Nonocclusive Ablation Technique Using a Novel Cryoballoon for Failed Left Superior Pulmonary Vein Isolation Despite Complete Occlusion

被引:0
|
作者
Murata, Kazuya [1 ,2 ]
Yamauchi, Yasuteru [1 ]
Yasui, Yumi [1 ]
Oda, Atsuhito [1 ]
Arai, Hirofumi [1 ]
Sagawa, Yuichiro [1 ]
Arima, Hideki [1 ]
Kurabayashi, Manabu [1 ]
Miyazaki, Shinsuke [2 ]
Sasano, Tetsuo [2 ]
机构
[1] Japan Red Cross Yokohama City Bay Hosp, Dept Cardiol, Yokohama, Kanagawa, Japan
[2] Inst Sci Tokyo, Dept Cardiovasc Med, Bunkyo Ku, Tokyo, Japan
关键词
atrial fibrillation; cryoballoon ablation; left superior pulmonary vein isolation; nonocclusive ablation; novel cryoballoon; PAROXYSMAL ATRIAL-FIBRILLATION; RADIOFREQUENCY ABLATION; PREDICTORS;
D O I
10.1111/pace.15136
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
Background: An indicator of successful cryoballoon (CB)-assisted pulmonary vein (PV) isolation is complete PV occlusion. However, CBs may exhibit a weaker freezing effect on the equatorial plane. This study investigates the predictors of failed left superior PV (LSPV) isolation despite complete occlusion with novel CBs. Methods: This retrospective analysis enrolled 300 consecutive patients who underwent first-time ablation with POLARx or POLARxFIT between November 2021 and October 2023. Of the total, complete occlusion of the LSPV was achieved in 200 patients. Patients in whom LSPV isolation was achieved with additional nonocclusive freezing of the LSPV roof due to nonisolation of LSPV despite complete occlusion (Group A) were compared with those in whom isolation was achieved with complete PV occlusion alone (Group B). Results: Group A had a larger LSPV diameter (21.5 +/- 4.6 mm vs. 18.8 +/- 3.3 mm, p = 0.052), larger left atrial volume on CT (142.3 +/- 47.8 cc vs. 117.8 +/- 39.0 cc, p = 0.028), higher nadir temperature (-54.1 +/- 5.1 degrees C vs. -60.2 +/- 4.4 degrees C, p < 0.001), and smaller northern latitude of the balloon contact site on the LSPV roof side (20.9 degrees +/- 3.8 degrees vs. 38.9 degrees +/- 6.7 degrees, p < 0.001) compared with Group B. A 27.5 degrees north latitude was observed in most of Group A (sensitivity, 100%; specificity, 96%). Conclusions: Adequate contact positioning of the northern hemisphere to the LSPV is critical for effective isolation, particularly when isolation is challenging despite complete occlusion. In such cases, nonocclusive cryoablation against the LSPV roof might be effective.
引用
收藏
页数:10
相关论文
共 50 条
  • [1] Efficacy of nonocclusive separate freezing technique of cryoballoon pulmonary vein isolation: comparison of novel and conventional cryoballoon system
    Arai, H.
    Oda, A.
    Murata, K.
    Sagawa, Y.
    Goya, M.
    Sasano, T.
    Yamauchi, Y.
    EUROPEAN HEART JOURNAL, 2023, 44
  • [2] Permanence of Pulmonary Vein Isolation Using Cryoballoon Observed in Repeat Left Atrial Ablation
    Chiu, Albert
    Al-Zubaidi, Muhanad
    Su, Wilber
    CIRCULATION, 2014, 130
  • [3] Left Phrenic Nerve Injury during Cryoballoon Ablation of the Left Superior Pulmonary Vein
    Andrie, Rene P.
    Schrickel, Jan W.
    Nickenig, Georg
    Lickfett, Lars
    PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY, 2012, 35 (11): : e334 - e336
  • [4] Pulmonary vein isolation using an occluding cryoballoon for circumferential ablation
    Van Belle, Y.
    Thornton, A. S.
    Rivero-Ayerza, M.
    Jansse, P.
    Jordaens, L.
    EUROPEAN HEART JOURNAL, 2006, 27 : 42 - 43
  • [5] Cryoballoon Pulmonary Vein Isolation Guided by Transesophageal Echocardiography: Novel Aspects on an Emerging Ablation Technique
    Siklody, Claudia Herrera
    Minners, Jan
    Allgeier, Martin
    Allgeier, Hans-Juergen
    Jander, Nikolaus
    Weber, Reinhold
    Schiebeling-Roemer, Jochen
    Neumann, Franz-Josef
    Kalusche, Dietrich
    Arentz, Thomas
    JOURNAL OF CARDIOVASCULAR ELECTROPHYSIOLOGY, 2009, 20 (11) : 1197 - 1202
  • [6] Beware of superior vena cava isolation during cryoballoon ablation of the right superior pulmonary vein
    Bergonti, Marco
    Heidbuchel, Hein
    Sarkozy, Andrea
    HEART RHYTHM O2, 2022, 3 (03): : 323 - 324
  • [7] Bailout cryoballoon ablation after failed pulmonary vein isolation with radiofrequency ablation: Acute success and safety
    Contractor, Tahmeed
    Bhardwaj, Rahul
    Mandapati, Ravi
    JOURNAL OF CARDIOVASCULAR ELECTROPHYSIOLOGY, 2019, 30 (09) : 1751 - 1751
  • [8] Comparison between superior vena cava ablation in addition to pulmonary vein isolation and standard pulmonary vein isolation in patients with paroxysmal atrial fibrillation with the cryoballoon technique
    Ingrid Overeinder
    Thiago Guimarães Osório
    Paul-Adrian Călburean
    Antonio Bisignani
    Gezim Bala
    Juan Sieira
    Erwin Ströker
    Maysam Al Houssari
    Joerelle Mojica
    Serge Boveda
    Gaetano Paparella
    Pedro Brugada
    Carlo de Asmundis
    Gian-Battista Chierchia
    Journal of Interventional Cardiac Electrophysiology, 2021, 62 : 579 - 586
  • [9] Comparison between superior vena cava ablation in addition to pulmonary vein isolation and standard pulmonary vein isolation in patients with paroxysmal atrial fibrillation with the cryoballoon technique
    Overeinder, Ingrid
    Osorio, Thiago Guimaraes
    Calburean, Paul-Adrian
    Bisignani, Antonio
    Bala, Gezim
    Sieira, Juan
    Stroker, Erwin
    Al Houssari, Maysam
    Mojica, Joerelle
    Boveda, Serge
    Paparella, Gaetano
    Brugada, Pedro
    de Asmundis, Carlo
    Chierchia, Gian-Battista
    JOURNAL OF INTERVENTIONAL CARDIAC ELECTROPHYSIOLOGY, 2021, 62 (03) : 579 - 586
  • [10] Novel occlusion tool software for pulmonary vein occlusion verification in atrial fibrillation cryoballoon ablation
    Solimene, F.
    Cauti, F. M.
    Stabile, G.
    Rossi, P.
    Schillaci, V
    Polselli, M.
    Arestia, A.
    Iaia, L.
    Shopova, G.
    Bianchi, S.
    EUROPEAN HEART JOURNAL, 2020, 41 : 435 - 435