Comparison of spin-echo echo planar imaging and gradient-recalled echo sequences in magnetic resonance elastography of liver at 1.5T same MRI scanner

被引:0
|
作者
Abe, Ryouna [1 ]
Fukuzawa, Kei [1 ]
Yoshihara, Chiharu [1 ]
Tano, Masakatsu [1 ]
Saitoh, Satoshi [2 ]
机构
[1] Toranomon Gen Hosp, Dept Radiol Technol, 2-2-2 Toranomon,Minato Ku, Tokyo, Japan
[2] Toranomon Gen Hosp, Dept Hepatol, 2-2-2 Toranomon,Minato Ku, Tokyo, Japan
关键词
Liver MRI; GRE MRE; SE-EPI MRE; R2*; Liver stiffness; HEPATIC-FIBROSIS; DISEASE; QUANTIFICATION; MULTIFREQUENCY; PERFORMANCE; STEATOSIS;
D O I
10.1007/s00261-023-04098-3
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Purpose Magnetic resonance elastography (MRE) is used to measure liver stiffness with gradient-recalled echo (GRE)-based and spin-echo echo planar imaging (SE-EPI)-based sequences. We compared the liver stiffness (LS) values of the two sequences on a 1.5-T MR imaging scanner.Methods This is a retrospective study. An MRE imaging section was obtained from a horizontal section of the liver. Region of interest was drawn on the elastogram, and the mean LS and pixel values were measured and compared. The correlations between proton density fat fraction, R2* values, and biochemical data from electronic medical records were confirmed, and multivariate analysis was performed.Results The mean LS values were 3.01 +/- 1.78 kPa for GRE and 3.13 +/- 1.57 kPa for SE-EPI, showing excellent agreement and a strong correlation between the two sequences (correlation coefficient r = 0.96). The mean pixel values were 369.5 +/- 142.7 pixels for GRE and 490.1 +/- 197.9 pixels for SE-EPI, showing a significant difference by the Wilcoxon rank sum test (p < 0.01). There were no LS unmeasurable cases in SE-EPI, but seven (2.5%) were unmeasurable in GRE, and multivariate analysis showed a significant difference with p < 0.001 in R2* values (mean, 92.7 Hz) for the GRE method.Conclusion The GRE and SE-EPI methods were comparable for LS measurements in 1.5-T liver MRE, indicating that SE-EPI MRE is more useful because GRE MRE may not measure cases with high R2* values and the region of interest tends to be smaller.
引用
收藏
页码:694 / 702
页数:9
相关论文
共 50 条
  • [41] MR imaging of the rotator cuff mechanism: Comparison of spin-echo and turbo spin-echo sequences
    Sonin, AH
    Peduto, AJ
    Fitzgerald, SW
    Callahan, CM
    Bresler, ME
    AMERICAN JOURNAL OF ROENTGENOLOGY, 1996, 167 (02) : 333 - 338
  • [42] Snapshot gradient-recalled echo-planar images of rat brains at long echo time at 9.4 T
    Lei, Hongxia
    Mlynarik, Vladimir
    Just, Nathalie
    Gruetter, Rolf
    MAGNETIC RESONANCE IMAGING, 2008, 26 (07) : 954 - 960
  • [43] CONTRIBUTION OF ORAL MAGNETIC PARTICLES IN MR IMAGING OF THE ABDOMEN WITH SPIN-ECHO AND GRADIENT-ECHO SEQUENCES
    BOUDGHENE, FP
    BACHGANSMO, T
    GRANGE, JD
    LAME, S
    NANTOIS, C
    WALLAYS, C
    BIGOT, JM
    JMRI-JOURNAL OF MAGNETIC RESONANCE IMAGING, 1993, 3 (01): : 107 - 112
  • [44] Diagnostic Performance of Spin-Echo Echo-Planar Imaging Magnetic Resonance Elastography in 3T System for Noninvasive Assessment of Hepatic Fibrosis
    Kim, Se Woo
    Lee, Jeong Min
    Park, Sungeun
    Joo, Ijin
    Yoon, Jeong Hee
    Chang, Won
    Kim, Haeryoung
    KOREAN JOURNAL OF RADIOLOGY, 2022, 23 (02) : 180 - 188
  • [45] INTRACRANIAL HEMORRHAGE - GRADIENT-ECHO MR IMAGING AT 1.5-T - COMPARISON WITH SPIN-ECHO IMAGING AND CLINICAL-APPLICATIONS
    ATLAS, SW
    MARK, AS
    GROSSMAN, RI
    GOMORI, JM
    RADIOLOGY, 1988, 168 (03) : 803 - 807
  • [46] MR imaging of focal liver lesions: Comparison between turbo spin-echo and conventional spin-echo pulse sequences
    Giovagnoni, A
    Paci, E
    Ercolani, P
    Kiefer, B
    Santino, P
    Piga, A
    MAGNETIC RESONANCE MATERIALS IN PHYSICS BIOLOGY AND MEDICINE, 1995, 3 (3-4): : 143 - 150
  • [47] The response of human tumors to carbogen breathing, monitored by Gradient-Recalled Echo Magnetic Resonance Imaging
    Griffiths, JR
    Taylor, NJ
    Howe, FA
    Saunders, MI
    Robinson, SP
    Hoskin, PJ
    Powell, MEB
    Thoumine, M
    Caine, LA
    Baddeley, H
    INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 1997, 39 (03): : 697 - 701
  • [48] Feasibility and agreement of stiffness measurements using gradient-echo and spin-echo MR elastography sequences in unselected patients undergoing liver MRI
    Cunha, Guilherme Moura
    Glaser, Kevin J.
    Bergman, Anke
    Luz, Rodrigo P.
    De Figueiredo, Eduardo H.
    Proenca Lobo Lopes, Flavia Paiva
    BRITISH JOURNAL OF RADIOLOGY, 2018, 91 (1087):
  • [49] MRI of the intracerebral lesions at 0.5 tesla: Comparison of fast spin-echo and conventional spin-echo sequences
    Incesu, L
    Gunes, M
    Akan, H
    Selcuk, MB
    COMPUTERIZED MEDICAL IMAGING AND GRAPHICS, 1996, 20 (02) : 105 - 113
  • [50] GRASE (GRADIENT-ECHO AND SPIN-ECHO) IMAGING - A NOVEL FAST MRI TECHNIQUE
    OSHIO, K
    FEINBERG, DA
    MAGNETIC RESONANCE IN MEDICINE, 1991, 20 (02) : 344 - 349