Nonocclusive Ablation Technique Using a Novel Cryoballoon for Failed Left Superior Pulmonary Vein Isolation Despite Complete Occlusion

被引:0
|
作者
Murata, Kazuya [1 ,2 ]
Yamauchi, Yasuteru [1 ]
Yasui, Yumi [1 ]
Oda, Atsuhito [1 ]
Arai, Hirofumi [1 ]
Sagawa, Yuichiro [1 ]
Arima, Hideki [1 ]
Kurabayashi, Manabu [1 ]
Miyazaki, Shinsuke [2 ]
Sasano, Tetsuo [2 ]
机构
[1] Japan Red Cross Yokohama City Bay Hosp, Dept Cardiol, Yokohama, Kanagawa, Japan
[2] Inst Sci Tokyo, Dept Cardiovasc Med, Bunkyo Ku, Tokyo, Japan
关键词
atrial fibrillation; cryoballoon ablation; left superior pulmonary vein isolation; nonocclusive ablation; novel cryoballoon; PAROXYSMAL ATRIAL-FIBRILLATION; RADIOFREQUENCY ABLATION; PREDICTORS;
D O I
10.1111/pace.15136
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
Background: An indicator of successful cryoballoon (CB)-assisted pulmonary vein (PV) isolation is complete PV occlusion. However, CBs may exhibit a weaker freezing effect on the equatorial plane. This study investigates the predictors of failed left superior PV (LSPV) isolation despite complete occlusion with novel CBs. Methods: This retrospective analysis enrolled 300 consecutive patients who underwent first-time ablation with POLARx or POLARxFIT between November 2021 and October 2023. Of the total, complete occlusion of the LSPV was achieved in 200 patients. Patients in whom LSPV isolation was achieved with additional nonocclusive freezing of the LSPV roof due to nonisolation of LSPV despite complete occlusion (Group A) were compared with those in whom isolation was achieved with complete PV occlusion alone (Group B). Results: Group A had a larger LSPV diameter (21.5 +/- 4.6 mm vs. 18.8 +/- 3.3 mm, p = 0.052), larger left atrial volume on CT (142.3 +/- 47.8 cc vs. 117.8 +/- 39.0 cc, p = 0.028), higher nadir temperature (-54.1 +/- 5.1 degrees C vs. -60.2 +/- 4.4 degrees C, p < 0.001), and smaller northern latitude of the balloon contact site on the LSPV roof side (20.9 degrees +/- 3.8 degrees vs. 38.9 degrees +/- 6.7 degrees, p < 0.001) compared with Group B. A 27.5 degrees north latitude was observed in most of Group A (sensitivity, 100%; specificity, 96%). Conclusions: Adequate contact positioning of the northern hemisphere to the LSPV is critical for effective isolation, particularly when isolation is challenging despite complete occlusion. In such cases, nonocclusive cryoablation against the LSPV roof might be effective.
引用
收藏
页数:10
相关论文
共 50 条
  • [41] Anatomical predictors to improve the outcome of cryoballoon ablation for pulmonary vein isolation using the 28 mm balloon
    Knecht, S.
    Kuehne, M.
    Ammann, P.
    Altmann, D.
    Osswald, S.
    Sticherling, C.
