Accuracy of digital and conventional implant impressions in edentulous jaws: A clinical comparative study

被引:0
|
作者
Kernen-Gintaute, A. [1 ,2 ]
Akulauskas, M. [4 ]
Kernen, F. [3 ]
Zitzmann, N. U. [2 ]
Spies, B. C. [1 ]
Burkhardt, F. [1 ]
机构
[1] Univ Freiburg, Fac Med, Ctr Dent Med, Dept Prosthet Dent,Med Ctr, Hugstetter Str 55, D-79106 Freiburg, Germany
[2] Univ Basel, Univ Ctr Dent Med, Dept Reconstruct Dent, Basel, Switzerland
[3] Univ Freiburg, Fac Med, Med Ctr, Dept Oral & Maxillofacial Surg,Translat Implantol, Freiburg, Germany
[4] Kaunas Univ Technol, Biomed Engn Inst, Kaunas, Lithuania
关键词
Digital impression; Intraoral scanner; Oral implants; Edentulous; Tactile scanner; PRECISION; SCANNER;
D O I
10.1016/j.jdent.2025.105559
中图分类号
R78 [口腔科学];
学科分类号
1003 ;
摘要
Objectives: This clinical study aimed to evaluate the accuracy of digital and conventional implant impressions in a fully edentulous maxilla and mandible. Methods: A 53-year-old edentulous patient with four maxillary and two mandibular implants was selected. Ten intraoral scans (IOS) and a conventional impression per jaw were taken. Clinically verified upper and lower plaster models were digitized using both optical (reference data, n = 10 per model) and tactile laboratory scanner (n = 10 per model). Accuracy was evaluated by comparing the precision and linear/angular deviations of the implants with the reference data. Statistical analyses were conducted using Student's t-test and Kruskal-Wallis test (alpha = 0.05). Results: In the maxilla, the most significant linear deviations exceeding the 100 mu m threshold were found with IOS between implants 1-4. In the mandible, all linear deviations remained below 55 mu m. Angular deviations between implants after IOS ranged from 0.01 degrees to 0.40 degrees in the mandible and <0.01 degrees to 1.86 degrees in the maxilla. After tactile scanning, linear deviations did not exceed 100 <mu>m threshold (except in one distance) and angular deviations ranged from 0.04 degrees to 0.54 degrees (mandible) and <0.01 degrees to 2.50 degrees (maxilla). The optical scanner demonstrated significantly higher precision (p < 0.001) compared to the IOS and tactile scanner. Conclusions: Given the significant deviations observed, especially in the maxilla, the optical scanner following conventional impressions remained the preferred method for fully edentulous cases due to its superior accuracy.
引用
收藏
页数:9
相关论文
共 50 条
  • [41] Implant assisted ortho-surgery in edentulous jaws: a clinical report
    Khojasteh, Arash
    Payaminia, Leila
    Alikhasi, Marzieh
    CLINICAL CASE REPORTS, 2015, 3 (11): : 920 - 926
  • [42] In Vitro Accuracy of Digital and Conventional Impressions for Full-Arch Implant-Supported Prostheses
    D'haese, Rani
    Vrombaut, Tom
    Roeykens, Herman
    Vandeweghe, Stefan
    JOURNAL OF CLINICAL MEDICINE, 2022, 11 (03)
  • [43] Do digital impressions have a greater accuracy for full-arch implant-supported reconstructions compared to conventional impressions? An in vitro study
    Shaikh, Mohsin
    Lakha, Tabrez
    Kheur, Supriya
    Qamri, Batul
    Kheur, Mohit
    THE JOURNAL OF INDIAN PROSTHODONTIC SOCIETY, 2022, 22 (04) : 398 - 404
  • [44] Effect of different impression coping and scan body designs on the accuracy of conventional versus digital implant impressions: An in vitro study
    Alkindi, Shaikha
    Hamdoon, Zaid
    Aziz, Ahmed M.
    JOURNAL OF DENTISTRY, 2024, 146
  • [45] Photogrammetry -: An alternative to conventional impressions in implant dentistry?: A clinical pilot study
    Jemt, T
    Bäck, T
    Petersson, A
    INTERNATIONAL JOURNAL OF PROSTHODONTICS, 1999, 12 (04) : 363 - 368
  • [46] Effect of reference objects on the accuracy of digital implant impressions in partially edentulous arches (vol 16, pg 302, 2024)
    Rutkunas, Vygandas
    Jegelevicius, Darius
    Pletkus, Justinas
    Auskalnis, Liudas
    Akulauskas, Mykolas
    Eyuboglu, Tan Firat
    Ozcan, Mutlu
    Gedrimiene, Agne
    JOURNAL OF ADVANCED PROSTHODONTICS, 2024, 16 (06): : 369 - 369
  • [47] Effect of Implant Divergence on the Accuracy of Definitive Casts Created from Traditional and Digital Implant-Level Impressions: An In Vitro Comparative Study
    Lin, Wei-Shao
    Harris, Bryan T.
    Elathamna, Eiad N.
    Abdel-Azim, Tamer
    Morton, Dean
    INTERNATIONAL JOURNAL OF ORAL & MAXILLOFACIAL IMPLANTS, 2015, 30 (01) : 102 - 109
  • [48] Accuracy of static fully guided implant placement in the posterior area of partially edentulous jaws: a cohort prospective study
    Gargallo-Albiol, Jordi
    Jose Zilleruelo-Pozo, Maria
    Lucas-Taule, Ernest
    Munoz-Penalver, Jesus
    Paternostro-Betancourt, Daniel
    Hernandez-Alfaro, Federico
    CLINICAL ORAL INVESTIGATIONS, 2022, 26 (03) : 2783 - 2791
  • [49] Accuracy of static fully guided implant placement in the posterior area of partially edentulous jaws: a cohort prospective study
    Jordi Gargallo-Albiol
    María José Zilleruelo-Pozo
    Ernest Lucas-Taulé
    Jesús Muñoz-Peñalver
    Daniel Paternostro-Betancourt
    Federico Hernandez-Alfaro
    Clinical Oral Investigations, 2022, 26 : 2783 - 2791
  • [50] Influence of intraoral scanning coverage on the accuracy of digital implant impressions - An in vitro study
    Wang, Zhen-yu
    Gong, Yu
    Liu, Fei
    Chen, Du
    Shen, Jie-fei
    JOURNAL OF DENTISTRY, 2024, 143