Comparison of the dose distribution of the VMAT radiotherapy technique depending on the beam used: FFF-X10MV and FFF-X15MV

被引:1
|
作者
Plaza, Dominika [1 ,2 ]
Sroka, Lukasz [1 ]
Orzechowska, Klaudia [1 ]
Slosarek, Krzysztof [1 ]
机构
[1] Maria Sklodowska Curie Natl Res Inst Oncol, Radiotherapy Planning Dept, Gliwice Branch, Gliwice, Poland
[2] Narodowy Inst Onkol, Panstwowy Inst Badawczy, Zaklad Planowan Radioterapii, Oddzial Gliwicach, Ul Wybrzeze Armii Krajowej 15, PL-44102 Gliwice, Poland
关键词
flattening filter-free beams (FFF); flattening filter beams (FF); volumetric modulated arc therapy (VMAT); radiation planning index (RPI); RADIATION-THERAPY; FLATTENING FILTER; CANCER;
D O I
10.5603/rpor.97508
中图分类号
R73 [肿瘤学];
学科分类号
100214 ;
摘要
Background: The aim of the study was to answer the question of whether flattening filter (FF) and flattening filter-free (FFF) beams can be used alternately in the volumetric modulated arc therapy (VMAT) treatment technique, regardless of the size of the irradiated volume [small (S) or large (L) planning target volume (PTV)].Materials and methods: Two groups of patients were examined: a group with a S-PTV-laryngeal cancer and a group with a L-PTV - gynecological volume. For each patient, two treatment plans were made for beams (energies): FFF-X10MV and FF-X15MV. Then, a statistical analysis, nonparametric test, and independent groups were performed, comparing the beams' impact on the analyzed treatment plans.Results: In the case of laryngeal irradiation (S-PTV), there are no statistically significant differences between the energy used and the assessed parameters of the plan. In the case of gynecological volume (L-PTV), only statistically significant differences were noted for the number of monitor units depending on the energy used. For a large irradiated volume (gynecological case), the use of FFF beams increases the number of monitor units by 39,4% in relation to the FF beam.Conclusions: In the case of gynecological neoplasms, statistically significant differences were found in the number of monitor units. Therefore, in the case of irradiation of L-PTV, it is recommended that flattening-filtering beams are used due to the smaller number of monitors. In the case of S-PTV, no statistically significant differences were found between the types of beams used (FF or FFF) and the treatment plan parameters analyzed in the study.
引用
收藏
页码:654 / 660
页数:7
相关论文
共 44 条
  • [1] Dosimetric Comparison of SBRT VMAT Treatment Plans Generated for 6 MV, 6 MV FFF, and 10 MV FFF Photon Beams
    Wilson, D.
    Wang, B.
    Dunlap, N.
    MEDICAL PHYSICS, 2015, 42 (06) : 3404 - 3404
  • [2] A comparison of the efficiency of delivery between 10 MV FFF VMAT versus conventional 6 MV VMAT for lung SABR and the impact on a busy radiotherapy department
    Moore, K.
    LUNG CANCER, 2014, 83 : S61 - S61
  • [3] Dosimetric comparison of mDCAT and VMAT techniques according to 6MV-FFF and 10MV-FFF energies in patients with single adrenal metastasis
    Durmus, Ismail Faruk
    Okumus, Ayse
    JOURNAL OF X-RAY SCIENCE AND TECHNOLOGY, 2022, 30 (05) : 1023 - 1032
  • [4] Surface Dose and Build-Up Dose Comparison of 6MV and 10MV Flattened with 6MV FFF and 10MV FFF Photon Beams with Different Detectors
    Nambiraj, Arunai
    Sigamani, Ashokkumar
    Kaviarasu, K.
    Rajesh, T.
    Karthikeyan, S.
    Subramanian, Narayanan
    Meerza, Kazem
    MEDICAL PHYSICS, 2015, 42 (06) : 3502 - 3502
  • [5] Technical note: Energy dependence of the new Gafchromic EBT4 film: Dose-response curves for 70 kV, 6 MV, 6 MV FFF, 10 MV FFF, and 15 MV x-ray beams
    Chan, Maria F.
    Park, Jeonghoon
    Aydin, Resat
    Lim, Seng-Boh
    MEDICAL PHYSICS, 2023, 50 (06) : 3738 - 3745
  • [6] 10X-FFF VMAT for Lung SABR: an Investigation of Peripheral Dose
    Mader, J.
    Mestrovic, A.
    MEDICAL PHYSICS, 2014, 41 (08) : 2 - 2
  • [7] Clinical Context of Neutron Dose Contribution From 10MV FFF Vs 15MV Photon Beams
    Vickress, J.
    Zeng, G.
    Sankreacha, R.
    MEDICAL PHYSICS, 2020, 47 (06) : E718 - E718
  • [8] Clinical experience with free-breathing expiration-gated 10MV FFF VMAT stereotactic lung radiotherapy
    de Vries, Isabel F. Remmerts
    Dahele, Max
    Rosario, Tezontl S.
    Slotman, Ben J.
    Verbakel, Wilko F. A. R.
    RADIOTHERAPY AND ONCOLOGY, 2025, 205
  • [9] Can a 10 MV-FFF Photon Beam be Used in Stereotactic Thalamotomy to Reduce Beam Delivery Times: A Dosimetric Evaluation
    Ding, G. X.
    Homann, K. L.
    INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2023, 117 (02): : E659 - E659
  • [10] Synergetic effect of high dose rate radiations (10x FFF/2400 MU/min/10 MV x-rays) and paclitaxel selectively eliminates melanoma cells
    Lodhi, Niraj
    Nagpal, Poonam
    Sarojini, Sreeja
    Keck, Michaela
    Chiu, Yuk Ming
    Parvez, Zeenath
    Adrianzen, Laura
    Suh, K. Stephen
    CANCER REPORTS, 2023, 6 (02)