Assessment of scattered and leakage radiation from ultra-portable X-ray systems in chest imaging: An independent study

被引:0
|
作者
Paulis, Leonie E. [1 ,2 ,3 ]
Schnerr, Roald S. [1 ,4 ]
Halton, Jarred [1 ]
Qin, Zhi Zhen [2 ]
Chua, Arlene [5 ]
机构
[1] Medecins Sans Frontieres Int, Amsterdam, Netherlands
[2] Dept Digital Hlth, Amsterdam, Netherlands
[3] Maxima Med Ctr, Dept Med Phys, Veldhoven, Netherlands
[4] Maastricht Univ, Dept Radiol & Nucl Med, Med Ctr, Maastricht, Netherlands
[5] Medecins Sans Frontieres Int, Geneva, Switzerland
来源
PLOS GLOBAL PUBLIC HEALTH | 2025年 / 5卷 / 01期
关键词
D O I
10.1371/journal.pgph.0003986
中图分类号
R1 [预防医学、卫生学];
学科分类号
1004 ; 120402 ;
摘要
Ultraportable (UP) X-ray devices are ideal to use in community-based settings, particularly for chest X-ray (CXR) screening of tuberculosis (TB). Unfortunately, there is insufficient guidance on the radiation safety of these devices. This study aims to determine the radiation dose by UP X-ray devices to both the public and radiographers compared to international dose limits. Radiation dose measurements were performed with four UP X-ray devices that met international criteria, utilizing a clinically representative CXR set-up made with a thorax phantom. Scatter and leakage radiation dose were measured at various positions surrounding the phantom and X-ray tube, respectively. These measurements were used to calculate yearly radiation doses for different scenarios based on the median of all UP X-ray devices. From the yearly scatter doses, the minimum distances from the phantom needed to stay below the international public dose limit (1 mSv/year) were calculated. This distance was longest in the direction back towards the X-ray tube and shortest to the left/right sides of the phantom, e.g., 4.5 m and 2.5 m resp. when performing 50 exams/day, at 90kV, 2.5 mAs and source skin distance (SSD) 1 m. Additional calculations including leakage radiation were conducted at a typical radiographer position (i.e., behind the X-ray tube), with a correction factor for wearing a lead apron. At 2 m behind the X-ray tube, a radiographer wearing a lead apron could perform 106 exams/day at 2.5 mAs and 29 exams/day at 10 mAs (90kV, SSD 1 m), while keeping his/her radiation dose below the public dose limit (1 mSv/year) and well below the radiographer dose limit (20 mSv/year). In most CXR screening scenarios, the radiation dose of UP X-ray devices can be kept below 1 mSv/year by employing basic radiation safety rules on time, distance and shielding and using appropriate CXR exposure parameters.
引用
收藏
页数:15
相关论文
共 50 条
  • [21] X-RAY INTERFERENCE STRUCTURE IN SCATTERED RADIATION FROM BARIUM STEARATE MULTILAYER FILMS
    KAPP, DS
    WAINFAN, N
    PHYSICAL REVIEW, 1965, 138 (5A): : 1490 - &
  • [22] Assessment of scattered radiation from hand-held dental x-ray equipment using the Monte Carlo method
    Batista, Wilson Otto
    Soares, Maria Rosangela
    Gurjao Rios, Jacqueline Machado
    Dos Santos Souza, Antonio Carlos
    Pinheiro, Isabelle Matos
    Javier Vega Ramirez, Jose Luis
    Caldas, Linda V. E.
    JOURNAL OF RADIOLOGICAL PROTECTION, 2021, 41 (04) : 654 - 668
  • [23] A study on electron density imaging using the Compton scattered X-ray CT technique
    Masuji, Ryota
    Watanabe, Kenichi
    Yamazaki, Atsushi
    Uritani, Akira
    NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 2011, 652 (01): : 620 - 624
  • [24] Personnel exposure to scattered radiation during radiography of the distal interphalangeal joint in the horse using a portable X-ray machine
    Sterc, J.
    Lepkova, R.
    ACTA VETERINARIA BRNO, 2007, 76 (01) : 105 - +
  • [25] A Monte Carlo study of the energy spectra and transmission characteristics of scattered radiation from x-ray computed tomography
    Platten, David John
    JOURNAL OF RADIOLOGICAL PROTECTION, 2014, 34 (02) : 445 - 456
  • [26] Using Natural Language Processing for Extracting Information from Portable Chest X-Ray Reports
    Wang, D. Y.
    Hwang, T. S.
    Rubin, D.
    Chambers, J.
    South, B. R.
    Goldstein, M. K.
    JOURNAL OF THE AMERICAN GERIATRICS SOCIETY, 2013, 61 : S103 - S103
  • [27] DUAL-ENERGY X-RAY PROJECTION IMAGING - 2 SAMPLING SCHEMES FOR THE CORRECTION OF SCATTERED RADIATION
    WAGNER, FC
    MACOVSKI, A
    NISHIMURA, DG
    MEDICAL PHYSICS, 1988, 15 (05) : 732 - 748
  • [28] A STUDY OF THE SHIELDING USED TO REDUCE LEAKAGE AND SCATTERED RADIATION TO THE FETUS IN A PREGNANT PATIENT TREATED WITH A 6-MV EXTERNAL X-RAY BEAM
    Han, Bin
    Bednarz, Bryan
    Xu, X. George
    HEALTH PHYSICS, 2009, 97 (06): : 581 - 589
  • [29] A DISTANCE TECHNIC FOR MINIMIZING SCATTERED RADIATION IN X-RAY IMAGES WITH SPECIAL REFERENCE TO HIGH-VOLTAGE CHEST RADIOLOGY
    GAJEWSKI, H
    MULLER, D
    STRAHLENTHERAPIE, 1969, 138 (06) : 689 - &
  • [30] X-Ray Acoustic Imaging: Towards Radiation Dose Assessment During Teletherapy
    Sampaio, D. R. T.
    Martins, A. A. F.
    Carneiro, A. A. O.
    Pavoni, J. F.
    Pavan, T. Z.
    MEDICAL PHYSICS, 2017, 44 (06) : 3201 - 3201