R2* Impact on Hepatic Fat Quantification With a Commercial Single Voxel Technique at 1.5 and 3.0 T

被引:1
|
作者
Fortier, Veronique [1 ,2 ,3 ,4 ]
Mohamed, Ahmed [5 ]
McNabb, Evan [1 ]
Dana, Jeremy [1 ]
Zakarian, Rita [6 ]
Levesque, Ives [3 ,4 ,6 ]
Reinhold, Caroline [1 ,2 ,6 ,7 ]
机构
[1] McGill Univ, Hlth Ctr, Dept Med Imaging, Glen Site C02 5505 1001 boul Decarie, Montreal, PQ H4A 3J1, Canada
[2] McGill Univ, Diag Radiol, Montreal, PQ, Canada
[3] McGill Univ, Gerald Bronfman Dept Oncol, Montreal, PQ, Canada
[4] McGill Univ, Med Phys Unit, Montreal, PQ, Canada
[5] Cairo Univ, Radiol Dept, NCI, Cairo, Egypt
[6] McGill Univ, Res Inst, Hlth Ctr, Montreal, PQ, Canada
[7] Montreal Imaging Experts Inc, Montreal, PQ, Canada
关键词
liver; fat quantification; HISTO; iron; R-2; IRON; MRI; LIVER; SPECTROSCOPY; DIXON; RELAXATION; FRACTION;
D O I
10.1177/08465371241255896
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Rationale and Objectives: Fat quantification accuracy using a commercial single-voxel high speed T2-corrected multi-echo (HISTO) technique and its robustness to R2* variations at 3.0 T, such as those introduced by iron in liver, has not been fully established. This study evaluated HISTO at 3.0 T and sought to reproduce results at 1.5 T. Methods: Phantoms were prepared with a range of fat content and R2*. Data were acquired at 1.5 T and 3.0 T, using HISTO and a Dixon technique. Fat quantification accuracy was evaluated as a function of R2*. The patient study included 239 consecutive patients. Data were acquired at 1.5 T or 3.0 T, using HISTO and Dixon techniques. The techniques were compared using Bland-Altman plots. Bias significance was evaluated using a one-sample t-test. Results: In phantoms, HISTO was accurate within 10% up to a R2* of 100 s-1 at both field strengths, while Dixon was accurate within 10% where R2* was accurately quantified (up to 350 s-1 at 1.5 T, and 550 s-1 at 3.0 T). In patients, where R2* was <100 s-1, fat quantification from both techniques agreed at 1.5 T (P = .71), but not at 3.0 T (P = .007), with a bias <1%. Conclusion: Results suggest that HISTO is reliable when R2* is <100 s-1, corresponding to patients with at most mild liver iron overload, and that it should be used with caution when R2* is >100 s-1. Dixon should be preferred for hepatic fat quantification due to its robustness to R2* variations. Visual Abstract This is a visual representation of the abstract.
引用
收藏
页码:838 / 846
页数:9
相关论文
共 50 条
  • [31] Effect of gadolinium on hepatic fat quantification using multi-echo reconstruction technique with T2* correction and estimation
    Mingmei Ge
    Jing Zhang
    Bing Wu
    Zhiqin Liu
    Hai Song
    Xiangfeng Meng
    Xinhuai Wu
    European Radiology, 2016, 26 : 1913 - 1920
  • [32] Cross-Sectional Correlation Between Hepatic R2☆ and Proton Density Fat Fraction (PDFF) in Children With Hepatic Steatosis
    Mamidipalli, Adrija
    Hamilton, Gavin
    Manning, Paul
    Hong, Cheng William
    Park, Charlie C.
    Wolfson, Tanya
    Hooker, Jonathan
    Heba, Elhamy
    Schlein, Alexandra
    Gamst, Anthony
    Durelle, Janis
    Paiz, Melissa
    Middleton, Michael S.
    Schwimmer, Jeffrey B.
    Sirlin, Claude B.
    JOURNAL OF MAGNETIC RESONANCE IMAGING, 2018, 47 (02) : 418 - 424
  • [33] Simulation of a virtual liver iron overload model and R2 estimation using multispectral fat-water models for GRE and UTE acquisitions at 1.5 T and 3 T
    Neupane, Prasiddhi
    Shrestha, Utsav
    Brasher, Sarah
    Abramson, Zachary
    Tipirneni-Sajja, Aaryani
    NMR IN BIOMEDICINE, 2023, 36 (12)
  • [34] Physiological gradient in lumbar spine fat fraction and R2*and its impact on osteoporosis diagnosis
    Lee, Haein
    Yun, Jae Sung
    Park, Sunghoon
    Kwack, Kyu-Sung
    SPINE JOURNAL, 2024, 24 (03): : 479 - 487
  • [35] Simultaneous quantification of hepatic MRI-PDFF and R2* in a rabbit model with nonalcoholic fatty liver disease
    Xiaomin Wang
    Xiaojing Zhang
    Lin Ma
    Shengli Li
    Science China(Life Sciences), 2018, 61 (09) : 1107 - 1114
  • [36] Simultaneous quantification of hepatic MRI-PDFF and R2* in a rabbit model with nonalcoholic fatty liver disease
    Xiaomin Wang
    Xiaojing Zhang
    Lin Ma
    Shengli Li
    Science China Life Sciences, 2018, 61 : 1107 - 1114
  • [37] Simultaneous quantification of hepatic MRI-PDFF and R2*in a rabbit model with nonalcoholic fatty liver disease
    Wang, Xiaomin
    Zhang, Xiaojing
    Ma, Lin
    Li, Shengli
    SCIENCE CHINA-LIFE SCIENCES, 2018, 61 (09) : 1107 - 1114
  • [38] Detection of hepatic and pancreatic fat infiltration in type II diabetes mellitus patients with IDEAL-Quant using 3.0T MR: comparison with single-voxel proton spectroscopy
    Ma Jing
    Song Zhiqiang
    Yan Fuhua
    CHINESE MEDICAL JOURNAL, 2014, 127 (20) : 3548 - 3552
  • [39] Does Fat Suppression via Chemically Selective Saturation Affect R2*-MRI for Transfusional Iron Overload Assessment? A Clinical Evaluation at 1.5T and 3T
    Krafft, Axel J.
    Loeffler, Ralf B.
    Song, Ruitian
    Bian, Xiao
    McCarville, M. Beth
    Hankins, Jane S.
    Hillenbrand, Claudia M.
    MAGNETIC RESONANCE IN MEDICINE, 2016, 76 (02) : 591 - 601
  • [40] R2* for myocardial iron quantification at 3T CMR using an advanced shimming algorithm
    Gabriel C Camargo
    Tamara Rothstein
    Flavia P Jjunqueira
    Daniel C Quintella
    Ralph Strecker
    Andreas Greiser
    Joao A Lima
    Patricia B Rizzi
    Maria Eduarda Derenne
    Ronaldo SL Lima
    Ilan Gottlieb