Evaluating the magnet response in deep subcutaneous implanted cardioverter defibrillator implants

被引:1
|
作者
Richman, Tuppence [1 ,4 ]
Stanton, Tony [1 ,2 ]
Fryer, Michael [1 ]
Dayananda, Naresh [1 ]
Tung, Matthew [1 ,3 ]
机构
[1] Sunshine Coast Univ Hosp, Birtinya, Australia
[2] Univ Sunshine Coast, Sippy Downs, Australia
[3] Griffith Univ, Sch Med & Dent, Sunshine Coast, Australia
[4] 6 Doherty St, Birtinya, Qld 4557, Australia
来源
关键词
implantable cardiac devices; magnet response; subcutaneous implantable cardioverter defibrillators; ELECTRONIC DEVICES; MANAGEMENT; ARRHYTHMIAS;
D O I
10.1111/pace.14609
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
IntroductionThe manufacturer of subcutaneous implantable cardioverter defibrillators (S-ICDs) acknowledges that 'deep implants' may fail to elicit a magnet response, however, does not define 'deep implant' or recommend a maximum implant depth. This study aims to systematically evaluate the effect of subcutaneous tissue depth and magnet types on evoked magnet response. MethodsSunshine Coast University Hospital's S-ICD cohort underwent magnet response evaluation; where bar and donut magnets were compared and the evoked magnet response was recorded in three separate zones, guided by a template. Ordinal regression (OR) models assessed the relationship between the evoked magnet response and tissue depth (TD), measured via post-implant X-Ray. The patient's ability to hear the magnet response audible tone was recorded. ResultsPatients (n = 39) with measurable TD (n = 30) were analyzed. The bar magnet evoked a magnet response in all zones in 53% of patients, compared with 73% of patients with the donut magnet (p = 0.18). The relationship between bar magnet response and TD showed the odds of an evoked magnet response decreased by 11% every 1 mm increase in TD (OR of 0.89, p < 0.01), whereas the donut magnet decreased by 16% per 1 mm (OR of 0.84, p < 0.01). Directly over the S-ICD was the most effective in evoking magnet response with the bar (85% of patients), and off-centre was most effective for the donut magnet (100%). BMI and Praetorian score were not significantly associated with magnet response. We found 23% of patients were unable to detect the audible tone. ConclusionWe observed a statistically significant association between TD and ability to evoke magnet response. The bar magnet was less reliable than the donut magnet for therapy inhibition in deep implants.
引用
收藏
页码:93 / 99
页数:7
相关论文
共 50 条
  • [1] Who Should Receive the Subcutaneous Implanted Defibrillator? Timing Is Not Right to Replace the Transvenous Implantable Cardioverter Defibrillator Response
    Poole, Jeanne E.
    Gold, Michael R.
    CIRCULATION-ARRHYTHMIA AND ELECTROPHYSIOLOGY, 2013, 6 (06): : 1251 - 1251
  • [2] Safety and Tolerability of Implanted Subcutaneous Cardioverter-Defibrillator Systems
    Vereshchagina, A. V.
    Uskach, T. M.
    Sapelnikov, O., V
    Amanatova, V. A.
    Grishin, I. R.
    Kulikov, A. A.
    Kostin, V. S.
    Akchurin, R. S.
    RATIONAL PHARMACOTHERAPY IN CARDIOLOGY, 2022, 18 (04) : 427 - 432
  • [3] Impairment of Quality of Life in Patients with Implanted Subcutaneous Cardioverter Defibrillator (S-ICD) Compared to Implanted Transvenous Cardioverter Defibrillator Therapy
    Thienel, Manuela
    Haum, Magda
    Sadoni, Sebastian
    Novotny, Julia
    Estner, Heidi L.
    Fichtner, Stephanie
    Lackermair, Korbinian
    PATIENT PREFERENCE AND ADHERENCE, 2022, 16 : 3027 - 3033
  • [4] Fracture of the subcutaneous patch electrode in a patient with an implanted cardioverter-defibrillator
    Milasinovic, G
    Jelic, V
    Savic, D
    Pavlovic, SU
    Velinovic, M
    CIRCULATION JOURNAL, 2005, 69 (01) : 116 - 118
  • [5] Interference of subcutaneous implantable cardioverter defibrillator by deep brain stimulation
    Nayak, Hemal M.
    Sauser, Emilie
    Xie, Tao
    PARKINSONISM & RELATED DISORDERS, 2020, 81 : 75 - 77
  • [6] Subcutaneous Implantable Cardioverter Defibrillator
    Rowley, Christopher P.
    Gold, Michael R.
    CIRCULATION-ARRHYTHMIA AND ELECTROPHYSIOLOGY, 2012, 5 (03): : 587 - 593
  • [7] The Subcutaneous Implantable Cardioverter Defibrillator
    Steffel, Jan
    EUROPEAN HEART JOURNAL, 2017, 38 (04) : 226 - 228
  • [8] The Subcutaneous Implantable Cardioverter Defibrillator
    Mangels, Daniel
    Frishman, William H.
    CARDIOLOGY IN REVIEW, 2016, 24 (05) : 248 - 255
  • [9] Who Should Receive the Subcutaneous Implanted Defibrillator? Timing Is Not Right to Replace the Transvenous Implantable Cardioverter Defibrillator
    Acha, Moshe Rav
    Milan, David
    CIRCULATION-ARRHYTHMIA AND ELECTROPHYSIOLOGY, 2013, 6 (06): : 1246 - 1251
  • [10] Who Should Receive the Subcutaneous Implanted Defibrillator? The Subcutaneous Implantable Cardioverter Defibrillator (ICD) Should Be Considered in all ICD Patients Who Do Not Require Pacing Response
    Acha, Moshe Rav
    Milan, David
    CIRCULATION-ARRHYTHMIA AND ELECTROPHYSIOLOGY, 2013, 6 (06): : 1244 - 1245