Statistical analysis of absolute photopeak efficiency of clover HPGe detectors

被引:0
|
作者
Dey, P. [1 ]
Palit, R. [1 ]
Das, Biswajit [1 ]
Kundu, A. [1 ]
Malik, Vishal [1 ]
机构
[1] Tata Inst Fundamental Res, Dept Nucl & Atom Phys, Mumbai 400005, India
关键词
Absolute photopeak efficiency; y-ray detectors; Machine learning method; Bayesian analysis; Markov chain Monte Carlo method; NUCLEAR-DATA SHEETS; GE(LI) DETECTOR; P VALUE;
D O I
10.1016/j.nima.2024.170129
中图分类号
TH7 [仪器、仪表];
学科分类号
0804 ; 080401 ; 081102 ;
摘要
The variation of absolute photopeak efficiency of an array of clover HPGe detectors as a function of y-ray energy has been obtained using statistical model analysis. Machine learning technique has been implemented for determining the energy-dependent polynomial of photopeak efficiency. The distribution of parameters of the best-fit model has been deduced by incorporating Markov chain Monte Carlo method working with the basic framework of Bayesian analysis which also provides the correlation between the parameters. This method has the advantage of calculating the posterior probability distribution by maximizing the likelihood function of the experimental data over standard least-squares method. Excellent agreement among the results procured from machine learning, Markov chain Monte Carlo, and least-squares method has been achieved and the correct prediction of the nature of the variation extrapolation region as well as interpolation region has been discussed.
引用
收藏
页数:10
相关论文
共 50 条
  • [1] ANALYSIS OF HPGE-DETECTOR RELATIVE PHOTOPEAK EFFICIENCY
    ELKHOSHT, MA
    NUOVO CIMENTO DELLA SOCIETA ITALIANA DI FISICA A-NUCLEI PARTICLES AND FIELDS, 1995, 108 (06): : 659 - 668
  • [2] Photopeak Efficiency and Coincidence Summing Factors for HPGe Detectors Using Bulk Sources
    Alharshan, Gharam A.
    HEALTH PHYSICS, 2020, 118 (04): : 396 - 401
  • [3] ABSOLUTE LOW-ENERGY PHOTOPEAK EFFICIENCY VALUES FOR NAI(TL) AND SI(LI) DETECTORS
    RIEPPO, R
    PHYSICS IN MEDICINE AND BIOLOGY, 1985, 30 (08): : 839 - 844
  • [4] METHOD FOR CALCULATION OF ABSOLUTE PHOTOPEAK EFFICIENCY VALUES OF NAI-DETECTORS WITH WATER AS MEDIUM IN SOURCE
    RIEPPO, R
    INTERNATIONAL JOURNAL OF APPLIED RADIATION AND ISOTOPES, 1978, 29 (02): : 65 - 67
  • [5] A software for simulation of efficiency of HPGe detectors
    Khiem L.H.
    Trong T.D.
    Physics of Particles and Nuclei Letters, 2015, 12 (3) : 398 - 399
  • [6] ON THE PHOTOPEAK EFFICIENCY OF GERMANIUM GAMMA-RAY DETECTORS
    JACKEL, B
    WESTMEIER, W
    PATZELT, P
    NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 1987, 261 (03): : 543 - 548
  • [7] Statistical pileup correction method for HPGe detectors
    Trigano, Thomas
    Souloumiac, Antoine
    Montagu, Thierry
    Roueff, Francois
    Moulines, Eric
    IEEE TRANSACTIONS ON SIGNAL PROCESSING, 2007, 55 (10) : 4871 - 4881
  • [8] Efficiency of segmented HPGe detectors: Design criteria for pulse shape analysis
    Wieland, O
    Camera, F
    Million, B
    Bracco, A
    Pignanelli, M
    Ripamonti, G
    Geraci, A
    Van der Marel, J
    IEEE TRANSACTIONS ON NUCLEAR SCIENCE, 2001, 48 (03) : 296 - 301
  • [9] EMPIRICAL EFFICIENCY AND VOLUME RELATIONSHIPS FOR HPGE DETECTORS
    OWENS, A
    PASCARELLE, SM
    GEHRELS, N
    TEEGARDEN, BJ
    NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 1991, 310 (03): : 681 - 684
  • [10] New Approach For Calibration The Efficiency Of HpGe Detectors
    Alnour, I. A.
    Wagiran, H.
    Ibrahim, N.
    Hamzah, S.
    Siong, W. B.
    Elias, M. S.
    ADVANCING NUCLEAR RESEARCH AND ENERGY DEVELOPMENT, 2014, 1584 : 38 - 44