Evaluating the magnet response in deep subcutaneous implanted cardioverter defibrillator implants

被引:1
|
作者
Richman, Tuppence [1 ,4 ]
Stanton, Tony [1 ,2 ]
Fryer, Michael [1 ]
Dayananda, Naresh [1 ]
Tung, Matthew [1 ,3 ]
机构
[1] Sunshine Coast Univ Hosp, Birtinya, Australia
[2] Univ Sunshine Coast, Sippy Downs, Australia
[3] Griffith Univ, Sch Med & Dent, Sunshine Coast, Australia
[4] 6 Doherty St, Birtinya, Qld 4557, Australia
来源
关键词
implantable cardiac devices; magnet response; subcutaneous implantable cardioverter defibrillators; ELECTRONIC DEVICES; MANAGEMENT; ARRHYTHMIAS;
D O I
10.1111/pace.14609
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
IntroductionThe manufacturer of subcutaneous implantable cardioverter defibrillators (S-ICDs) acknowledges that 'deep implants' may fail to elicit a magnet response, however, does not define 'deep implant' or recommend a maximum implant depth. This study aims to systematically evaluate the effect of subcutaneous tissue depth and magnet types on evoked magnet response. MethodsSunshine Coast University Hospital's S-ICD cohort underwent magnet response evaluation; where bar and donut magnets were compared and the evoked magnet response was recorded in three separate zones, guided by a template. Ordinal regression (OR) models assessed the relationship between the evoked magnet response and tissue depth (TD), measured via post-implant X-Ray. The patient's ability to hear the magnet response audible tone was recorded. ResultsPatients (n = 39) with measurable TD (n = 30) were analyzed. The bar magnet evoked a magnet response in all zones in 53% of patients, compared with 73% of patients with the donut magnet (p = 0.18). The relationship between bar magnet response and TD showed the odds of an evoked magnet response decreased by 11% every 1 mm increase in TD (OR of 0.89, p < 0.01), whereas the donut magnet decreased by 16% per 1 mm (OR of 0.84, p < 0.01). Directly over the S-ICD was the most effective in evoking magnet response with the bar (85% of patients), and off-centre was most effective for the donut magnet (100%). BMI and Praetorian score were not significantly associated with magnet response. We found 23% of patients were unable to detect the audible tone. ConclusionWe observed a statistically significant association between TD and ability to evoke magnet response. The bar magnet was less reliable than the donut magnet for therapy inhibition in deep implants.
引用
收藏
页码:93 / 99
页数:7
相关论文
共 50 条
  • [31] The experiences of women with a subcutaneous implantable cardioverter defibrillator
    Van der Stuijt, W.
    Quast, A. B. E.
    Baalman, S. W. E.
    Wilde, A. A. M.
    Knops, R. E.
    EUROPEAN HEART JOURNAL, 2019, 40 : 3651 - 3651
  • [32] Subcutaneous implantable cardioverter defibrillator in cardiomyopathies and channelopathies
    Migliore, Federico
    Pelliccia, Francesco
    Autore, Camillo
    Bertaglia, Emanuele
    Cecchi, Franco
    Curcio, Antonio
    Bontempi, Luca
    Curnis, Antonio
    De Filippo, Paolo
    D'Onofrio, Antonio
    Francia, Pietro
    Maurizi, Niccolo
    Musumeci, Beatrice
    Proclemer, Alessandro
    Zorzi, Alessandro
    Corrado, Domenico
    JOURNAL OF CARDIOVASCULAR MEDICINE, 2018, 19 (11) : 633 - 642
  • [33] Who Should Receive the Subcutaneous Implanted Defibrillator? The Subcutaneous Implantable Cardioverter Defibrillator (ICD) Should Be Considered in all ICD Patients Who Do Not Require Pacing
    Poole, Jeanne E.
    Gold, Michael R.
    CIRCULATION-ARRHYTHMIA AND ELECTROPHYSIOLOGY, 2013, 6 (06): : 1236 - 1244
  • [34] Role of deep learning methods in screening for subcutaneous implantable cardioverter defibrillator in heart failure
    ElRefai, Mohamed
    Abouelasaad, Mohamed
    Wiles, Benedict M.
    Dunn, Anthony J.
    Coniglio, Stefano
    Zemkoho, Alain B.
    Morgan, John M.
    Roberts, Paul R.
    ANNALS OF NONINVASIVE ELECTROCARDIOLOGY, 2023, 28 (01)
  • [35] Efficacy of sotalol and metoprolol in patients with an implanted cardioverter defibrillator
    Kuhlkamp, V
    Mewis, C
    Dornberger, V
    Suchalla, R
    Mermi, J
    Seipel, L
    JOURNAL OF THE AMERICAN COLLEGE OF CARDIOLOGY, 1997, 29 (02) : 7453 - 7453
  • [36] The occupational routine of patients implanted with cardioverter-defibrillator
    Mannini, Juliana
    Nascimento, Janaina Santos
    Pelosi, Miryam Bonadiu
    CADERNOS DE TERAPIA OCUPACIONAL DA UFSCAR, 2015, 23 (01): : 31 - 42
  • [37] Fitness to drive in patients with implanted cardioverter-defibrillator
    Klein, H. H.
    DEUTSCHE MEDIZINISCHE WOCHENSCHRIFT, 2012, 137 (39) : 1927 - 1930
  • [38] Troubleshooting Implanted Cardioverter Defibrillator Sensing Problems I
    Swerdlow, Charles D.
    Asirvatham, Samuel J.
    Ellenbogen, Kenneth A.
    Friedman, Paul A.
    CIRCULATION-ARRHYTHMIA AND ELECTROPHYSIOLOGY, 2014, 7 (06): : 1237 - 1261
  • [39] Cognitive behavioural therapy in patients with an implanted cardioverter defibrillator
    Hoefer, Stefan
    EUROPEAN JOURNAL OF PREVENTIVE CARDIOLOGY, 2020, 27 (03) : 256 - 257
  • [40] Quality of life over time with an implanted cardioverter defibrillator
    McKinley, S
    Pelletier, D
    MittenLewis, S
    Gallagher, RD
    Squire, JK
    CIRCULATION, 1997, 96 (08) : 2454 - 2454