Design and operation of a whole-body MRI scanner without RF shielding

被引:0
|
作者
Biber, Stephan [1 ]
Kannengiesser, Stephan [1 ]
Nistler, Juergen [1 ]
Braun, Markus [1 ]
Blaess, Stefan [1 ]
Gebhardt, Matthias [1 ]
Grodzki, David [1 ]
Ritter, Dieter [1 ]
Seegerer, Georg [1 ]
Vester, Markus [1 ]
Schneider, Rainer [1 ]
机构
[1] Siemens Healthineers AG, Allee Roethelheimpk 2, D-91052 Erlangen, Germany
关键词
accessible MRI; noise cancellation; RF shielding; whole-body MRI;
D O I
暂无
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Purpose: To investigate how a clinical MRI system can be operated without an RF cabin. Receive interference cancellation via signal processing and active transmit emission suppression are evaluated for mitigating image artifacts and achieving electromagnetic compatibility. Methods: A clinical whole-body MR scanner with B-0=0.55T was placed in an environment without RF shielding. The signals of 16 additional auxiliary receive antenna elements outside the scanner bore were used to remove interference from the wanted MR signal via a dedicated sidelobe cancellation algorithm. In vivo measurements with different MR sequences were performed in the presence of ambient RF noise and additional artificial interferers, covering brain and abdominal applications. To investigate the transmit electromagnetic compatibility, simulations with different auxiliary transmit antenna arrangements and body coil loading conditions were performed, and their emission suppression performance in the far field was evaluated. Results: The MR sidelobe cancellation algorithm successfully removed interference artifacts up to more than 3 times of the wanted MR image energy; difference images carried little to no visibly removed wanted signal. According to the simulations, electromagnetic compliance can be achieved with 32 or more auxiliary transmit antennas. The number and placement of sensor points are essential. Conclusions: A whole-body 0.55T MR scanner can be operated without RF shielding. Image artifacts from external interference can be removed, requiring only a modest number of auxiliary antennas. Electromagnetic compliance appears to be achievable using active transmit cancellation, but the required hardware efforts may not be practical.
引用
收藏
页码:1842 / 1855
页数:14
相关论文
共 50 条
  • [1] Design and operation of a whole-body MRI scanner without RF shielding
    Biber, Stephan
    Kannengiesser, Stephan
    Nistler, Juergen
    Braun, Markus
    Blaess, Stefan
    Gebhardt, Matthias
    Grodzki, David
    Ritter, Dieter
    Seegerer, Georg
    Vester, Markus
    Schneider, Rainer
    MAGNETIC RESONANCE IN MEDICINE, 2024,
  • [3] An RF power amplifier for whole-body MRI scanner applications
    不详
    MICROWAVE JOURNAL, 1999, 42 (02) : 168 - +
  • [4] WHOLE-BODY COUNTING WITHOUT MASSIVE SHIELDING
    RAHOLA, T
    TILLANDER, M
    HEALTH PHYSICS, 1975, 28 (05): : 640 - 640
  • [5] A NEW WHOLE-BODY COUNTER SHIELDING DESIGN
    POULHEIM, KF
    HOESSELBARTH, H
    HEALTH PHYSICS, 1965, 11 (01): : 52 - +
  • [6] CLINICAL WHOLE-BODY COUNTING - WHOLE-BODY SCANNER WITH 2 CRYSTALS
    REIZENSTEIN, P
    KARLSSON, HA
    ACTA RADIOLOGICA-THERAPY PHYSICS BIOLOGY, 1966, 4 (03): : 209 - +
  • [7] Whole-Body Overdiagnosis: Perils of Whole-Body MRI
    Jha, Saurabh
    JOURNAL OF THE AMERICAN COLLEGE OF RADIOLOGY, 2024, 21 (11) : 1819 - 1821
  • [8] NEW ASPECTS IN THE DESIGN OF COMPUTERIZED DIGITAL WHOLE-BODY SCANNER
    LESJAK, M
    RUPNIK, Z
    SNAJDER, J
    KRALJI, H
    KOLBAS, J
    EUROPEAN JOURNAL OF NUCLEAR MEDICINE, 1979, 4 (02): : 146 - 146
  • [9] Design and assembly of an 8 tesla whole-body MR scanner
    Robitaille, PML
    Warner, R
    Jagadeesh, J
    Abduljalil, AM
    Kangarlu, A
    Burgess, RE
    Yu, Y
    Yang, LN
    Zhu, H
    Jiang, ZC
    Bailey, RE
    Chung, W
    Somawiharja, Y
    Feynan, P
    Rayner, DL
    JOURNAL OF COMPUTER ASSISTED TOMOGRAPHY, 1999, 23 (06) : 808 - 820
  • [10] Design and assembly of an 8 Tesla whole-body MR scanner
    Department of Radiology, Ohio State University, Columbus, OH, United States
    不详
    不详
    不详
    不详
    不详
    J. Comput. Assisted Tomogr., 6 (808-820):