Nonocclusive Ablation Technique Using a Novel Cryoballoon for Failed Left Superior Pulmonary Vein Isolation Despite Complete Occlusion

被引:0
|
作者
Murata, Kazuya [1 ,2 ]
Yamauchi, Yasuteru [1 ]
Yasui, Yumi [1 ]
Oda, Atsuhito [1 ]
Arai, Hirofumi [1 ]
Sagawa, Yuichiro [1 ]
Arima, Hideki [1 ]
Kurabayashi, Manabu [1 ]
Miyazaki, Shinsuke [2 ]
Sasano, Tetsuo [2 ]
机构
[1] Japan Red Cross Yokohama City Bay Hosp, Dept Cardiol, Yokohama, Kanagawa, Japan
[2] Inst Sci Tokyo, Dept Cardiovasc Med, Bunkyo Ku, Tokyo, Japan
关键词
atrial fibrillation; cryoballoon ablation; left superior pulmonary vein isolation; nonocclusive ablation; novel cryoballoon; PAROXYSMAL ATRIAL-FIBRILLATION; RADIOFREQUENCY ABLATION; PREDICTORS;
D O I
10.1111/pace.15136
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
Background: An indicator of successful cryoballoon (CB)-assisted pulmonary vein (PV) isolation is complete PV occlusion. However, CBs may exhibit a weaker freezing effect on the equatorial plane. This study investigates the predictors of failed left superior PV (LSPV) isolation despite complete occlusion with novel CBs. Methods: This retrospective analysis enrolled 300 consecutive patients who underwent first-time ablation with POLARx or POLARxFIT between November 2021 and October 2023. Of the total, complete occlusion of the LSPV was achieved in 200 patients. Patients in whom LSPV isolation was achieved with additional nonocclusive freezing of the LSPV roof due to nonisolation of LSPV despite complete occlusion (Group A) were compared with those in whom isolation was achieved with complete PV occlusion alone (Group B). Results: Group A had a larger LSPV diameter (21.5 +/- 4.6 mm vs. 18.8 +/- 3.3 mm, p = 0.052), larger left atrial volume on CT (142.3 +/- 47.8 cc vs. 117.8 +/- 39.0 cc, p = 0.028), higher nadir temperature (-54.1 +/- 5.1 degrees C vs. -60.2 +/- 4.4 degrees C, p < 0.001), and smaller northern latitude of the balloon contact site on the LSPV roof side (20.9 degrees +/- 3.8 degrees vs. 38.9 degrees +/- 6.7 degrees, p < 0.001) compared with Group B. A 27.5 degrees north latitude was observed in most of Group A (sensitivity, 100%; specificity, 96%). Conclusions: Adequate contact positioning of the northern hemisphere to the LSPV is critical for effective isolation, particularly when isolation is challenging despite complete occlusion. In such cases, nonocclusive cryoablation against the LSPV roof might be effective.
引用
收藏
页数:10
相关论文
共 50 条
  • [21] EFFICACY OF PULMONARY VEIN ISOLATION WITH A SINGLE APPLICATION USING A NOVEL CRYOBALLOON CATHETER
    Andrade, J.
    Dubuc, M.
    Guerra, P.
    Coulombe, N.
    Landry, E.
    Rivard, L.
    Macle, L.
    Thibault, B.
    Talajic, M.
    Roy, D.
    Khairy, P.
    CANADIAN JOURNAL OF CARDIOLOGY, 2012, 28 (05) : S411 - S411
  • [22] Performance of two tools for pulmonary vein occlusion assessment with a novel navigation system in cryoballoon ablation procedure
    Qiao, Yu
    Zhao, Zhen
    Fu, Mingpeng
    Cai, Xiang
    Guo, Yulong
    Liu, Ke
    Guo, Jinrui
    Guo, Tao
    Niu, Guodong
    JOURNAL OF CARDIOVASCULAR ELECTROPHYSIOLOGY, 2023, 34 (01) : 16 - 23
  • [23] Catheter ablation of atrial fibrillation: radiofrequency catheter ablation for redo procedures after pulmonary vein isolation with the cryoballoon technique
    Kettering, K.
