Comparison of spin-echo echo planar imaging and gradient-recalled echo sequences in magnetic resonance elastography of liver at 1.5T same MRI scanner

被引:0
|
作者
Abe, Ryouna [1 ]
Fukuzawa, Kei [1 ]
Yoshihara, Chiharu [1 ]
Tano, Masakatsu [1 ]
Saitoh, Satoshi [2 ]
机构
[1] Toranomon Gen Hosp, Dept Radiol Technol, 2-2-2 Toranomon,Minato Ku, Tokyo, Japan
[2] Toranomon Gen Hosp, Dept Hepatol, 2-2-2 Toranomon,Minato Ku, Tokyo, Japan
关键词
Liver MRI; GRE MRE; SE-EPI MRE; R2*; Liver stiffness; HEPATIC-FIBROSIS; DISEASE; QUANTIFICATION; MULTIFREQUENCY; PERFORMANCE; STEATOSIS;
D O I
10.1007/s00261-023-04098-3
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Purpose Magnetic resonance elastography (MRE) is used to measure liver stiffness with gradient-recalled echo (GRE)-based and spin-echo echo planar imaging (SE-EPI)-based sequences. We compared the liver stiffness (LS) values of the two sequences on a 1.5-T MR imaging scanner.Methods This is a retrospective study. An MRE imaging section was obtained from a horizontal section of the liver. Region of interest was drawn on the elastogram, and the mean LS and pixel values were measured and compared. The correlations between proton density fat fraction, R2* values, and biochemical data from electronic medical records were confirmed, and multivariate analysis was performed.Results The mean LS values were 3.01 +/- 1.78 kPa for GRE and 3.13 +/- 1.57 kPa for SE-EPI, showing excellent agreement and a strong correlation between the two sequences (correlation coefficient r = 0.96). The mean pixel values were 369.5 +/- 142.7 pixels for GRE and 490.1 +/- 197.9 pixels for SE-EPI, showing a significant difference by the Wilcoxon rank sum test (p < 0.01). There were no LS unmeasurable cases in SE-EPI, but seven (2.5%) were unmeasurable in GRE, and multivariate analysis showed a significant difference with p < 0.001 in R2* values (mean, 92.7 Hz) for the GRE method.Conclusion The GRE and SE-EPI methods were comparable for LS measurements in 1.5-T liver MRE, indicating that SE-EPI MRE is more useful because GRE MRE may not measure cases with high R2* values and the region of interest tends to be smaller.
引用
收藏
页码:694 / 702
页数:9
相关论文
共 50 条
  • [21] MR IMAGING OF THE CAVERNOUS SINUS - VALUE OF SPIN-ECHO AND GRADIENT RECALLED ECHO IMAGES
    DANIELS, DL
    CZERVIONKE, LF
    BONNEVILLE, JF
    CATTIN, F
    MARK, LP
    PECH, P
    HENDRIX, LE
    SMITH, DF
    HAUGHTON, VM
    WILLIAMS, AL
    AMERICAN JOURNAL OF NEURORADIOLOGY, 1988, 9 (05) : 947 - 952
  • [22] MR IMAGING OF THE CAVERNOUS SINUS - VALUE OF SPIN-ECHO AND GRADIENT RECALLED ECHO IMAGES
    DANIELS, DL
    CZERVIONKE, LF
    BONNEVILLE, JF
    CATTIN, F
    MARK, LP
    PECH, P
    HENDRIX, LE
    SMITH, DF
    HAUGHTON, VM
    WILLIAMS, AL
    AMERICAN JOURNAL OF ROENTGENOLOGY, 1988, 151 (05) : 1009 - 1014
  • [23] COMPARISON OF GRADIENT-RECALLED-ECHO AND T2-WEIGHTED SPIN-ECHO PULSE SEQUENCES IN INTRAMEDULLARY SPINAL LESIONS
    KATZ, BH
    QUENCER, RM
    HINKS, RS
    AMERICAN JOURNAL OF NEURORADIOLOGY, 1989, 10 (04) : 815 - 822
  • [24] Comparison of spin echo, gradient echo and fat saturation magnetic resonance imaging sequences for imaging the canine elbow
    Snaps, FR
    Saunders, JH
    Park, RD
    Daenen, B
    Balligand, MH
    Dondelinger, RF
    VETERINARY RADIOLOGY & ULTRASOUND, 1998, 39 (06) : 518 - 523
  • [25] MR elastography of liver fibrosis: preliminary results comparing spin-echo and echo-planar imaging
    Laurent Huwart
    Najat Salameh
    Leon ter Beek
    Eric Vicaut
    Frank Peeters
    Ralph Sinkus
    Bernard E. Van Beers
    European Radiology, 2008, 18 : 2535 - 2541
  • [26] MR elastography of liver fibrosis: preliminary results comparing spin-echo and echo-planar imaging
    Huwart, Laurent
    Salameh, Najat
    ter Beek, Leon
    Vicaut, Eric
    Peeters, Frank
    Sinkus, Ralph
    Van Beers, Bernard E.
    EUROPEAN RADIOLOGY, 2008, 18 (11) : 2535 - 2541
  • [27] Fast detection of diffuse axonal damage in severe traumatic brain injury: Comparison of gradient-recalled echo and turbo Proton Echo Planar Spectroscopic Imaging MRI sequences
    Giugni, E
    Sabatini, U
    Hagberg, GE
    Formisano, R
    Castriota-Scanderbeg, A
    AMERICAN JOURNAL OF NEURORADIOLOGY, 2005, 26 (05) : 1140 - 1148
  • [28] MR-IMAGING OF THE BRAIN - COMPARISON OF GRADIENT-ECHO AND SPIN-ECHO PULSE SEQUENCES
    PUI, MH
    FOK, ECM
    AMERICAN JOURNAL OF ROENTGENOLOGY, 1995, 165 (04) : 959 - 962
  • [30] MR IMAGING OF PARKINSON DISEASE WITH SPIN-ECHO AND GRADIENT-ECHO SEQUENCES
    BRAFFMAN, BH
    GROSSMAN, RI
    GOLDBERG, HI
    STERN, MB
    HURTIG, HI
    HACKNEY, DB
    BILANIUK, LT
    ZIMMERMAN, RA
    AMERICAN JOURNAL OF ROENTGENOLOGY, 1989, 152 (01) : 159 - 165