Prospective Evaluation of Accelerated Brain MRI Using Deep Learning-Based Reconstruction: Simultaneous Application to 2D Spin-Echo and 3D Gradient-Echo Sequences

被引:0
|
作者
Choi, Kyu Sung [1 ,2 ]
Park, Chanrim [1 ,4 ]
Lee, Ji Ye [1 ,2 ]
Lee, Kyung Hoon [1 ,5 ]
Jeon, Young Hun [1 ]
Hwang, Inpyeong [1 ,2 ]
Yoo, Roh Eul [1 ,2 ]
Yun, Tae Jin [1 ,2 ]
Lee, Mi Ji [3 ]
Jung, Keun-Hwa [3 ]
Kang, Koung Mi [1 ,2 ,6 ]
机构
[1] Seoul Natl Univ Hosp, Dept Radiol, Seoul, South Korea
[2] Seoul Natl Univ, Coll Med, Dept Radiol, Seoul, South Korea
[3] Seoul Natl Univ Hosp, Dept Neurol, Seoul, South Korea
[4] Seoul Natl Univ Hosp, Biomed Res Inst, Seoul, South Korea
[5] Sungkyunkwan Univ, Kangbuk Samsung Hosp, Sch Med, Dept Radiol, Seoul, South Korea
[6] Seoul Natl Univ, Seoul Natl Univ Hosp, Coll Med, Dept Radiol, 101 Daehak Ro, Seoul 03080, South Korea
基金
新加坡国家研究基金会;
关键词
Deep learning; Artifacts; Signal-to-noise ratio; Image reconstruction; Volumetric analysis; ITERATIVE RECONSTRUCTION;
D O I
10.3348/kjr.2024.0653
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Objective: To prospectively evaluate the effect of accelerated deep learning-based reconstruction (Accel-DL) on improving brain magnetic resonance imaging (MRI) quality and reducing scan time compared to that in conventional MRI. Materials and Methods: This study included 150 participants (51 male; mean age 57.3 +/- 16.2 years). Each group of 50 participants was scanned using one of three 3T scanners from three different vendors. Conventional and Accel-DL MRI images were obtained from each participant and compared using 2D T1- and T2-weighted and 3D gradient-echo sequences. Accel-DL acquisition was achieved using optimized scan parameters to reduce the scan time, with the acquired images reconstructed using U-Net-based software to transform low-quality, undersampled k-space data into high-quality images. The scan times of Accel-DL and conventional MRI methods were compared. Four neuroradiologists assessed the overall image quality, structural delineation, and artifacts using Likert scale (5- and 3-point scales). Inter-reader agreement was assessed using Fleiss' kappa coefficient. Signal-to-noise ratio (SNR) and contrast-to-noise ratio (CNR) were calculated, and volumetric quantification of regional structures and white matter hyperintensities (WMHs) was performed. Results: Accel-DL showed a mean scan time reduction of 39.4% (range, 24.2%-51.3%). Accel-DL improved overall image quality (3.78 +/- 0.71 vs. 3.36 +/- 0.61, P < 0.001), structure delineation (2.47 +/- 0.61 vs. 2.35 +/- 0.62, P < 0.001), and artifacts (3.73 +/- 0.72 vs. 3.71 +/- 0.69, P = 0.016). Inter-reader agreement was fair to substantial (kappa = 0.34-0.50). SNR and CNR increased in Accel-DL (82.0 +/- 23.1 vs. 31.4 +/- 10.8, P = 0.02; 12.4 +/- 4.1 vs. 4.4 +/- 11.2, P = 0.02). Bland-Altman plots revealed no significant differences in the volumetric measurements of 98.2% of the relevant regions, except in the deep gray matter, including the thalamus. Five of the six lesion categories showed no significant differences in WMH segmentation, except for leukocortical lesions (r = 0.64 +/- 0.29). Conclusion: Accel-DL substantially reduced the scan time and improved the quality of brain MRI in both spin-echo and gradient- echo sequences without compromising volumetry, including lesion quantification.