    Mollnau, H.
    Gramley, F.
    Weiss, C.
    EUROPEAN HEART JOURNAL, 2009, 30 : 272 - 272
  • [24] Effectiveness of Simple Pulmonary Vein Isolation using Cryoballoon Ablation in Patients with Persistent Atrial Fibrillation. -A Comparison with Pulmonary Vein Antrum Isolation using Radiofrequency Catheter Ablation
    Tokano, Takashi
    Shiozawa, Tomoyuki
    Kimura, Yuki
    Odagiri, Fuminori
    Tabuchi, Haruna
    Hayuashi, Hidemori
    Sekita, Gaku
    Sumiyoshi, Masataka
    Nakazato, Yuji
    JOURNAL OF CARDIOVASCULAR ELECTROPHYSIOLOGY, 2020, 31 (03) : 789 - 789
  • [25] Pulmonary vein isolation during cryoballoon ablation using the novel Achieve inner lumen mapping catheter: a feasibility study
    Chierchia, Gian-Battista
    de Asmundis, Carlo
    Namdar, Mehdi
    Westra, Sjoerd
    Kuniss, Malte
    Sarkozy, Andrea
    Bayrak, Fatih
    Ricciardi, Danilo
    Casado-Arroyo, Ruben
    Manero, Moises Rodriguez
    Rao, Jayakeerthi Y.
    Smeets, Joep
    Brugada, Pedro
    EUROPACE, 2012, 14 (07): : 962 - 967
  • [26] Pulmonary Vein Isolation Using the First-Generation Cryoballoon Technique in Chinese Patients
    Zhou, Gong-Bu
    Guo, Xiao-Gang
    Liu, Xu
    Yang, Jian-Du
    Sun, Qi
    Ma, Jian
    Zhang, Shu
    PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY, 2015, 38 (09): : 1073 - 1081
  • [27] Accuracy and acute efficacy of a novel occlusion tool to guide cryoballoon-based pulmonary vein isolation
    Rottner, L.
    Moser, F.
    Schleberger, R.
    Weimann, J.
    Moser, J.
    Lemoine, M.
    Muenkler, P.
    Dinshaw, L.
    Risius, T.
    Kirchhof, P.
    Ouyang, F.
    Reissmann, B.
    Metzner, A.
    Rillig, A.
    EUROPEAN HEART JOURNAL, 2021, 42 : 354 - 354
  • [28] Sutureless technique for left pulmonary vein occlusion with persistent left superior vena cava: Case report
    Tamura, Kazuki
    Hoshino, Joji
    Ezure, Masahiko
    Hasegawa, Yutaka
    Yamada, Yasuyuki
    Okada, Shuichi
    Morishita, Hiroyuki
    Seki, Masahiro
    Soda, Takashi
    JTCVS TECHNIQUES, 2024, 27 : 119 - 121
  • [29] Troponin release after pulmonary vein isolation using pulsed field ablation compared to radiofrequency and cryoballoon ablation
    Krisai, Philipp
    Knecht, Sven
    Badertscher, Patrick
    Muhl, Aline
    Osswald, Stefan
    Roten, Laurent
    Reichlin, Tobias
    Sticherling, Christian
    Kuhne, Michael
    HEART RHYTHM, 2022, 19 (09) : 1471 - 1472
  • [30] Durability of a right superior pulmonary vein isolation after an inevitably interrupted single short freeze during cryoballoon ablation
    Miyazaki, Shinsuke
    Hasegawa, Kanae
    Iesaka, Yoshito
    JOURNAL OF CARDIOVASCULAR ELECTROPHYSIOLOGY, 2021, 32 (09) : 2418 - 2423