Efficacy and safety of a novel polytetrafluoroethylene-coated self-expandable metal stent for distal malignant biliary obstruction

被引:0
|
作者
Nakagawa, Hiroki [1 ,2 ]
Takeda, Tsuyoshi [1 ]
Okamoto, Takeshi [1 ]
Hirai, Tatsuki [1 ]
Mie, Takafumi [1 ]
Furukawa, Takaaki [1 ]
Kasuga, Akiyoshi [1 ]
Sasaki, Takashi [1 ]
Ozaka, Masato [1 ]
Matsuda, Takahisa [2 ]
Igarashi, Yoshinori [2 ]
Sasahira, Naoki [1 ]
机构
[1] Canc Inst Hosp Japanese Fdn Canc Res, Dept Hepatobiliary Pancreat Med, 3-8-31 Ariake, Koto, Tokyo 1358550, Japan
[2] Toho Univ, Omori Med Ctr, Dept Internal Med, Div Gastroenterol & Hepatol, Tokyo, Japan
来源
DEN OPEN | 2025年 / 5卷 / 01期
关键词
adverse event; distal malignant biliary obstruction; polytetrafluoroethylene; recurrent biliary obstruction; self-expandable metal stent; RISK-FACTORS; MIGRATION;
D O I
10.1002/deo2.70010
中图分类号
R57 [消化系及腹部疾病];
学科分类号
摘要
Background Stent migration and sludge formation remain significant problems associated with covered self-expandable metal stents (CSEMSs). The EGIS biliary stent fully covered flare type (EGIS biliary stent), a new type of polytetrafluoroethylene-coated self-expandable metal stent with low axial force and an anti-migration system, was developed to overcome these disadvantages. We conducted this study to evaluate the efficacy and safety of this stent in comparison with conventional CSEMS (c-CSEMS). Methods We retrospectively analyzed consecutive patients with unresectable pancreatic cancer who received initial CSEMS for distal malignant biliary obstruction. The primary outcome was time to recurrent biliary obstruction (RBO). Secondary outcomes included technical success rate, functional success rate, stent-related adverse events, causes of RBO, and re-intervention. Results A total of 40 patients were included (EGIS group: 20; c-CSEMS group: 20). The technical and functional success rates were similar between the two groups. Stent-related adverse event rates (20% vs. 15%, p > 0.99) and overall RBO rates (56% vs. 50%, p > 0.99) were not significantly different between the two groups. Stent migration was the most common cause of RBO in the EGIS group, while stent occlusion was in the c-CSEMS group. The median time to RBO (102 vs. 434 days, p = 0.10) was not significantly different between the two groups. Endoscopic transpapillary re-intervention was successful in most patients in both groups. Conclusions The EGIS biliary stent was not associated with a longer time to RBO compared to c-CSEMS. Further improvements, especially against stent migration, are needed to improve its efficacy.
引用
收藏
页数:8
相关论文
共 50 条
  • [21] Self-expandable metallic stents for malignant biliary obstruction: Efficacy on proximal and distal tumors
    Chen, Jui-Hao
    Sun, Cheuk-Kay
    Liao, Chao-Sheng
    Chua, Chain-Smoke
    WORLD JOURNAL OF GASTROENTEROLOGY, 2006, 12 (01) : 119 - 122
  • [22] COVERED VERSUS UNCOVERED SELF-EXPANDABLE METAL STENTS FOR DISTAL MALIGNANT BILIARY OBSTRUCTION
    Upreti, Alina
    Madala, Seetharamprasad
    GASTROINTESTINAL ENDOSCOPY, 2024, 99 (06) : AB673 - AB673
  • [23] Efficacy of multi-hole self-expandable metal stent compared to fully covered and uncovered self-expandable metal stents in patients with unresectable malignant distal biliary obstruction: a propensity analysis
    Santi Kulpatcharapong
    Panida Piyachaturawat
    Parit Mekaroonkamol
    Phonthep Angsuwatcharakon
    Wiriyaporn Ridtitid
    Pradermchai Kongkam
    Rungsun Rerknimitr
    Surgical Endoscopy, 2024, 38 : 212 - 221
  • [24] Efficacy of multi-hole self-expandable metal stent compared to fully covered and uncovered self-expandable metal stents in patients with unresectable malignant distal biliary obstruction: a propensity analysis
    Kulpatcharapong, Santi
    Piyachaturawat, Panida
    Mekaroonkamol, Parit
    Angsuwatcharakon, Phonthep
    Ridtitid, Wiriyaporn
    Kongkam, Pradermchai
    Rerknimitr, Rungsun
    SURGICAL ENDOSCOPY AND OTHER INTERVENTIONAL TECHNIQUES, 2023, 37 (12): : 9062 - 9069
  • [25] Biliary self-expandable metallic stent for unresectable malignant distal biliary obstruction: Which is better: covered or uncovered?
    Isayama, Hiroyuki
    Nakai, Yousuke
    Kogure, Hirofumi
    Yamamoto, Natsuyo
    Koike, Kazuhiko
    DIGESTIVE ENDOSCOPY, 2013, 25 : 71 - 74
  • [26] Nitinol versus steel partially covered self-expandable metal stent for malignant distal biliary obstruction: a randomized trial
    Soderlund, Claes
    Linder, Stefan
    Bergenzaun, Per E.
    Grape, Tomas
    Hakansson, Hans-Olof
    Kilander, Anders
    Lindell, Gert
    Ljungman, Martin
    Ohlin, Bo
    Nielsen, Jorgen
    Rudberg, Claes
    Stotzer, Per-Ove
    Svartholm, Erik
    Toth, Ervin
    Frozanpor, Farshad
    ENDOSCOPY, 2014, 46 (11) : 941 - 948
  • [27] Covered self-expandable metallic stents for distal malignant biliary obstruction
    Tsujino, Takeshi
    Isayama, Hiroyuki
    Togawa, Osamu
    Nakai, Yousuke
    Omata, Masao
    JOURNAL OF GASTROENTEROLOGY AND HEPATOLOGY, 2007, 22 : A84 - A84
  • [28] EFFICACY OF MULTI-HOLE STENT COMPARE TO FULLY COVERED AND UNCOVERED SELF-EXPANDABLE METAL STENTS FOR THE TREATMENT OF MALIGNANT DISTAL BILIARY OBSTRUCTION
    Kulpatcharapong, Santi
    Piyachaturawat, Panida
    Ridtitid, Wiriyaporn
    Kongkam, Pradermchai
    Rerknimitr, Rungsun
    GASTROINTESTINAL ENDOSCOPY, 2019, 89 (06) : AB236 - AB236
  • [29] Novel self-expandable metallic stent with dumbbell-shape and spiral outer cover for malignant distal biliary obstruction
    Miwa, Haruo
    Ishino, Yugo
    Tsunoda, Shotaro
    Endo, Kazuki
    Oishi, Ritsuko
    Suzuki, Yuichi
    Maeda, Shin
    ENDOSCOPY, 2024, 56 : E849 - E850
  • [30] Efficacy and safety of winged partially covered self-expandable metal stent for malignant gastric outlet obstruction: Newly designed stent to prevent distal migration
    Choi, Young Kwon
    Ahn, Ji Yong
    Na, Hee Kyong
    Jung, Kee Wook
    Kim, Do Hoon
    Lee, Jeong Hoon
    Choi, Kee Don
    Song, Ho June
    Lee, Gin Hyug
    Jung, Hwoon-Yong
    JOURNAL OF GASTROENTEROLOGY AND HEPATOLOGY, 2018, 33 : 264 - 264