Reducing electromagnetic interference in MR thermometry: A comparison of setup configurations for MR-guided microwave ablations

被引:1
|
作者
Schroeer, Simon [1 ,3 ]
Duex, Daniel [1 ,3 ]
Caballero, Josef Joaquin Loening [1 ,3 ]
Glandorf, Julian [1 ,3 ]
Gerlach, Thomas [2 ,3 ]
Horstmann, Dominik [1 ,3 ]
Belker, Othmar [1 ,3 ]
Gutt, Moritz [1 ,3 ]
Wacker, Frank [1 ,3 ]
Speck, Oliver [2 ,3 ]
Hensen, Bennet [1 ,3 ]
Gutberlet, Marcel [1 ,3 ]
机构
[1] Hannover Med Sch, Carl Neuberg Str 1, D-30625 Hannover, Lower Saxony, Germany
[2] Otto von Guericke Univ, Univ Pl 2, D-39106 Magdeburg, Saxony Anhalt, Germany
[3] Res Campus STIMULATE, Otto Hahn Str 2, D-39106 Magdeburg, Saxony Anhalt, Germany
来源
ZEITSCHRIFT FUR MEDIZINISCHE PHYSIK | 2025年 / 35卷 / 01期
关键词
MR thermometry; MR-guided microwave ablation; Electromagnetic compatibility; Electromagnetic interfer- ence; Shielding; CANCER; LIVER;
D O I
10.1016/j.zemedi.2024.07.004
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Magnetic Resonance (MR) thermometry is used for the monitoring of MR-guided microwave ablations (MWA), and for the intraoperative evaluation of ablation regions. Nevertheless, the accuracy of temperature mapping may be compromised by electromagnetic interference emanating from the microwave (MW) generator. This study evaluated different setups for improving magnetic resonance imaging (MRI) during MWA with a modified MW generator. MWA was performed in 15 gel phantoms comparing three setups: The MW generator was placed outside the MR scanner room, either connected to the MW applicator using a penetration panel with a radiofrequency (RF) filter and a 7 m coaxial cable (Setup 1), or through a waveguide using a 5 m coaxial cable (Setup 2). Setup 3 employed the MW generator within the MR scan room, connected by a 5 m coaxial cable. The coaxial cables in setups 2 and 3 were modified with custom shielding to reduce interference. The setups during ablation (active setup) were compared to a reference setup without the presence of the MW system. Thermometry and thermal dose maps (CEM43 model) were compared for the three configurations. Primary endpoints for assessment were signal-to-noise ratio (SNR), temperature precision, S & oslash;rensen-Dice-Coefficient (DSC), and RF-noise spectra. Setup 3 showed highly significant electromagnetic interference during ablation with a SNR decrease by -60.4%+/- 13.5% (p<0.001) compared to reference imaging. For setup 1 and setup 2 no significant decrease in SNR was measured with differences of -2.9%+/- 9.8% (p=0.6) and -1.5%+/- 12.8% (p=0.8), respectively. SNR differences were significant between active setups 1 and 3 with -51.2%+/- 16.1% (p<0.001) and between active setups 2 and 3 with -59.0%+/- 15.5% (p<0.001) but not significant between active setups 1 and 2 with 19.0%+/- 13.7% (p=0.09). Furthermore, no significant differences were seen in temperature precision or DSCs between all setups, ranging from 0.33 degrees C +/- 0.04 degrees C (Setup 1) to 0.38 degrees C +/- 0.06 degrees C (Setup 3) (p=0.6) and from 87.0%+/- 1.6% (Setup 3) to 88.1%+/- 1.6% (Setup 2) (p=0.58), respectively. Both setups (1 and 2) with the MW generator outside the MR scanner room were beneficial to reduce electromagnetic interference during MWA. Moreover, provided that a shielded cable is utilized in setups 2 and 3, all configurations displayed negligible differences in temperature precision and DSCs, indicating that the location of the MW generator does not significantly impact the accuracy of thermometry during MWA.
引用
收藏
页码:59 / 68
页数:10
相关论文
共 50 条
  • [41] Multi-gradient echo MR thermometry for monitoring of the near-field area during MR-guided high intensity focused ultrasound heating
    Lam, Mie K.
    de Greef, Martijn
    Bouwman, Job G.
    Moonen, Chrit T. W.
