Verification of the Vitek Reveal System for Direct Antimicrobial Susceptibility Testing in Gram-Negative Positive Blood Cultures

被引:1
|
作者
Menchinelli, Giulia [1 ]
Squitieri, Damiano [2 ]
Magri, Carlotta [2 ]
De Maio, Flavio [1 ]
D'Inzeo, Tiziana [1 ,2 ]
Cacaci, Margherita [2 ]
De Angelis, Giulia [1 ,2 ]
Sanguinetti, Maurizio [1 ,2 ]
Posteraro, Brunella [2 ,3 ]
机构
[1] Fdn Policlin Univ A Gemelli IRCCS, Dipartimento Sci Lab Ematolog, I-00168 Rome, Italy
[2] Univ Cattolica Sacro Cuore, Dipartimento Sci Biotecnol Base, Clin Intensivol & Perioperatorie, I-00168 Rome, Italy
[3] Fdn Policlin Univ A Gemelli IRCCS, Unita Operativa Med Precis Microbiol Clin, Direz Sci, I-00168 Rome, Italy
来源
ANTIBIOTICS-BASEL | 2024年 / 13卷 / 11期
关键词
antimicrobial susceptibility testing; blood cultures; international standard for organization; verification; Vitek Reveal assay; PSEUDOMONAS-AERUGINOSA; ENTEROBACTERIACEAE;
D O I
10.3390/antibiotics13111058
中图分类号
R51 [传染病];
学科分类号
100401 ;
摘要
Background/Objectives: The International Organization for Standardization (ISO) 20776-2:2021, which replaces ISO 20776-2:2007, focuses solely on the performance of antimicrobial susceptibility testing (AST) assays, emphasizing the ISO 20776-1 broth microdilution method as the reference standard. Consequently, categorical agreement (CA) and associated errors should not be applied. We verified the Vitek Reveal AST assay according to both ISO 20776-2:2021 and ISO 20776-2:2007 criteria. Methods: Samples from 100 simulated and clinical Gram-negative (GN) positive blood cultures (PBCs) were tested at a large teaching hospital. The simulated GN-PBCs were obtained from a hospital collection of isolates selected to represent diverse antimicrobial resistance profiles. The Reveal assay results were compared with those from the reference assay, and the time to result (TTR) for the Reveal assay was calculated. Results: The essential agreement rates were 96.1% (816/849) for simulated and 98.8% (929/940) for clinical GN-PBC samples. The bias values were -3.1 for simulated and -11.0 for clinical samples. The CA rates were 97.7% (808/827) for simulated and 99.2% (924/931) for clinical samples. The mean TTR +/- SD (hours) for resistant organisms was significantly lower (4.40 +/- 1.15) than that for susceptible, increased exposure (5.52 +/- 0.48) and susceptible (5.54 +/- 0.49) organisms. Conclusions: Our findings reinforce the potential of the Reveal assay as a valuable tool and support its implementation in clinical microbiology laboratories.
引用
收藏
页数:13
相关论文
共 50 条
  • [1] Evaluation of the Vitek® Reveal™ system for rapid antimicrobial susceptibility testing of Gram-negative pathogens, including ESBL, CRE and CRAB, from positive blood cultures
    Antonelli, Alberto
    Cuffari, Sara
    Casciato, Benedetta
    Giani, Tommaso
    Rossolini, Gian Maria
    DIAGNOSTIC MICROBIOLOGY AND INFECTIOUS DISEASE, 2024, 110 (04)
  • [2] Evaluation of the VITEK 2 system for rapid direct identification and susceptibility testing of gram-negative bacilli from positive blood cultures
    Ling, TKW
    Liu, ZK
    Cheng, AFB
    JOURNAL OF CLINICAL MICROBIOLOGY, 2003, 41 (10) : 4705 - 4707
  • [3] Direct inoculation of positive blood cultures using the Phoenix system for antimicrobial susceptibility testing of both Gram-positive and Gram-negative bacteria
    Florio, Walter
    Barnini, Simona
    Morici, Paola
    Lupetti, Antonella
    JOURNAL OF MEDICAL MICROBIOLOGY, 2015, 64 : 582 - 585
  • [4] Direct antimicrobial susceptibility testing of gram-negative bacilli in blood cultures by an electrochemical method
    Huang, AH
    Wu, JJ
    Weng, YM
    Ding, HC
    Chang, TC
    JOURNAL OF CLINICAL MICROBIOLOGY, 1998, 36 (10) : 2882 - 2886
  • [5] Rapid direct identification and susceptibility testing of gram-negative rods from positive blood cultures by VITEK 2 in common practice
    Matten, J.
    Scharmann, C.
    Schwarz, R.
    INTERNATIONAL JOURNAL OF MEDICAL MICROBIOLOGY, 2008, 298 : 6 - 7
  • [6] Saponin promotes rapid identification and antimicrobial susceptibility profiling of Gram-positive and Gram-negative bacteria in blood cultures with the Vitek 2 system
    Lupetti, A.
    Barnini, S.
    Morici, P.
    Ghelardi, E.
    Nibbering, P. H.
    Campa, M.
    EUROPEAN JOURNAL OF CLINICAL MICROBIOLOGY & INFECTIOUS DISEASES, 2013, 32 (04) : 493 - 502
  • [7] Saponin promotes rapid identification and antimicrobial susceptibility profiling of Gram-positive and Gram-negative bacteria in blood cultures with the Vitek 2 system
    A. Lupetti
    S. Barnini
    P. Morici
    E. Ghelardi
    P. H. Nibbering
    M. Campa
    European Journal of Clinical Microbiology & Infectious Diseases, 2013, 32 : 493 - 502
  • [8] Accuracy of the VITEK® 2 system for a rapid and direct identification and susceptibility testing of Gram-negative rods and Gram-positive cocci in blood samples
    Nimer, N. A.
    Al-Saa'da, R. J.
    Abuelaish, O.
    EASTERN MEDITERRANEAN HEALTH JOURNAL, 2016, 22 (03) : 193 - 200
  • [9] Rapid antimicrobial susceptibility testing by VITEK®2 directly from blood cultures in patients with Gram-negative rod bacteremia
    Hogan, C. A.
    Watz, N.
    Budvytiene, I.
    Banaei, N.
    DIAGNOSTIC MICROBIOLOGY AND INFECTIOUS DISEASE, 2019, 94 (02) : 116 - 121
  • [10] Identification and antimicrobial susceptibility testing of Gram-positive and Gram-negative bacteria from positive blood cultures using the Accelerate Pheno™ system
    Ullberg, Mans
    Ozenci, Volkan
    EUROPEAN JOURNAL OF CLINICAL MICROBIOLOGY & INFECTIOUS DISEASES, 2020, 39 (01) : 139 - 149