Eye lens dosimetry: does the direction of rotation (vertical or horizontal) play a role in type testing?

被引:0
|
作者
Khanbabaee, Behnam [1 ]
Behrens, Rolf [1 ]
Erdem, Onur [1 ]
机构
[1] Phys Tech Bundesanstalt, Bundesallee 100, D-38116 Braunschweig, Germany
关键词
H-p(3) dosemeters; type testing; cylinder phantom; eye lens dosimetry; H-P(3); RADIONUCLIDES; CYLINDER; PHANTOM;
D O I
10.1088/1361-6498/adaaff
中图分类号
X [环境科学、安全科学];
学科分类号
08 ; 0830 ;
摘要
With the International Commission on Radiological Protection (ICRP) lowering the annual dose limit for the eye lens to 20 mSv, precise monitoring of eye lens exposure has become essential. The personal dose equivalent at a depth of 3 mm, Hp(3), is the measurement method for monitoring the dose to the lens of the eye. Usual dosemeter type-test irradiations at non-normal angles of radiation incidence (alpha not equal 0 degrees) primarily use lateral radiation exposure scenarios, where radiation approaches from the left or right, necessitating rotation of the dosemeter-phantom setup around a vertical axis. However, this method does not adequately account for bottom-to-top radiation exposures which are common in real-world situations (such as radiation scattered by a patient reaching medical staff). This study examines oblique radiation exposure conditions using a typical eye lens thermoluminescent dosemeter (TLD), Eye-D, placed on a cylindrical phantom to assess dose response at different angles and exposure energies. The study employs both low-energy (N-30 radiation quality with a mean photon energy of 25 keV) and medium-energy (N-100 radiation quality with a mean photon energy of 83 keV) x-rays at irradiation angles of -60 degrees, 0 degrees, and +60 degrees, measured along the vertical and horizontal rotation axes of the dosemeter-phantom setup. The results show no significant difference between horizontal and vertical (polar and radial) rotation orientations of the dosemeter-phantom setup: recorded relative doses stayed well within +/- 1 %, i.e. by far within the attributed combined uncertainty of +/- 2 %.
引用
收藏
页数:6
相关论文
共 50 条
  • [21] Type testing of a head band dosemeter for measuring eye lens dose in terms of Hp(3)
    Gilvin, P. J.
    Baker, S. T.
    Gibbens, N. J.
    Roberts, G. H.
    Tanner, R. J.
    Eakins, J. S.
    Hager, L. G.
    Daniels, T. J.
    RADIATION PROTECTION DOSIMETRY, 2013, 157 (03) : 430 - 436
  • [22] Does the LDL Receptor Play a Role in the Risk of Developing Type 2 Diabetes?
    Preiss, David
    Sattar, Naveed
    JAMA-JOURNAL OF THE AMERICAN MEDICAL ASSOCIATION, 2015, 313 (10): : 1016 - 1017
  • [23] Preferences for different questions when testing hypotheses in an abstract task: Positivity does play a role, asymmetry does not
    Cherubini, Paolo
    Rusconi, Patrice
    Russo, Selena
    Di Bari, Selenia
    Sacchi, Simona
    ACTA PSYCHOLOGICA, 2010, 134 (02) : 162 - 174
  • [24] Does ferritin secreted by canine lens epithelial cells play a role in maintaining cellular iron homeostasis?
    Goralska, M
    McGahan, MC
    INVESTIGATIVE OPHTHALMOLOGY & VISUAL SCIENCE, 2003, 44 : U324 - U324
  • [25] If Eyes are the Window to Our Soul, What Role does Eye-Tracking Play in Accounting Research?
    Lynch, Edward J.
    Andiola, Lindsay M.
    BEHAVIORAL RESEARCH IN ACCOUNTING, 2019, 31 (02) : 107 - 133
  • [26] Does endogenous serum oestrogen play a role in meibomian gland dysfunction in postmenopausal women with dry eye?
    Golebiowski, Blanka
    Badarudin, Noor
    Eden, John
    You, Jingjing
    Hampel, Ulrike
    Stapleton, Fiona
    BRITISH JOURNAL OF OPHTHALMOLOGY, 2017, 101 (02) : 218 - 222
  • [27] DISPARITIES IN HIV TESTING RATES: DOES PREDOMINANT CLINIC RACIAL/ETHNIC POPULATION PLAY A ROLE?
    Twito, Victoria
    Schubert, Finn D.
    Bhat, Sandeep
    Dapkins, Isaac
    JOURNAL OF GENERAL INTERNAL MEDICINE, 2020, 35 (SUPPL 1) : S89 - S89
  • [28] Type-testing of the TLD system used for the assessing the eye lens doses due to ionising radiation
    Petrovic, Jelena Stankovic
    Danilovic, Natasa
    Krzanovic, Nikola
    Topalovic, Dusan
    Zivanovic, Milos
    Bozovic, Predrag
    Todorovic, Natasa
    Nikolov, Jovana
    RADIATION MEASUREMENTS, 2023, 168
  • [29] Does Vitamin D deficiency play a role in peripheral neuropathy in Type 2 diabetes?
    Shehab, D.
    Al-Jarallah, K.
    Mojiminiyi, O. A.
    Al Mohamedy, H.
    Abdella, N. A.
    DIABETIC MEDICINE, 2012, 29 (01) : 43 - 49
  • [30] CIGARETTE-SMOKING IN PREGNANCY AND FETAL GROWTH - DOES THE TYPE OF TOBACCO PLAY A ROLE
    OLSEN, J
    INTERNATIONAL JOURNAL OF EPIDEMIOLOGY, 1992, 21 (02) : 279 - 284