Fatigue cyclic loading test of an auro-galvanoforming ceramic bridge

被引:0
|
作者
ZHANG XianghaoSUN FengWANG Huan and XU Manqiong Outpatient Dental CenterPeking University School of StomatologyBeijing China Department of Engineering MechanicsTsinghua University Beijing China [100034 ,100084 ]
机构
关键词
D O I
暂无
中图分类号
R78 [口腔科学];
学科分类号
1003 ;
摘要
Background In dental clinics,dentition defects are commonly restored with conventional porcelain-fused-to-metal fixed bridges.However,Ni-Cr alloy ceramic fixed bridges are known to have several drawbacks such as marginal coloration of the neck,low casting precision and,most seriously,poor biocompatibility.These problems could be circumvented by using noble metal ceramic bridges;however,one negative issue related to the conventional noble metal ceramic bridges is its high price due to the use of a large amount of gold for pontic.Therefore,an auro-galvanoforming ceramic bridge would be ideal to retain the advantages of a conventional material,yet reduce the amount of noble metal used.This study aimed to investigate whether any destructive changes occur to the auro-galvanoforming ceramic bridge under a fatigue cyclic loading test. Methods On standard models which the left maxillary first premolar is lost and with the cuspid teeth and the second premolar as the fixed bridge abutment teeth,six maxillary auro-galvanoforming ceramic bridges and six corresponding nichrome ceramic jointed crowns were made as group A,six nichrome ceramic bridges and six corresponding nichrome ceramic jointed crowns were made as group B(control group).And then all specimens were fixed and tested on a fatigue cyclic loading machine;the changes occurring to the surfaces of occlusal contact with large functional area and to the porcelain-metal interfaces of the nichrome primary copings margins were observed through a scanning electron microscope(SEM). Results In 120 hours’ continual observation,none of the specimens had porcelain coating fractures or scraping occurrence and all of the porcelain coatings had been kept intact under sinusoidal cyclic loading with the load range of 120-200 N and frequency of 5 Hz. Conclusions Auro-galvanoforming ceramic bridges exhibited excellent fatigue strength in the fatigue cyclic loading test and may satisfy the clinical demand in theory,while the practical application shall be evaluated by observations in long-term clinical usage.
引用
收藏
页码:1896 / 1899
相关论文
共 50 条
  • [31] CYCLIC FATIGUE OF CERAMIC MATERIALS - INFLUENCE OF CRACK PATH AND FATIGUE MECHANISMS
    HORIBE, S
    HIRAHARA, R
    ACTA METALLURGICA ET MATERIALIA, 1991, 39 (06): : 1309 - 1317
  • [32] Cyclic loading test of self-centering precast segmental unbonded posttensioned UHPFRC bridge columns
    Wang, Jingquan
    Wang, Zhen
    Tang, Yuchuan
    Liu, Tongxu
    Zhang, Jian
    BULLETIN OF EARTHQUAKE ENGINEERING, 2018, 16 (11) : 5227 - 5255
  • [33] Cyclic loading test of self-centering precast segmental unbonded posttensioned UHPFRC bridge columns
    Jingquan Wang
    Zhen Wang
    Yuchuan Tang
    Tongxu Liu
    Jian Zhang
    Bulletin of Earthquake Engineering, 2018, 16 : 5227 - 5255
  • [34] Stability analysis of bridge piers subjected to cyclic loading
    Yamaguchi, E
    Goto, Y
    Abe, K
    Hayashi, M
    Kubo, Y
    LIGHT-WEIGHT STEEL AND ALUMINIUM STRUCTURES, 1999, : 829 - 836
  • [35] Fracture load of ceramic restorations after fatigue loading
    Baladhandayutham, Balasudha
    Lawson, Nathaniel C.
    Burgess, John O.
    JOURNAL OF PROSTHETIC DENTISTRY, 2015, 114 (02): : 266 - 271
  • [36] Failure Characteristics of PZT Ceramic During Cyclic Loading
    Mitsuhiro Okayasu
    Tsukasa Ogawa
    Journal of Electronic Materials, 2020, 49 : 5534 - 5541
  • [37] Fatigue of a porcelain ceramic under repeated impact loading
    Maity, S
    Sarkar, BK
    TRANSACTIONS OF THE INDIAN INSTITUTE OF METALS, 1996, 49 (04): : 601 - 604
  • [38] Cyclic plastic response and fatigue life in symmetric and asymmetric cyclic loading
    Polak, J.
    Petrenec, M.
    Kruml, T.
    Chlupova, A.
    11TH INTERNATIONAL CONFERENCE ON THE MECHANICAL BEHAVIOR OF MATERIALS (ICM11), 2011, 10
  • [39] Failure Characteristics of PZT Ceramic During Cyclic Loading
    Okayasu, Mitsuhiro
    Ogawa, Tsukasa
    JOURNAL OF ELECTRONIC MATERIALS, 2020, 49 (09) : 5534 - 5541
  • [40] LOADING OF BRIDGE ABUTMENT TEETH (INVITRO TEST)
    KORBER, E
    LINDEMANN, W
    DEUTSCHE ZAHNARZTLICHE ZEITSCHRIFT, 1982, 37 (03): : 305 - 308