    EUROPEAN HEART JOURNAL, 2012, 33 : 215 - 216
  • [42] Efficacy and safety of pulmonary vein isolation with pulsed field ablation vs. novel cryoballoon ablation system for atrial fibrillation
    Badertscher, Patrick
    Weidlich, Simon
    Knecht, Sven
    Stauffer, Niklas
    Krisai, Philipp
    Voellmin, Gian
    Osswald, Stefan
    Sticherling, Christian
    Kuhne, Michael
    EUROPACE, 2023, 25 (12):
  • [43] Pulmonary vein isolation using cryoballoon ablation versus RF ablation using ablation index following the CLOSE protocol: A prospective randomized trial
    Theis, Cathrin
    Kaiser, Bastian
    Kaesemann, Philipp
    Hui, Felix
    Pirozzolo, Giancarlo
    Bekeredjian, Raffi
    Huber, Carola
    JOURNAL OF CARDIOVASCULAR ELECTROPHYSIOLOGY, 2022, 33 (05) : 866 - 873
  • [44] Right ventricular rapid pacing in catheter ablation of atrial fibrillation: a novel application for cryoballoon pulmonary vein isolation
    K. R. Julian Chun
    Alexander Fürnkranz
    Boris Schmidt
    Andreas Metzner
    Roland Tilz
    Thomas Zerm
    Ilka Köster
    Buelent Koektuerk
    Melanie Konstantinidou
    Feifan Ouyang
    Karl Heinz Kuck
    Clinical Research in Cardiology, 2009, 98 : 493 - 500
  • [45] Novel spiral mapping catheter facilitates observation of the time-to-pulmonary vein isolation during cryoballoon ablation
    Pott, Alexander
    Petscher, Kerstin
    Baumhardt, Michael
    Stephan, Tilman
    Rattka, Manuel
    Paliskyte, Rima
    Bothner, Carlo
    Kessler, Mirjam
    Rottbauer, Wolfgang
    Dahme, Tillman
    HEART AND VESSELS, 2019, 34 (03) : 496 - 502
  • [46] The impact of left atrium size on selection of the pulmonary vein isolation method for atrial fibrillation: Cryoballoon or radiofrequency catheter ablation
    Ikenouchi, Takashi
    Inaba, Osamu
    Takamiya, Tomomasa
    Inamura, Yukihiro
    Sato, Akira
    Matsumura, Yutaka
    Sato, Hiroyuki
    Hirakawa, Akihiro
    Takahashi, Yoshihide
    Goya, Masahiko
    Sasano, Tetsuo
    Nitta, Junichi
    AMERICAN HEART JOURNAL, 2021, 231 : 82 - 92
  • [47] Novel spiral mapping catheter facilitates observation of the time-to-pulmonary vein isolation during cryoballoon ablation
    Alexander Pott
    Kerstin Petscher
    Michael Baumhardt
    Tilman Stephan
    Manuel Rattka
    Rima Paliskyte
    Carlo Bothner
    Mirjam Keßler
    Wolfgang Rottbauer
    Tillman Dahme
    Heart and Vessels, 2019, 34 : 496 - 502
  • [48] Right ventricular rapid pacing in catheter ablation of atrial fibrillation: a novel application for cryoballoon pulmonary vein isolation
    Chun, K. R. Julian
    Fuernkranz, Alexander
    Schmidt, Boris
    Metzner, Andreas
    Tilz, Roland
    Zerm, Thomas
    Koester, Ilka
    Koektuerk, Buelent
    Konstantinidou, Melanie
    Ouyang, Feifan
    Kuck, Karl Heinz
    CLINICAL RESEARCH IN CARDIOLOGY, 2009, 98 (08) : 493 - 500
  • [49] Novel cryoballoon ablation system for pulmonary vein isolation: multicenter assessment of efficacy and safety-ANTARCTICA study
    Heeger, Christian H.
    Pott, Alexander
    Sohns, Christian
    Riesinger, Lisa
    Sommer, Philipp
    Gasperetti, Alessio
    Tondo, Claudio
    Fassini, Gaetano
    Moser, Fabian
    Lucas, Philipp
    Weinmann, Karolina
    Bohnen, Jan Eric
    Dahme, Tillman
    Rillig, Andreas
    Kuck, Karl Heinz
    Wakili, Reza
    Metzner, Andreas
    Tilz, Roland R.
    EUROPACE, 2022, 24 (12): : 1917 - 1925
  • [50] Novel or established cryoballoon ablation system for pulmonary vein isolation: the prospective ICE-AGE-1 study
    Heeger, Christian-Hendrik
    Popescu, Sorin Stefan
    Inderhees, Tim
    Nussbickel, Noemi
    Eitel, Charlotte
    Kirstein, Bettina
    Phan, Huong-Lan
    Hatahet, Sascha
    Subin, Behnam
    Traub, Anna
    Grosse, Niels
    Kuck, Karl-Heinz
    Vogler, Julia
    Tilz, Roland R.
    EUROPACE, 2023, 25 (09):