Guidelines for Accurate and Precise Stable Isotope Analysis of Calcite, Dolomite, and Magnesite Using a Carbonate Device for Continuous Flow-Isotope Ratio Mass Spectrometry (CF-IRMS)

被引:0
|
作者
Kim, Soujung [1 ]
Kim, Sang-Tae [1 ]
Knyf, Martin [1 ]
机构
[1] McMaster Univ, Sch Earth Environm & Soc, Hamilton, ON, Canada
基金
加拿大创新基金会; 加拿大自然科学与工程研究理事会;
关键词
calcite; Ca-Mg carbonates; carbon and oxygen isotope compositions; dolomite; isotope ratio mass spectrometry; magnesite; PHOSPHORIC-ACID; OXYGEN; FRACTIONATION; TEMPERATURE; DELTA-O-18; DIOXIDE; CO2;
D O I
10.1002/rcm.9958
中图分类号
Q5 [生物化学];
学科分类号
071010 ; 081704 ;
摘要
Rationale: Carbonate minerals are one of the most popular samples for an automated sample preparation system for CF-IRMS, such as GasBench II and iso FLOW, but no standardized analytical protocols exist. This study gives guidelines on optimal analytic conditions for carbon and oxygen isotope analysis of Ca-Mg carbonates when using the carbonate-phosphoric acid reaction method. Methods: Calcite (CaCO3-McMaster Carrara), dolomite (CaMg(CO3)(2)-MRSI Dolomite), and magnesite (MgCO3-ROM Brazil Magnesite) with two grain size fractions (< 74 and 149-250 mu m) were reacted with 103% (specific gravity of 1.92) phosphoric acid under He atmosphere in 12-mL borosilicate glass vials to examine the full delta C-13 and delta O-18 evolution of acid-liberated CO2 for an extended reaction time of up to 12-30 days at 25 degrees C and up to 3-7 days at 72 degrees C. Results: At 25 degrees C, the optimal reaction time of calcite is 1 day for both grain size fractions while the optimal reaction time of 2-10 day is suggested for dolomite with a grain size of < 74 mu m. At 72 degrees C, 30-min to 12-h or 45-min to 12-h reaction is optimal for calcite with < 74-mu m or 149- to 250-mu m grain size fraction, respectively, whereas dolomite requires 12-h to 1-day reaction for both grain size fractions. The only optimal condition for magnesite is 6-7 days of reaction with < 74-mu m grain size at 72 degrees C. Conclusions: To determine precise and accurate delta C-13 and delta O-18 values of a carbonate mineral using the carbonate-phosphoric acid reaction method, an optimal reaction time must be assessed for a given analytical condition to avoid nonequilibrium isotope effects and unnecessary oxygen isotope exchange of acid-liberated CO2 in the carbonate reaction vessel. Our experimental result provides a guideline for the accurate and precise stable isotope analysis of Ca-Mg carbonate minerals.
引用
收藏
页数:19
相关论文
共 50 条
  • [1] CF-IRMS (CONTINUOUS FLOW-ISOTOPE RATIO MASS-SPECTROMETRY), AN ANALYTICAL STABLE-ISOTOPE METHOD FOR BIOLOGICAL SAMPLES
    WAGNER, G
    ISOTOPENPRAXIS, 1992, 28 (01): : 114 - 115
  • [2] An online method to determine chlorine stable isotope composition by continuous flow isotope ratio mass spectrometry (CF-IRMS) coupled with a Gasbench Ⅱ
    刘运德
    周爱国
    甘义群
    刘存富
    余婷婷
    李小倩
    JournalofCentralSouthUniversity, 2013, 20 (01) : 193 - 198
  • [3] An online method to determine chlorine stable isotope composition by continuous flow isotope ratio mass spectrometry (CF-IRMS) coupled with a Gasbench II
    Yun-de Liu
    Ai-guo Zhou
    Yi-qun Gan
    Cun-fu Liu
    Ting-ting Yu
    Xiao-qian Li
    Journal of Central South University, 2013, 20 : 193 - 198
  • [4] An online method to determine chlorine stable isotope composition by continuous flow isotope ratio mass spectrometry (CF-IRMS) coupled with a Gasbench II
    Liu Yun-de
    Zhou Ai-guo
    Gan Yi-qun
    Liu Cun-fu
    Yu Ting-ting
    Li Xiao-qian
    JOURNAL OF CENTRAL SOUTH UNIVERSITY, 2013, 20 (01) : 193 - 198
  • [5] Sulfur isotope analysis of sulfide and sulfate minerals by continuous flow-isotope ratio mass spectrometry
    Grassineau, NV
    Mattey, DP
    Lowry, D
    ANALYTICAL CHEMISTRY, 2001, 73 (02) : 220 - 225
  • [6] Analysis of Compound-Specific Chlorine Stable Isotopes of Vinyl Chloride by Continuous Flow-Isotope Ratio Mass Spectrometry (FC-IRMS)
    Shouakar-Stash, Orfan
    Frape, S. K.
    Aravena, R.
    Gargini, A.
    Pasini, M.
    Drimmie, R. J.
    ENVIRONMENTAL FORENSICS, 2009, 10 (04) : 299 - 306
  • [7] Triple oxygen stable isotope analysis of nitrite measured using continuous flow isotope ratio mass spectrometry *
    Walters, Wendell W.
    Hastings, Meredith G.
    METHODSX, 2023, 11
  • [8] Isotope ratio analysis of carbon and nitrogen by elemental analyser continuous flow isotope ratio mass spectrometry (EA-CF-IRMS) without the use of a reference gas
    Bay, L. J.
    Chan, S. H.
    Walczyk, T.
    JOURNAL OF ANALYTICAL ATOMIC SPECTROMETRY, 2015, 30 (01) : 310 - 314
  • [9] Oxygen isotope analysis of silicates using continuous flow isotope ratio mass spectrometry
    Kingston, A. W.
    Weis, D.
    Francois, R.
    GEOCHIMICA ET COSMOCHIMICA ACTA, 2008, 72 (12) : A474 - A474
  • [10] Organic matter removal for continuous flow isotope ratio mass spectrometry analysis of carbon and oxygen isotope compositions of calcite or dolomite in organic-rich samples
    Chaduteau, Carine
    Ader, Magali
    Lebeau, Oanez
    Landais, Guillaume
    Busigny, Vincent
    LIMNOLOGY AND OCEANOGRAPHY-METHODS, 2021, 19 (08): : 523 - 539