Assessment of ultrasound shear wave elastography: An animal ex-vivo study

被引:1
|
作者
Wang, Xiuming [1 ,2 ]
Zhu, Jiaan [1 ]
Liu, Yiqun [1 ]
Li, Wenxue [1 ]
Chen, Si [1 ]
Zhang, Huabin [2 ]
机构
[1] Peking Univ Peoples Hosp, Dept Ultrasound, 11 Xizhimen South St, Beijing 100044, Peoples R China
[2] Tsinghua Univ, Beijing Tsinghua Changgung Hosp, Sch Clin Med, Dept Ultrasound, Beijing, Peoples R China
来源
关键词
lesion size; rectangular sampling box; shear wave elastography; WFUMB GUIDELINES; CLINICAL-USE; SKELETAL-MUSCLE; RECOMMENDATIONS; ELASTICITY; REGION;
D O I
10.1002/acm2.13924
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
ObjectivesTo explore the influence of the surrounding environment of the target tissue, lesion size, and rectangular sampling box size on shear wave speed (SWS). MethodsThe tendon SWS was acquired ex-vivo. Then the tendons were dissected and buried in the couplant (gel) and evaluated by two-dimensional shear wave elastography (2D-SWE). Finally, the tendons were placed in the isolated muscles to simulate the intramuscular lesions, and their elasticity was tested under two rectangular sampling box conditions. The isolated complete liver SWS was acquired. Similarly, the large and small pieces of livers were cut out, placed in the muscles, and assessed by SWE under two rectangular sampling box conditions. The SWS acquired under different conditions was compared. Variability was evaluated using the coefficient of variation (CV). The intraclass correlation coefficient (ICC) was used to evaluate repeatability. ResultsThe SWS of the tendons ex-vivo, buried in the couplant and placed in the isolated muscles showed significant differences (p < 0.001). The ex-vivo condition produced the highest SWS and CV values. There were significant differences in SWS of livers with different sizes placed in muscles (p < 0.001). The highest SWS value was associated with small pieces of livers. No significant difference was found in SWS acquired under different rectangular box sizes (p > 0.05). ConclusionsUnder the present study conditions, the surrounding environment of the target tissue makes a big difference to lesion SWS values. The lesion size will affect the assessment of its inherent elasticity. The size of the sampling frame has no significant effect on the tissue SWS.
引用
收藏
页数:10
相关论文
共 50 条
  • [21] Measurement of shear wave speed dispersion in the placenta by transient elastography: A preliminary ex vivo study
    Simon, Emmanuel G.
    Calle, Samuel
    Perrotin, Franck
    Remenieras, Jean-Pierre
    PLOS ONE, 2018, 13 (04):
  • [22] QUANTITATIVE ASSESSMENT OF KELOIDS USING ULTRASOUND SHEAR WAVE ELASTOGRAPHY
    Guo, Ruiqian
    Xiang, Xi
    Wang, Liyun
    Zhu, Bihui
    Cheng, Shan
    Qiu, Li
    ULTRASOUND IN MEDICINE AND BIOLOGY, 2020, 46 (05): : 1169 - 1178
  • [23] ENDOSCOPIC ULTRASOUND SHEAR WAVE ELASTOGRAPHY FOR LIVER FIBROSIS ASSESSMENT
    Chin, Jerry Yung-Lun
    Vargas, Eric
    Abu Dayyeh, Barham
    Garimella, Vishal
    Storm, Andrew
    Law, Ryan
    Levy, Michael
    Chandrasekhara, Vinay
    GASTROINTESTINAL ENDOSCOPY, 2023, 97 (06) : AB833 - AB834
  • [24] Ex-vivo systems for neuromodulation: A comparison of ex-vivo and in-vivo large animal nerve electrophysiology
    Ribeiro, Mafalda
    Andreis, Felipe R.
    Jabban, Leen
    Nielsen, Thomas G. N. dS.
    Smirnov, Sergey, V
    Lutteroth, Christof
    Proulx, Michael J.
    Rocha, Paulo R. F.
    Metcalfe, Benjamin
    JOURNAL OF NEUROSCIENCE METHODS, 2024, 406
  • [25] Ultrasound shear wave elastography detects bowel wall fibrosis in ex vivo human Crohn's disease specimens
    Higgins, P. D.
    Dillman, J. R.
    Stidham, R. W.
    Johnson, L. A.
    Rubin, J. M.
    JOURNAL OF CROHNS & COLITIS, 2014, 8 : S106 - S106
  • [26] Ultrasound Shear Wave Elastography Detects Bowel Wall Fibrosis in Ex Vivo Human Crohn's Disease Specimens
    Dillman, Jonathan R.
    Stidham, Ryan W.
    Johnson, Laura A.
    Rubin, Jonathan M.
    Higgins, Peter D.
    GASTROENTEROLOGY, 2014, 146 (05) : S152 - S152
  • [27] EVALUATION OF MICROWAVE ABLATION EFFICACY BY STRAIN ELASTOGRAPHY AND SHEAR WAVE ELASTOGRAPHY IN EX VIVO PORCINE LIVER
    Luo, Chunyue
    Li, Tao
    Li, Zuojia
    Zuo, Yanling
    He, Guangmin
    Lin, Jianying
    Liu, Guancheng
    Dai, Lu
    ULTRASOUND IN MEDICINE AND BIOLOGY, 2021, 47 (09): : 2636 - 2645
  • [28] Ultrasound Shear Wave Elastography in Cardiology
    Caenen, Annette
    Bezy, Stephanie
    Pernot, Mathieu
    Nightingale, Kathryn R.
    Vos, Hendrik J.
    Voigt, Jens-Uwe
    Segers, Patrick
    D'hooge, Jan
    JACC-CARDIOVASCULAR IMAGING, 2024, 17 (03) : 314 - 329
  • [29] Comparison of shear wave velocity measurements assessed with two different ultrasound systems in an ex-vivo tendon strain phantom
    Rosskopf, Andrea B.
    Bachmann, Elias
    Snedeker, Jess G.
    Pfirrmann, Christian W. A.
    Buck, Florian M.
    SKELETAL RADIOLOGY, 2016, 45 (11) : 1541 - 1551
  • [30] Comparison of shear wave velocity measurements assessed with two different ultrasound systems in an ex-vivo tendon strain phantom
    Andrea B. Rosskopf
    Elias Bachmann
    Jess G. Snedeker
    Christian W. A. Pfirrmann
    Florian M. Buck
    Skeletal Radiology, 2016, 45 : 1541 - 1551