    Viergever, Max A.
    Bartels, Lambertus W.
    PHYSICS IN MEDICINE AND BIOLOGY, 2015, 60 (19): : 7729 - 7746
  • [42] MR-guided conformal microwave imaging for enhanced inclusion detection within irregularly shaped volumes
    Epstein, Neil R.
    Meaney, Paul M.
    Paulsen, Keith D.
    MEDICAL IMAGING 2013: BIOMEDICAL APPLICATIONS IN MOLECULAR, STRUCTURAL, AND FUNCTIONAL IMAGING, 2013, 8672
  • [43] Feasibility of Respiratory Triggering for MR-Guided Microwave Ablation of Liver Tumors Under General Anesthesia
    Shigehiro Morikawa
    Toshiro Inubushi
    Yoshimasa Kurumi
    Shigeyuki Naka
    Koichiro Sato
    Koichi Demura
    Tohru Tani
    Hasnine A Haque
    CardioVascular and Interventional Radiology, 2004, 27 : 370 - 373
  • [44] Feasibility of respiratory triggering for MR-guided microwave ablation of liver tumors under general anesthesia
    Morikawa, S
    Inubushi, T
    Kurumi, Y
    Naka, S
    Sato, K
    Demura, K
    Tani, T
    Haque, HA
    CARDIOVASCULAR AND INTERVENTIONAL RADIOLOGY, 2004, 27 (04) : 370 - 373
  • [45] Temperature estimation for MR-guided microwave hyperthermia using block-based compressed sensing
    Faridi, Pegah
    Shrestha, Tej B.
    Pyle, Marla
    Basel, Matthew T.
    Bossmann, Stefan H.
    Prakash, Punit
    Natarajan, Balasubramaniam
    42ND ANNUAL INTERNATIONAL CONFERENCES OF THE IEEE ENGINEERING IN MEDICINE AND BIOLOGY SOCIETY: ENABLING INNOVATIVE TECHNOLOGIES FOR GLOBAL HEALTHCARE EMBC'20, 2020, : 5057 - 5060
  • [46] Comparison and Validation of Distortion Correction on a Low-Field MR-Guided Radiotherapy System
    Gao, Y.
    Pham, J.
    Kalbasi, A.
    Raldow, A.
    Hu, P.
    Yang, Y.
    MEDICAL PHYSICS, 2020, 47 (06) : E414 - E415
  • [47] Enhancing MR-guided Laser Interstitial Thermal Therapy Planning using U-Net: A Data-Driven Approach for Predicting MR Thermometry Images
    Sadatamin, Saba
    Ketabi, Sara
    Donszelmann-Lund, Elise
    Abtahi, Saba
    Chaban, Yuri
    Robbins, Steven
    Tyc, Richard
    Khalvati, Farzad
    Waspe, Adam C.
    Kahrs, Lueder A.
    Drake, James M.
    IMAGE-GUIDED PROCEDURES, ROBOTIC INTERVENTIONS, AND MODELING, MEDICAL IMAGING 2024, 2024, 12928
  • [48] Automatic spatial and temporal temperature control for MR-guided focused ultrasound using fast 3D MR thermometry and multispiral trajectory of the focal point
    Mougenot, C
    Salomir, R
    Palussière, J
    Grenier, N
    Moonen, CTW
    MAGNETIC RESONANCE IN MEDICINE, 2004, 52 (05) : 1005 - 1015
  • [49] MR-Guided High-Power Microwave Ablation in Hepatic Malignancies: Initial Results in Clinical Routine
    Winkelmann, Moritz T.
    Gohla, Georg
    Kuebler, Jens
    Weiss, Jakob
    Clasen, Stephan
    Nikolaou, Konstantin
    Hoffmann, Ruediger
    CARDIOVASCULAR AND INTERVENTIONAL RADIOLOGY, 2020, 43 (11) : 1631 - 1638
  • [50] Dosimetry & toxicity comparison between CBCT-Guided and MR-Guided Prostate Ultra-hypofractionated RT
    Kong, V.
    Padayachee, J.
    Dang, J.
    Li, W.
    Malkov, V.
    Winter, J.
    Navarro, I.
    Berlin, A.
    Helou, J.
    Raman, S.
    Chung, P.
    RADIOTHERAPY AND ONCOLOGY, 2022, 170 : S844 - S845