引用
收藏
页码:54 / 64
页数:11
相关论文
共 50 条
  • [21] Knee Derangements: Comparison of Isotropic 3D Fast Spin-Echo, Isotropic 3D Balanced Fast Field-Echo, and Conventional 2D Fast Spin-Echo MR Imaging
    Jung, Jin Young
    Yoon, Young Cheol
    Kim, Hye Rin
    Choe, Bong-Keun
    Wang, Joon Ho
    Jung, Jee Young
    RADIOLOGY, 2013, 268 (03) : 802 - 813
  • [22] Detecting brain gadolinium retention in MS with MRI: a comparison of 2D Spin Echo, 3D gradient echo, quantitative susceptibility mapping and magnetization transfer
    Cananau, C.
    Forslin, Y.
    Sjostrom, H.
    Bergendal, A.
    Fink, K.
    Wiberg, M. Kristoffersen
    Ouellette, R.
    Fredrikson, S.
    Granberg, T.
    MULTIPLE SCLEROSIS JOURNAL, 2021, 27 (2_SUPPL) : 408 - 409
  • [23] MR-IMAGING OF THE NORMAL MENINGES - COMPARISON OF CONTRAST-ENHANCEMENT PATTERNS ON 3D GRADIENT-ECHO AND SPIN-ECHO IMAGES
    FARN, JW
    MIROWITZ, SA
    AMERICAN JOURNAL OF ROENTGENOLOGY, 1994, 162 (01) : 131 - 135
  • [24] MRI of the Knee Ligaments and Menisci: Comparison of Isotropic-Resolution 3D and Conventional 2D Fast Spin-Echo Sequences at 3 T
    Subhas, Naveen
    Kao, Amy
    Freire, Maxime
    Polster, Joshua M.
    Obuchowski, Nancy A.
    Winalski, Carl S.
    AMERICAN JOURNAL OF ROENTGENOLOGY, 2011, 197 (02) : 442 - 450
  • [25] T2 quantification in brain using 3D fast spin-echo imaging with long echo trains
    Snyder, Jeff
    McPhee, Kelly C.
    Wilman, Alan H.
    MAGNETIC RESONANCE IN MEDICINE, 2022, 87 (05) : 2145 - 2160
  • [26] Comparison of 3D multi-echo gradient-echo and 2D T2*MR sequences for the detection of arterial thrombus in patients with acute stroke
    Hodel, Jerome
    Leclerc, Xavier
    Khaled, Wassef
    Tamazyan, Ruben
    Rodallec, Mathieu
    Gerber, Sophie
    Blanc, Raphael
    Benadjaoud, Mohamed
    Lambert, Oriane
    Rabrait, Cecile
    Zuber, Mathieu
    Rahmouni, Alain
    Zins, Marc
    EUROPEAN RADIOLOGY, 2014, 24 (03) : 762 - 769
  • [27] Comparison of 3D multi-echo gradient-echo and 2D T2* MR sequences for the detection of arterial thrombus in patients with acute stroke
    Jérôme Hodel
    Xavier Leclerc
    Wassef Khaled
    Ruben Tamazyan
    Mathieu Rodallec
    Sophie Gerber
    Raphael Blanc
    Mohamed Benadjaoud
    Oriane Lambert
    Cécile Rabrait
    Mathieu Zuber
    Alain Rahmouni
    Marc Zins
    European Radiology, 2014, 24 : 762 - 769
  • [28] Diagnostic Utility of 3D Gradient-Echo MR Imaging Sequences through the Filum Compared with Spin-Echo T1 in Children with Concern for Tethered Cord
    Rafiee, F.
    Mehan, W. A.
    Rincon, S.
    Rohatgi, S.
    Rapalino, O.
    Buch, K.
    AMERICAN JOURNAL OF NEURORADIOLOGY, 2023, 44 (03) : 323 - 327
  • [29] Advantages of a T1-Weighted Gradient-Recalled Echo (GRE) Sequence With a Radial 3D Sampling Approach Versus 2D Turbo Spin-Echo and Cartesian 3D GRE Sequences in Head and Neck MRI
    Fahlenkamp, Ute Lina
    Siepmann, Stefan
    Fehrenbach, Uli
    Thieme, Nadine
    Doellinger, Felix
    Hamm, Bernd
    Denecke, Timm
    AMERICAN JOURNAL OF ROENTGENOLOGY, 2020, 214 (04) : 747 - 752
  • [30] Direct MR arthrography of the shoulder: 2D vs. 3D gradient-echo imaging
    Wutke, R
    Fellner, FA
    Fellner, C
    Stangl, R
    Dobritz, M
    Bautz, WA
    MAGNETIC RESONANCE IMAGING, 2001, 19 (09) : 1183 - 1191