Is thoron a problem in radon measurements with NRPB/SSI passive radon detectors-experimental study

被引:0
|
作者
Wasikiewicz, J. M. [1 ]
Dimitrova, I [2 ]
Ibrahimi, Z-f [1 ]
Georgiev, S. [2 ]
Mitev, K. [2 ]
Daraktchieva, Z. [1 ]
机构
[1] UK Hlth Secur Agcy, RCCE, Didcot OX11 0RQ, Oxon, England
[2] Sofia Univ St Kliment Ohridski, Fac Phys, Sofia 1164, Bulgaria
关键词
ATMOSPHERE; EXPOSURE;
D O I
10.1093/rpd/ncaf012
中图分类号
X [环境科学、安全科学];
学科分类号
08 ; 0830 ;
摘要
The UKHSA radon detectors with polyallyldiglycol carbonate sensing material produced from 2010 onward were used to confirm that they are purely radon detectors only with negligible interference from thoron presence. The statistical analysis of results, after following standard detector processing procedures, showed that such devices can only detect 2.4% +/- 0.3% of the total thoron activity concentration compared to the calibrated reference device. It was also demonstrated that thoron, unlike radon, can only travel a very short distance from the source to be effectively measured. Based on the above, it has been concluded that interference of thoron in radon measurement is negligible (within the statistical measurement error), and hence can be ignored in standard indoor and outdoor measurements.
引用
收藏
页数:7
相关论文
共 50 条
  • [11] Sensitivity to thoron of an SSNTD-based passive radon measuring device: Experimental evaluation and implications for radon concentration measurements and risk assessment
    Bochicchio, F.
    Ampollini, M.
    Tommasino, L.
    Sorimachi, A.
    Tokonami, S.
    RADIATION MEASUREMENTS, 2009, 44 (9-10) : 1024 - 1027
  • [12] Radon and thoron parallel measurements in Hungary
    Kavasi, N.
    Nemeth, Cs.
    Kovacs, T.
    Tokonami, S.
    Jobbagy, V.
    Varhegyi, A.
    Gorjanacz, Z.
    Vigh, T.
    Somlai, J.
    RADIATION PROTECTION DOSIMETRY, 2007, 123 (02) : 250 - 253
  • [13] THORON CONTRIBUTIONS IN RADON MEASUREMENTS IN THE ENVIRONMENT
    SHWEIKANI, R
    DURRANI, SA
    RADIATION MEASUREMENTS, 1995, 25 (1-4) : 615 - 616
  • [14] Thoron contributions to radon measurements in the environment
    Shweikani, R.
    Durrani, S.A.
    Radiation Measurements, 25 (1-4):
  • [15] INDOOR THORON AND RADON PROGENY MEASUREMENTS
    TU, KW
    GEORGE, AC
    LOWDER, WM
    GOGOLAK, CV
    RADIATION PROTECTION DOSIMETRY, 1992, 45 (1-4) : 557 - 560
  • [16] THORON AND RADON MEASUREMENTS IN ROMANIAN SCHOOLS
    Burghele, B. D.
    Cosma, C.
    RADIATION PROTECTION DOSIMETRY, 2012, 152 (1-3) : 38 - 41
  • [17] Passive methods for environmental radon and thoron monitoring
    Iida, T
    RADON AND THORON IN THE HUMAN ENVIRONMENT, 1998, : 18 - 27
  • [18] The analysis of results of radon/thoron measurements performed with the use of nuclear track detectors
    Chalupnik, Stanislaw
    JOURNAL OF RADIOLOGICAL PROTECTION, 2020, 40 (04) : N39 - N45
  • [19] Experimental study of radon and thoron diffusion through barriers
    Durcik, M
    Havlik, F
    JOURNAL OF RADIOANALYTICAL AND NUCLEAR CHEMISTRY-ARTICLES, 1996, 209 (02): : 307 - 313
  • [20] Comparative analysis of radon, thoron and thoron progeny concentration measurements
    Janik, Miroslaw
    Tokonami, Shinji
    Kranrod, Chutima
    Sorimachi, Atsuyuki
    Ishikawa, Tetsuo
    Hosoda, Masahiro
    Mclaughlin, James
    Chang, Byung-Uck
    Kim, Yong Jae
    JOURNAL OF RADIATION RESEARCH, 2013, 54 (04) : 597 - 610