Optimizing Stem Length in Conversion Total Hip Arthroplasty: An Expanded Finite Element Analysis

被引:0
|
作者
Shimasaki, Koshiro [1 ]
Nishino, Tomofumi [1 ]
Yoshizawa, Tomohiro [1 ]
Watanabe, Ryunosuke [1 ]
Hirose, Fumi [1 ]
Yasunaga, Shota [1 ]
Mishima, Hajime [1 ]
机构
[1] Univ Tsukuba, Inst Med, Dept Orthopaed Surg, 1-1-1 Tennodai, Tsukuba 3058575, Japan
关键词
conversion total hip arthroplasty (cTHA); finite element analysis (FEA); stress concentration; periprosthetic fracture; stem length; FAILED INTERNAL-FIXATION; INTERTROCHANTERIC FRACTURE; PERIPROSTHETIC FRACTURES; CORTICAL DEFECTS; FEMORAL-NECK; SALVAGE; FEMUR; COMPLICATIONS; STRESS; BIOMECHANICS;
D O I
10.3390/jcm14041141
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
Background/Objectives: Stress concentration around distal screw-removal holes confers a major risk for periprosthetic fractures following conversion total hip arthroplasty (cTHA) for intertrochanteric femoral fractures. Optimal stem-selection criteria and guidelines for cTHA can improve clinical outcomes. We determined the influence of the cementless stem length on the stress distribution around distal screw-removal holes. Methods: For the finite element analysis, institutional data from preoperative CT scans of contralateral femurs of patients who underwent THA were used. To replicate the post-nail-removal state, we used 3D registration of standard triangulated language data of the intramedullary nail as an unused material to simulate distal screw-removal holes, located 135 mm from the proximal end of the intramedullary nail. Cementless stems of 130, 140, 150, and 160 mm were individually registered using STL data, and cTHA models were constructed accordingly. Using simulations under load conditions representing normal walking and stair climbing, the mean and maximum equivalent stress values around the distal screw-removal holes were calculated. For multiple comparisons, repeated-measures ANOVA with Bonferroni correction was employed. Results: Compared to the 130 mm stem, the 150 mm and 160 mm stems similarly reduced the maximum equivalent stress around the distal screw-removal holes, although the 140 mm stem showed no significant difference with other stems. Conclusions: A >= 150 mm stem length reliably mitigated stress concentration around distal screw-removal holes post-cTHA; it is the optimal choice for balancing effectiveness and risk of complications and may contribute to improved long-term clinical outcomes. This study provides practical evidence for stem selection in cTHA and offers valuable insights for future treatment guidelines.
引用
收藏
页数:18
相关论文
共 50 条
  • [1] Stress Analysis in Conversion Total Hip Arthroplasty: A Finite Element Analysis on Stem Length and Distal Screw Hole
    Shimasaki, Koshiro
    Nishino, Tomofumi
    Yoshizawa, Tomohiro
    Watanabe, Ryunosuke
    Hirose, Fumi
    Yasunaga, Shota
    Mishima, Hajime
    JOURNAL OF CLINICAL MEDICINE, 2025, 14 (01)
  • [2] Optimizing leg length correction in total hip arthroplasty
    Fujita, Kenji
    Kabata, Tamon
    Kajino, Yoshitomo
    Tsuchiya, Hiroyuki
    INTERNATIONAL ORTHOPAEDICS, 2020, 44 (03) : 437 - 443
  • [3] Optimizing leg length correction in total hip arthroplasty
    Kenji Fujita
    Tamon Kabata
    Yoshitomo Kajino
    Hiroyuki Tsuchiya
    International Orthopaedics, 2020, 44 : 437 - 443
  • [4] Total Hip Arthroplasty in Hip Osteoarthritis with Subtrochanteric Localized Periosteal Thickening: Preoperative Planning Using Finite Element Analysis to Determine the Optimal Stem Length
    Shimasaki, Koshiro
    Nishino, Tomofumi
    Yoshizawa, Tomohiro
    Watanabe, Ryunosuke
    Hirose, Fumi
    Yasunaga, Shota
    Mishima, Hajime
    JOURNAL OF CLINICAL MEDICINE, 2024, 13 (19)
  • [5] Effect of stem position and length on bone-stem constructs after cementless hip arthroplasty A FINITE ELEMENT ANALYSIS
    Kwak, D-K
    Bang, S-H
    Lee, S-J
    Park, J-H
    Yoo, J-H
    BONE & JOINT RESEARCH, 2021, 10 (04): : 250 - 258
  • [6] Finite Element Analysis of Proximal Cement Fixation in Total Hip Arthroplasty
    Ahmad, Maizatul Afirah
    Zulkifli, Nurul Nadhirah Mohamed Elias
    Shuib, Solehuddin
    Sulaiman, Shahrul Hisham
    Abdullah, Abdul Halim
    INTERNATIONAL JOURNAL OF TECHNOLOGY, 2020, 11 (05) : 1046 - 1055
  • [7] Finite Element Analysis of Cemented Hip Arthroplasty: Influence of Stem Tapers
    Abdullah, Abdul H.
    Asri, Mohd N. Mohd
    Alias, Mohd S.
    Giha, Tardan
    INTERNATIONAL MULTICONFERENCE OF ENGINEERS AND COMPUTER SCIENTISTS (IMECS 2010), VOLS I-III, 2010, : 2241 - +
  • [8] Finite element model of the failure of stem-cement interface in total hip arthroplasty
    Amiri, Riad
    Bouiadjra, Bel Abbes Bachir
    Amiri, Ahmed
    UPB Scientific Bulletin, Series D: Mechanical Engineering, 2019, 81 (03): : 201 - 212
  • [9] Static structural analysis of different stem designs used in total hip arthroplasty using finite element method
    Chethan, K. N.
    Zuber, Mohammad
    Bhat, Shyamasunder N.
    Shenoy, Satish B.
    Kini, Chandrakant R.
    HELIYON, 2019, 5 (06)
  • [10] A Conversion Total Hip Arthroplasty Is Not a Primary Total Hip Arthroplasty