Amperometric sensor for gaseous H2O2 based on copper redox mediator incorporated electrolyte

被引:2
|
作者
Klun, Ursa [1 ]
Zorko, David [1 ]
Stojanov, Leon [2 ]
Mirceski, Valentin [2 ,3 ]
Jovanovski, Vasko [1 ]
机构
[1] Natl Inst Chem, Dept Mat Chem, Hajdrihova 19, Ljubljana 1000, Slovenia
[2] Ss Cyril & Methodius Univ, Inst Chem, Fac Nat Sci & Math, POB 162, Skopje 1000, North Macedonia
[3] Univ Lodz, Dept Inorgan & Analyt Chem, Tamka 12, PL-91403 Lodz, Poland
关键词
Electrochemical gas sensor; Hydrogen peroxide; Amperometry; Gel electrolyte; Copperion mediator; ELECTROCHEMICAL GAS SENSOR; HYDROGEN-PEROXIDE; POLY(ACRYLIC ACID); CURRENT TRENDS; KINETICS; COMPLEXES; NANOPARTICLES; STABILITY; CONSTANTS; WATER;
D O I
10.1016/j.snr.2023.100144
中图分类号
Q81 [生物工程学(生物技术)]; Q93 [微生物学];
学科分类号
071005 ; 0836 ; 090102 ; 100705 ;
摘要
The determination of gases and other volatile compounds using electrochemical gas sensors remains a challenge with respect to sensitivity and selectivity. Many real-life situations require fast, sensitive, and yet easy-to-operate sensors for gaseous hydrogen peroxide (H2O2) that can operate at room temperature. Herein, we present an H2O2 gas sensor comprising a polyacrylic acid and NaOH gel electrolyte that included Cu(II) ions acting as a redox mediator and commercially available screen-printed carbon electrodes. Aqueous polyacrylate gel electrolyte containing Cu(II) provided for sensing material that enabled the accumulation and stabilization of the gaseous analyte and via redox interaction with Cu(II) its fast and sensitive detection. The designed sensor exhibited good sensitivity in the low mg m- 3 range with a wide linear response in the examined concentration range of 10-100 mg m- 3 and a rapid detection after only 2 min accumulation under ambient conditions. Sensor's relatively simple setup and solid electroanalytical performance have strong application potential in emerging fields such as clinical diagnostics, explosive detection, environmental monitoring and occupational health and safety.
引用
收藏
页数:6
相关论文
共 50 条
  • [21] MWCNT Based Non-Enzymatic H2O2 Sensor: Influence of Amine Functionalization on the Electrochemical H2O2 Sensing
    Revathi, C.
    Rajavel, K.
    Saranya, M.
    Kumar, R. T. Rajendra
    JOURNAL OF THE ELECTROCHEMICAL SOCIETY, 2016, 163 (13) : B627 - B632
  • [22] Electropolymerized film of cobalt tetrabenzimidazolephthalocyanine for the amperometric detection of H2O2
    Mohammed, Imadadulla
    Nemakal, Manjunatha
    Sajjan, Veeresh A.
    Puttappashetty, Dayananda B.
    Sannegowda, Lokesh Koodlur
    JOURNAL OF ELECTROANALYTICAL CHEMISTRY, 2018, 826 : 96 - 103
  • [23] Calorimetric Sensor for H2O2/H2O Mist Streams
    Porkovich, Alexander J.
    Arnold, Matthew D.
    Kouzmina, Galina
    Hingley, Brian
    Cortie, Michael B.
    IEEE SENSORS JOURNAL, 2012, 12 (07) : 2392 - 2398
  • [24] ELECTRODIC REACTION OF H2O2 ON PALLADIUM IN AN ACIDIC ELECTROLYTE
    VITVITSK.GV
    STRAKHOV.VV
    KOZELKOV.NI
    ZHURNAL PRIKLADNOI KHIMII, 1972, 45 (11) : 2429 - 2435
  • [25] LAWS OF THE AUTODISSOLUTION OF COPPER IN THE PRESENCE OF H2O2
    MOLODOV, AI
    MARKOSYAN, GN
    LOSEV, VV
    SOVIET ELECTROCHEMISTRY, 1982, 18 (09): : 1052 - 1059
  • [26] Regulation of redox signaling mediated by H2O2 transporters
    Medrano-Fernandez, Iria
    Sorrentino, Ilaria
    Galli, Mauro
    Bestetti, Stefano
    Sitia, Roberto
    FREE RADICAL BIOLOGY AND MEDICINE, 2021, 177 : S2 - S2
  • [27] Chemiluminescent flow sensor for H2O2 based on the decomposition of H2O2 catalyzed by cobalt(II)-ethanolamine complex immobilized on resin
    Hanaoka, S
    Lin, JM
    Yamada, M
    ANALYTICA CHIMICA ACTA, 2001, 426 (01) : 57 - 64
  • [28] H2O2 signaling by redox modifications of cysteine residues
    Toledano, Michel B.
    TOXICOLOGY LETTERS, 2006, 164 : S5 - S5
  • [29] THP as a sensor for the electrochemical detection of H2O2
    Failla, Mariacristina
    Ferlazzo, Angelo
    Abbate, Vincenzo
    Neri, Giovanni
    Saccullo, Erika
    Gulino, Antonino
    Rescifina, Antonio
    Patamia, Vincenzo
    Floresta, Giuseppe
    BIOORGANIC CHEMISTRY, 2024, 152
  • [30] H2O2 is an important mediator of physiological and pathological healing responses
    Domínguez-Rosales, JA
    Mavi, G
    Levenson, SM
    Rojkind, M
    ARCHIVES OF MEDICAL RESEARCH, 2000, 31 (01) : 15 - 20