Validation of Noninvasive Derivation of the Central Aortic Pressure Waveform from Fingertip Photoplethysmography Using a Novel Selective Transfer Function Method

被引:0
|
作者
Cox, James R. [1 ]
Akeila, Ehad [2 ]
Avolio, Alberto P. [1 ,2 ]
Butlin, Mark [1 ]
Liao, Catherine [2 ]
Bentley, Gisele J. [1 ]
Qasem, Ahmad [1 ,2 ]
机构
[1] Macquarie Univ, Macquarie Med Sch, Fac Med Hlth & Human Sci, Sydney, NSW, Australia
[2] Cardiex Ltd, Sydney, NSW, Australia
关键词
Photoplethysmography; Transfer function; Waveform analysis; Aortic pressure; Blood pressure; BLOOD-PRESSURE; ARTERIAL STIFFNESS; AUGMENTATION INDEX; RISK; REFLECTIONS;
D O I
10.1159/000540666
中图分类号
R6 [外科学];
学科分类号
1002 ; 100210 ;
摘要
Introduction: Central aortic pressure waveform analyses can provide clinically relevant information beyond conventional brachial blood pressure (BP) assessment. This waveform can be reproduced noninvasively through application of a generalized transfer function (GTF) on a peripheral waveform, as conventionally performed by applanation tonometry. Photoplethysmography (PPG) is an alternate approach; however, differences in measurement site and modality demand the use of a transfer function (TF) specific for those differences. This study aimed to compare central aortic waveform features generated from radial tonometry (reference) using a proprietary GTF with a central aortic waveform and its features generated from a simultaneous fi ngertip PPG measurement using a selective method where one of three different TFs is chosen based on the input signal harmonic profile. Methods: Brachial BP was measured in triplicate under resting conditions and was used for subsequent calibration. Multiple simultaneous radial tonometry (SphygmoCor CVMS) and fi ngertip PPG measurements were then performed in individual participants (n = 21, 10 females, age: 39 +/- 16 years). Measurements were converted into central aortic waveforms with their respective TFs. Twenty central aortic pressure waveform parameters were compared through correlation analysis, Bland-Altman plots, and a repeated measure mixed-effects ANOVA model. Central aortic waveform shape was compared using the root-mean-squared error (RMSE). Results: Correlation (r) of PPG-derived parameters with radially tonometry-derived central aortic parameters was high ranging from 0.79 to 0.99. Mean differences of pressure-related parameters were within 1.3 mm Hg, and differences of time-related parameters ranged from - 2.2 to 3.4%. While some parameters were statistically different, these differences are not physiologically meaningful. Central aortic waveform shape had an average RMSE of 1.8 +/- 0.9%. Conclusion: Fingertip PPG-derived central aortic waveform parameters using a novel selective TF were comparable to central aortic waveform features derived from radial tonometry using a previously validated GTF. (c) 2024 The Author(s). Published by S. Karger AG, Basel
引用
收藏
页码:95 / 105
页数:11
相关论文
共 50 条
  • [21] Impact of radial artery pressure waveform calibration on estimated central pressure using a transfer function approach
    Segers, Patrick
    Mahieu, Dries
    Rietzschel, Ernst R.
    De Buyzere, Marc L.
    Van Bortel, Luc M.
    HYPERTENSION, 2008, 52 (03) : E24 - E25
  • [22] Reconstruction of the aortic pressure waveform using a two-level adaptive transfer function strategy
    Du, Shuo
    Liu, Wenyan
    Yao, Yang
    Sun, Guozhe
    He, Ying
    Alastruey, Jordi
    Xu, Lisheng
    Yao, Yudong
    Qian, Wei
    Measurement: Journal of the International Measurement Confederation, 2022, 204
  • [23] Reconstruction of the aortic pressure waveform using a two-level adaptive transfer function strategy
    Du, Shuo
    Liu, Wenyan
    Yao, Yang
    Sun, Guozhe
    He, Ying
    Alastruey, Jordi
    Xu, Lisheng
    Yao, Yudong
    Qian, Wei
    MEASUREMENT, 2022, 204
  • [24] Use of a Chebyshev filter as an inverse transfer function for derivation of central from the prepheral pressure
    Qasem, A
    Avolio, A
    Frangakis, G
    PROCEEDINGS OF THE 23RD ANNUAL INTERNATIONAL CONFERENCE OF THE IEEE ENGINEERING IN MEDICINE AND BIOLOGY SOCIETY, VOLS 1-4: BUILDING NEW BRIDGES AT THE FRONTIERS OF ENGINEERING AND MEDICINE, 2001, 23 : 2132 - 2134
  • [25] Comparison of estimates of central blood pressure obtained from the late systolic shoulder of the radial pressure waveform and from the SphygmoCor radial-to-aortic transfer function
    Kamalesh, T.
    Chowienczyk, P.
    JOURNAL OF HYPERTENSION, 2008, 26 : S509 - S509
  • [26] Gender specific and generalized arterial transfer functions in the derivation of aortic waveform characteristics from radial blood pressure waves
    Hope, SA
    Tay, DB
    Meredith, IT
    Cameron, JD
    JOURNAL OF HYPERTENSION, 2002, 20 : S183 - S183
  • [27] Impact of radial artery pressure waveform calibration on estimated central pressure using a transfer function approach - Response
    McEniery, Carmel M.
    Cockcroft, John R.
    Wilkinson, Ian B.
    HYPERTENSION, 2008, 52 (03) : E26 - E26
  • [28] A New Approach to Reconstruction of Central Aortic Blood Pressure Using "Adaptive" Transfer Function
    Hahn, Jin-Oh
    Asada, H. Harry
    Reisner, Andrew T.
    Jaffer, Farouc A.
    2008 30TH ANNUAL INTERNATIONAL CONFERENCE OF THE IEEE ENGINEERING IN MEDICINE AND BIOLOGY SOCIETY, VOLS 1-8, 2008, : 813 - +
  • [29] Validation of a novel method to determine non-invasively the rate of central aortic pressure changes
    Gorenberg, M.
    Marmor, A.
    Journal of Medical Engineering and Technology, 2008, 32 (04): : 257 - 262
  • [30] A novel method to directly derive central aortic systolic pressure (CASP) from the radial artery pressure waveform - Validation with data from the CAFE study (vol 26, pg S40, 2008)
    Williams, B.
    Lacy, P.
    Ting, C. M.
    Chua, N. H.
    Chen, L.
    JOURNAL OF HYPERTENSION, 2011, 29 (01) : 177 - 177