous blood and plasma (r = 0.974-0.983; n = 44). Paired replicate precision had mean coefficients of variation of 6.4 % (plasma), 6.6 % (capillary direct) and 8.1 % (venous blood); overall error rates were 2.9 %. The quantitative LumiraDx CRP Test showed robust analytical performance across sample matrices and close agreement compared to the laboratory reference method when used at the point of care.
引用
收藏
页数:6
相关论文
共 50 条
  • [31] Point-of-care microfluidic devices for pathogen detection
    Nasseri, Behzad
    Soleimani, Neda
    Rabiee, Navid
    Kalbasi, Alireza
    Karimi, Mahdi
    Hamblin, Michael R.
    BIOSENSORS & BIOELECTRONICS, 2018, 117 : 112 - 128
  • [32] Facile and Ultrafast Microfluidic Photothermal PCR for Autonomous and Quantitative Point-of-Care Pathogen Detection
    Zou, Yaowei
    Liu, Mingxu
    Gao, Zixi
    Xue, Yaohua
    Qiu, Jieyu
    Zhang, Huizhen
    Li, Xinying
    Zhang, Chunsun
    Shu, Bowen
    SMALL, 2025,
  • [33] Pregnancy evaluation with a point-of-care pregnancy test in dairy cattle
    Szelenyi, Zoltan
    Lipthay, Ildiko
    Santa, Attila
    Lenart, Lea
    Repasi, Atilla
    Szenci, Otto
    THERIOGENOLOGY, 2024, 214 : 201 - 205
  • [34] Evaluation of a rapid oral fluid point-of-care test for MDMA
    Wilson, Lisa
    Jehanli, Ahmed
    Hand, Chris
    Cooper, Gail
    Smith, Robert
    JOURNAL OF ANALYTICAL TOXICOLOGY, 2007, 31 (02) : 98 - 104
  • [35] Clinical performance of a point-of-care Coccidioides antibody test in dogs
    Reagan, Krystle L.
    McHardy, Ian
    Thompson, George R., III
    Sykes, Jane E.
    JOURNAL OF VETERINARY INTERNAL MEDICINE, 2021, 35 (02) : 965 - 969
  • [36] Evaluation of a new quantitative point-of-care test platform for urine-based detection of bladder cancer
    Ritter, Rene
    Hennenlotter, Joerg
    Kuehs, Ursula
    Aufderklamm, Stefan
    Blutbacher, Pia
    Deja, Angelika
    Hohneder, Andrea
    Gerber, Valentina
    Gakis, Georgios
    Stenzl, Arnulf
    Schwentner, Christian
    Todenhoefer, Tilman
    UROLOGIC ONCOLOGY-SEMINARS AND ORIGINAL INVESTIGATIONS, 2014, 32 (03) : 337 - 344
  • [37] A Point-of-Care Prothrombin Time Test on a Microfluidic Disk Analyzer Using Alternate Spinning
    Lin, Chia-Hui
    Lin, Kun-Wei
    Yen, Daniel
    Shih, Chih-Hsin
    Lu, Chien-Hsing
    Wang, Jiunn-Min
    Lin, Chi-Yu
    JOURNAL OF NANOSCIENCE AND NANOTECHNOLOGY, 2015, 15 (02) : 1401 - 1407
  • [39] Evaluation of quantitative cardiac biomarker point-of-care testing in the emergency department
    Caragher, TE
    Fernandez, BB
    Jacobs, FL
    Barr, LA
    JOURNAL OF EMERGENCY MEDICINE, 2002, 22 (01): : 1 - 7
  • [40] Evaluation of a point-of-care system for quantitative determination of troponin T and myoglobin
    Müller-Bardorff, M
    Sylvén, C
    Rasmanis, G
    Jorgensen, B
    Collinson, PO
    Waldenhofer, U
    Hirschl, MM
    Laggner, AN
    Gerhardt, W
    Hafner, G
    Labaere, I
    Leinberger, R
    Zerback, R
    Katus, HA
    CLINICAL CHEMISTRY AND LABORATORY MEDICINE, 2000, 38 (06) : 567 - 574