Standard-sample bracketing calibration method combined with Mg as an internal standard for silicon isotopic compositions using multi-collector inductively coupled plasma mass spectrometry

被引:0
|
作者
Honglin Yuan [1 ]
Cheng Cheng [1 ]
Kaiyun Chen [1 ]
Zhian Bao [1 ]
机构
[1] State Key Laboratory of Continental Dynamics, Department of Geology, Northwest University, Collaborative Innovation Center of Continental Tectonics
基金
中国国家自然科学基金;
关键词
Si isotope; Mg internal standard; MC-ICPMS; Rock samples; High resolution;
D O I
暂无
中图分类号
P597 [同位素地质学与地质年代学];
学科分类号
070902 ;
摘要
Silicon isotope analysis traditionally uses a standard-sample bracketing(SSB) method that relies upon greater instrument stability than can be consistently expected. The following proposed method reduces the level of instrumental stability required for the analysis process and provides a valid solution for high-precision and accurate studies of Si isotopic compositions. Rock samples were dissolved by using alkali fusion and acidification.Silicon isotopes were purified with an ion exchange resin.Interfering peaks for isotopes were separated by using a Nu Plasma 1700 multi-collector inductively coupled plasma mass spectrometry(MS) system in high-resolution mode(M/DM [ 8000 RP). Two magnesium isotopes(Mg andMg) and three silicon isotopes(Si,Si, andSi) were analyzed in the same data collection cycle. Mg isotopes were used as an internal standard to calibrate the mass discrimination effects in MS analysis of Si isotopes in combination with the SSB method in order to reduce the effects of MS interference and instrumental mass discrimination on the accuracy of measurements. The conventional SSB method without the Mg internal standard and the proposed SSB method with Mg calibration delivered consistent results within two standard deviations.When Mg was used as an internal standard for calibration,the analysis precision was better than 0.05 % amu.
引用
收藏
页码:421 / 427
页数:7
相关论文
共 50 条
  • [1] Standard-sample bracketing calibration method combined with Mg as an internal standard for silicon isotopic compositions using multi-collector inductively coupled plasma mass spectrometry
    Yuan H.
    Cheng C.
    Chen K.
    Bao Z.
    Acta Geochimica, 2016, 35 (4) : 421 - 427
  • [2] Standard-sample bracketing calibration method combined with Mg as an internal standard for silicon isotopic compositions using multi-collector inductively coupled plasma mass spectrometry
    Honglin Yuan
    Cheng Cheng
    Kaiyun Chen
    Zhian Bao
    Acta Geochimica, 2016, 35 (04) : 421 - 427
  • [3] Determination of Eu isotopic ratio by multi-collector inductively coupled plasma mass spectrometry using a Sm internal standard
    Lee, Seung-Gu
    Tanaka, Tsuyoshi
    SPECTROCHIMICA ACTA PART B-ATOMIC SPECTROSCOPY, 2019, 156 : 42 - 50
  • [4] Magnesium isotopic homogeneity of San Carlos olivine: a potential standard for Mg isotopic analysis by multi-collector inductively coupled plasma mass spectrometry
    Hu, Yan
    Harrington, Melissa D.
    Sun, Yang
    Yang, Zhe
    Konter, Jasper
    Teng, Fang-Zhen
    RAPID COMMUNICATIONS IN MASS SPECTROMETRY, 2016, 30 (19) : 2123 - 2132
  • [5] Analysis of stable isotope ratios of Ba by double-spike standard-sample bracketing using multiple-collector inductively coupled plasma mass spectrometry
    Miyazaki, Takashi
    Kimura, Jun-Ichi
    Chang, Qing
    JOURNAL OF ANALYTICAL ATOMIC SPECTROMETRY, 2014, 29 (03) : 483 - 490
  • [6] In situ Analysis of Mg Isotopic Compositions of Basalt Glasses by Femtosecond Laser Ablation Multi-collector Inductively Coupled Plasma Mass Spectrometry
    Dai Meng-Ning
    Bao Zhi-An
    Chen Kai-Yun
    Yuan Hong-Lin
    CHINESE JOURNAL OF ANALYTICAL CHEMISTRY, 2016, 44 (02) : 173 - 177
  • [7] Measurement of isotopic ratios of uranium by multi-collector inductively coupled plasma mass spectrometry
    Li, Li-Li
    Li, Jin-Ying
    Zhao, Yong-Gang
    Liu, Jun-Ling
    Zhang, Ji-Long
    He-Huaxue yu Fangshe Huaxue/Journal of Nuclear and Radiochemistry, 2006, 28 (03): : 152 - 156
  • [8] Determination of Stable Silicon Isotopes Using Multi-Collector Inductively Coupled Plasma Mass Spectrometry
    Zhang An-Yu
    Zhang Jing
    Zhang Rui-Feng
    Xue Yun
    CHINESE JOURNAL OF ANALYTICAL CHEMISTRY, 2015, 43 (09) : 1353 - 1358
  • [9] Revised exponential model for mass bias correction using an internal standard for isotope abundance ratio measurements by multi-collector inductively coupled plasma mass spectrometry
    Baxter, DC
    Rodushkin, I
    Engström, E
    Malinovsky, D
    JOURNAL OF ANALYTICAL ATOMIC SPECTROMETRY, 2006, 21 (04) : 427 - 430
  • [10] Absolute isotopic composition and atomic weight of selenium using multi-collector inductively coupled plasma mass spectrometry
    Wang, Jun
    Ren, Tongxiang
    Lu, Hai
    Zhou, Tao
    Zhao, Motian
    INTERNATIONAL JOURNAL OF MASS SPECTROMETRY, 2011, 308 (01) : 65 - 70