Measurement of central corneal thickness and precorneal tear film thickness of rabbits using the Scheimpflug system
被引:0
|作者:
Jing Dong
[1
]
Qiang Wu
[2
]
Xiao-Gang Wang
[2
]
机构:
[1] Department of Ophthalmology,the First Hospital of Shanxi Medical University
[2] Department of Ophthalmology,the Sixth People's Hospital Affiliated to Shanghai Jiaotong University
关键词:
tear film;
Scheimpflug;
rabbits;
central corneal thickness;
fluorescein;
D O I:
暂无
中图分类号:
R77 [眼科学];
学科分类号:
100212 ;
摘要:
AIM:To measure central corneal thickness(CCT) and pre-corneal tear film thickness using the Galilei dual Scheimpflug analyzer(GSA) in New Zealand White rabbits.METHODS:Ten normal New Zealand White rabbits(20 eyes) were included in this study.With the assistance of 0.1% fluorescein,the pre-corneal tear film can be well visualized.Both eyes of each rabbit were scanned once with the GSA pre- and post-instillation of 1μL 0.1% fluorescein.The difference between the two measurements of CCT(4-mm diameter) was recorded as the pachymetric values of the central tear film.RESULTS:The CCT of pre- and post-instillation was 388.8±9.5μm and 407.0±10.5μm,respectively.After a paired t-test analysis,the central pre-corneal tear film thickness of 4mm diameter was 18.2±5.31μm with a 95% confidence interval of(15.7,20.6)μm(P<0.001).CONCLUSION:GSA can be used to measure CCT and analyze central tear film thickness of rabbits with the help of fluorescein.
引用
收藏
页码:584 / 587
页数:4