Radiopharmaceutical imaging of phantoms and patients with MACACO III Compton camera

被引:0
|
作者
Roser, Jorge [1 ,3 ]
Barrientos, Luis [1 ]
Bello, Pilar [2 ]
Borja-Lloret, Marina [1 ]
Calatayud, Jose Manuel [2 ]
Casana, Jose V. [1 ]
Hueso-Gonzalez, Fernando [1 ]
Perez-Curbelo, Javier [1 ]
Prado, Stefan [2 ]
Ros, Ana [1 ]
Senra, Cesar [1 ]
Viegas, Rita [1 ]
Torres-Espallardo, Irene [2 ]
Llosa, Gabriela [1 ]
机构
[1] UV, CSIC, Inst Fis Corpuscular IFIC, Valencia, Spain
[2] La Fe Hosp HUiP, Nucl Med Dept, Valencia, Spain
[3] Univ Lubeck, Inst Med Engn, Lubeck, Schleswig Holst, Germany
关键词
Compton cameras; Gamma cameras; Radiopharmaceutical imaging;
D O I
10.1016/j.ejmp.2025.104928
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Background and purpose: The use of Compton cameras for imaging radioisotopes in nuclear medicine offers potential advantages over conventional gamma cameras in some aspects. In this work, we present the first results in this field with the Compton camera prototype MACACO III after a measurement campaign in La Fe Hospital (Valencia, Spain). Methods: Images of Derenzo-like phantoms filled with 18F-FDG and 131I-NaI have been obtained with MACACO III and, in the latter case, with gamma cameras available in the hospital. Subsequently, a metastatic lesion in a patient treated with 131I-NaI after thyroid gland resection was imaged. Results: The 18F-FDG phantom images show the prototype capability of identifying up to 4 mm diameter rods. The measurements with 131I-NaI-filled phantoms demonstrate the clear improvement of the spatial resolution attainable with MACACO III when compared with images obtained with the gamma cameras available in the hospital. Finally, the image of the patient metastatic lesion obtained with the Compton camera was found to be consistent with the gamma camera inspection image. Conclusion: These encouraging results demonstrate MACACO III's capability for radiopharmaceutical imaging and open a door for its application to nuclear medicine.
引用
收藏
页数:8
相关论文
共 50 条
  • [31] A flexible geometry Compton camera for industrial gamma ray imaging
    Royle, GJ
    Speller, RD
    1996 IEEE NUCLEAR SCIENCE SYMPOSIUM - CONFERENCE RECORD, VOLS 1-3, 1997, : 821 - 824
  • [32] Design of a small laboratory Compton camera for the imaging of positron emitters
    Univ Coll London, London, United Kingdom
    IEEE Transactions on Nuclear Science, 2000, 47 (3 III) : 1155 - 1162
  • [33] Gamma-ray Compton camera imaging with a segmented HPGe
    Schmid, GJ
    Beckedahl, DA
    Kammeraad, JE
    Blair, JJ
    Vetter, K
    Kuhn, A
    NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 2001, 459 (03): : 565 - 576
  • [34] Study on the Imaging Resolution of a Multiply-stacked Compton Camera
    Kim, N. Y.
    Lee, J. H.
    Lee, C. S.
    JOURNAL OF THE KOREAN PHYSICAL SOCIETY, 2009, 55 (04) : 1409 - 1415
  • [35] Investigating the Impact of Secondaries Neutrons On Compton Camera for Medical Imaging
    Peterson, S.
    Maggi, P.
    Panthi, R.
    Mackin, D.
    Beddar, S.
    Polf, J.
    MEDICAL PHYSICS, 2020, 47 (06) : E632 - E632
  • [36] DESIGN OF A COMPTON CAMERA FOR IMAGING 662 KEV RADIONUCLIDE DISTRIBUTIONS
    ROYLE, GJ
    SPELLER, RD
    NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 1994, 348 (2-3): : 623 - 626
  • [37] Multi nuclides simultaneous imaging with the Electron Tracking Compton Camera
    Kabuki, Shiaeto
    Kubo, Hidetoshi
    Miuchi, Kentaro
    Kimura, Hiroyuki
    Amano, Hiroo
    Kawashima, Hidekazu
    Ueda, Masashi
    Nakamoto, Yuji
    Togashi, Kori
    Saji, Hideo
    Tanimori, Toru
    YAKUGAKU ZASSHI-JOURNAL OF THE PHARMACEUTICAL SOCIETY OF JAPAN, 2008, 128 : 46 - 46
  • [38] Development of a Compton camera for prompt-gamma medical imaging
    Aldawood, S.
    Thirolf, P. G.
    Miani, A.
    Boehmer, M.
    Dedes, G.
    Gernhauser, R.
    Lang, C.
    Liprandi, S.
    Maier, L.
    Marinsek, T.
    Mayerhofer, M.
    Schaart, D. R.
    Lozano, I. Valencia
    Parodi, K.
    RADIATION PHYSICS AND CHEMISTRY, 2017, 140 : 190 - 197
  • [39] Hodoscope Coincidence Imaging for Hadron Therapy Using a Compton Camera
    Gillam, J. E.
    Torres-Espallardo, I.
    Lacasta, C.
    Llosa, G.
    Barrio, J.
    Stankova, V.
    Solaz, C.
    Rafecas, M.
    2011 IEEE NUCLEAR SCIENCE SYMPOSIUM AND MEDICAL IMAGING CONFERENCE (NSS/MIC), 2011, : 3508 - 3513
  • [40] Development of semiconductor imaging detectors for a Si/CdTe Compton camera
    Watanabe, Shin
    Takeda, Shin'ichiro
    Ishikawa, Shin-nosuke
    Odaka, Hirokazu
    Ushio, Masayoshi
    Tanaka, Takaaki
    Nakazawa, Kazuhiro
    Takahashi, Tadayuki
    Tajima, Hiroyasu
    Fukazawa, Yasushi
    Kuroda, Yoshikatsu
    Onishi, Mitsunobu
    NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 2007, 579 (02): : 871 - 877