摘要
PurposeTo computationally characterize the LET distribution during dynamic collimation in PBS and quantify its impact on the resultant dose distribution.MethodsMonte Carlo simulations using Geant4 were used to model the production of low-energy proton scatter produced in the collimating components of a novel PBS collimator. Custom spectral tallies were created to quantify the energy, track- and dose-averaged LET resulting from individual beamlet and composite fields simulated from a model of the IBA dedicated nozzle system. The composite dose distributions were optimized to achieve a uniform physical dose coverage of a cubical and pyramidal target, and the resulting dose-average LET distributions were calculated for uncollimated and collimated PBS deliveries and used to generate RBE-weighted dose distributions.ResultsFor collimated beamlets, the scattered proton energy fluence is strongly dependent on collimator position relative to the central axis of the beamlet. When delivering a uniform profile, the distribution of dose-average LET was nearly identical within the target and increased between 1 and 2keV/mu m$2 \,{\rm keV}/\mathrm{\umu }\mathrm{m}$ within 10 mm surrounding the target. Dynamic collimation resulted in larger dose-average LET changes: increasing the dose-average LET between 1 and 3keV/mu m$3 \,{\rm keV}/\mathrm{\umu }\mathrm{m}$ within 10 mm of a pyramidal target while reducing the dose-average LET outside this margin by as much as 10keV/mu m$10 \,{\rm keV}/\mathrm{\umu }\mathrm{m}$. Biological dose distributions are improved with energy-specific collimation in reducing the lateral penumbra.ConclusionThe presence of energy-specific collimation in PBS can lead to dose-average LET changes relative to an uncollimated delivery. In some clinical situations, the placement and application of energy-specific collimation may require additional planning considerations based on its reduction to the lateral penumbra and increase in high-dose conformity. Future applications may embody these unique dosimetric characteristics to redirect high-LET portions of a collimated proton beamlet from healthy tissues while enhancing the dose-average LET distribution within target.
引用
收藏
页数:11
相关论文
共 50 条
  • [41] Monte Carlo Dose Verification of Pencil Beam Scanning Proton Therapy
    Giantsoudi, D.
    Clasie, B.
    Grassberger, C.
    Dowdell, S.
    Depauw, N.
    Madden, T.
    Kooy, H.
    Paganetti, H.
    MEDICAL PHYSICS, 2013, 40 (06)
  • [42] Initial Experience of Patient-Specific QA Using a Pencil Beam Scanning Proton Therapy System
    Piskulich, F.
    Zhang, Y.
    Perles, L.
    Mascia, A.
    Lepage, R.
    Giebeler, A.
    Dong, L.
    MEDICAL PHYSICS, 2014, 41 (06) : 384 - 384
  • [43] Pencil Beam Dose Algorithm for Intensity Modulated Proton Therapy with Scanning Proton Beams
    Wang, H.
    Zheng, H.
    Cao, R.
    Jia, J.
    Song, J.
    He, T.
    Wu, Y.
    MEDICAL PHYSICS, 2017, 44 (06)
  • [44] Characterizing the Entrance Dose Induced by a Dynamic Collimation System for Pencil Beam Scanning Proton Therapy Using a Commercial Multilayer Ionization Chamber
    Nelson, N.
    Culberson, W.
    Smith, B.
    Flynn, R.
    Hyer, D.
    Hill, P.
    MEDICAL PHYSICS, 2021, 48 (06)
  • [45] Investigating beam matching for multi-room pencil beam scanning proton therapy
    Rana, Suresh
    Bennouna, Jaafar
    PHYSICAL AND ENGINEERING SCIENCES IN MEDICINE, 2020, 43 (04) : 1241 - 1251
  • [46] Secondary Neutron Dose From a Dynamic Collimation System During Intracranial Pencil Beam Scanning Proton Therapy: A Monte Carlo Investigation
    Smith, Blake R.
    Hyer, Daniel E.
    Hill, Patrick M.
    Culberson, Wesley S.
    INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2019, 103 (01): : 241 - 250
  • [47] Investigating beam matching for multi-room pencil beam scanning proton therapy
    Suresh Rana
    Jaafar Bennouna
    Physical and Engineering Sciences in Medicine, 2020, 43 : 1241 - 1251
  • [48] Pencil beam scanning proton therapy for the treatment of craniopharyngioma complicated with radiation-induced cerebral vasculopathies: A dosimetric and linear energy transfer (LET) evaluation
    Bolsi, Alessandra
    Placidi, Lorenzo
    Pica, Alessia
    Ahlhelm, Frank J.
    Walser, Marc
    Lomax, Antony J.
    Weber, Damien C.
    RADIOTHERAPY AND ONCOLOGY, 2020, 149 : 197 - 204
  • [49] Toxicity outcomes of hypofractionated pencil beam scanning proton beam therapy for spinal Chordomas
    Burela, N.
    Chilukuri, S.
    Nangia, S.
    Patro, K.
    Mp, N.
    Gaikwad, U.
    Sundar, S.
    Thimma, R.
    Rajendran, A.
    Sulaiman, A.
    Sharma, D.
    Panda, P. K.
    Jalali, R.
    RADIOTHERAPY AND ONCOLOGY, 2021, 161 : S1166 - S1166
  • [50] Applications of a novel detector for pencil beam scanning proton therapy beam quality assurance
    Yang, Pei-Ying
    Hsieh, Yang-Wei
    Kang, Chen-Lin
    Tseng, Chin-Dar
    Lin, Chih-Hsueh
    Shieh, Chin-Shiuh
    Chao, Pei-Ju
    Lee, Tsair-Fwu
    INTERNATIONAL JOURNAL OF MODERN PHYSICS B, 2021, 35 (14N16):