Simulation study of a planar quadrupole mass filter for MEMS mass spectrometer

被引:1
|
作者
Cheng, Yu -Peng [1 ,2 ]
Shen, Chen [1 ,2 ]
Wu, Zhang-Xu [1 ,2 ]
Li, Shan [2 ]
Liu, You-Jiang [2 ]
Wang, Han [1 ,2 ]
Chen, Chi-Lai [2 ]
机构
[1] Univ Sci & Technol China, Hefei 230026, Peoples R China
[2] Chinese Acad Sci, Hefei Inst Phys Sci, Hefei 230031, Peoples R China
基金
国家重点研发计划;
关键词
Stacked-layer structure; Mass spectrometer chip; Ion trajectory simulation; High-pressure tolerance; Stability diagram; GAS-ANALYSIS; PERFORMANCE; FIELD;
D O I
10.1016/j.cjac.2024.100364
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
Micro-electro-mechanical system (MEMS) based mass spectrometry, MEMS mass spectrometry, emerges a new technique branch of miniature mass spectrometry. One of the main challenges hindering the development of MEMS mass spectrometry is the poor compatibility between the complex electrode structure of traditional ion optics and MEMS processes. This paper proposed a planar quadrupole mass filter utilizing a highly simplified stacked-layer structure and film electrodes, which is easy to fabricate with high precision by MEMS surface micromachining. The ion trajectory simulation model of the MEMS mass spectrometry chip involving all necessary ion optics was elaborated for investigating the preliminary performance of the planar quadrupole mass filter. By optimizing the crucial parameters such as the frequency of radio frequency voltage, the ratio of direct current voltage and radio frequency voltage, pre -pole length, main -pole length, and injection kinetic energy, the preliminary simulation result of planar quadrupole mass filter achieved a 1-2 Da peak width (at 50% of peak height) and 20-80% ion transmission efficiency. Importantly, the planar quadrupole mass filter demonstrated a maximum working pressure of 0.1 Pa (using air as the buffer gas), 100 times higher than that of conventional quadrupole mass filters, while maintaining a 0.7-1.8 Da peak width and similar to 5% ion transmission efficiency. Additionally, to modify the theoretical formulas for the non -ideal quadrupolar field of planar quadrupole mass filter, a novel approach based on stability diagrams obtained by simulation was also introduced. The pretty good linearity ( R2 = 1) between ion masses in the range of 10 to 100 Da and the corresponding radio frequency voltages was verified through the simulated mass spectra. Notably, the proportional coefficient of ion mass and radio frequency voltage exhibited a minimal 4% deviation between the simulated value (4.7132 V0P/Da) and the calculated value (4.9036 V0P/Da) through the modified formula. In conclusion, the preliminary simulation results demonstrated that the planar quadrupole mass filter possessed moderate performance, high -pressure tolerance, and excellent theoretical consistency, which proved it a promising technology of MEMS mass spectrometry. It can find applications in the fields of gas detection, including the rapid response to hazardous gas emissions and in situ detection of dissolved gases in the deep sea.
引用
收藏
页数:10
相关论文
共 50 条
  • [11] STUDY OF A QUADRUPOLE MASS FILTER USING COMPUTER SIMULATION.
    Zikes, Vaclav
    Tesla electronics, 1984, 17 (03): : 86 - 90
  • [12] QUADRUPOLE MASS SPECTROMETER
    MATHIESON, E
    HARRIS, TJ
    AMERICAN JOURNAL OF PHYSICS, 1969, 37 (10) : 1054 - +
  • [13] Simulation of a Quadrupole Mass Filter with Octupole Field
    Bugrov, P. V.
    Sysoev, A. A.
    Konenkov, N. V.
    JOURNAL OF ANALYTICAL CHEMISTRY, 2023, 78 (13) : 1864 - 1867
  • [14] Simulation of a Quadrupole Mass Filter with Octupole Field
    P. V. Bugrov
    A. A. Sysoev
    N. V. Konenkov
    Journal of Analytical Chemistry, 2023, 78 : 1864 - 1867
  • [15] COMPUTER-SIMULATION OF THE QUADRUPOLE MASS FILTER
    TAKEBE, M
    KUMASHIRO, S
    NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 1995, 363 (1-2): : 411 - 415
  • [16] COMPUTER-SIMULATION OF THE QUADRUPOLE MASS FILTER
    CAMPANA, JE
    JURS, PC
    INTERNATIONAL JOURNAL OF MASS SPECTROMETRY, 1980, 33 (02) : 119 - 137
  • [17] Miniature Mass Spectrometer Systems Based on a Microengineered Quadrupole Filter
    Malcolm, Andrew
    Wright, Steven
    Syms, Richard R. A.
    Dash, Neil
    Schwab, Marc-Andre
    Finlay, Alan
    ANALYTICAL CHEMISTRY, 2010, 82 (05) : 1751 - 1758
  • [18] THE QUADRUPOLE MASS-SPECTROMETER - FILTER THEORY AND INSTRUMENT ADJUSTMENTS
    OTT, F
    RITTY, B
    MANQUENOUILLE, R
    VIDE-SCIENCE TECHNIQUE ET APPLICATIONS, 1983, 38 (219): : 479 - 490
  • [19] Silicon micromachined mass filter for a low power, low cost quadrupole mass spectrometer
    Tunstall, JJ
    Taylor, S
    Syms, RRA
    Tate, T
    Ahmad, MM
    MICRO ELECTRO MECHANICAL SYSTEMS - IEEE ELEVENTH ANNUAL INTERNATIONAL WORKSHOP PROCEEDINGS, 1998, : 438 - 442
  • [20] Study of the performance comparison between micro-chamber quadrupole mass spectrometer and miniature quadrupole mass spectrometer
    Xiao, Mei
    Zhang, Xiaobing
    Zhao, Zhiwei
    2009 IEEE INTERNATIONAL VACUUM ELECTRONICS CONFERENCE, 2009, : 257 - 258