In vitro biomechanical evaluation of a strutted intradiscal spacer for lumbar discectomy

被引:0
|
作者
Schmoelz, Werner [1 ]
Spicher, Anna [1 ]
Lindtner, Richard [1 ]
Hoermann, Romed [2 ]
Srour, Robin [3 ]
机构
[1] Med Univ Innsbruck, Dept Orthoped & Traumatol, Anichstr 35, A-6020 Innsbruck, Austria
[2] Med Univ Innsbruck, Dept Anat Histol & Embryol, Div Clin & Funct Anat, Mullerstr 59, A-6020 Innsbruck, Austria
[3] Hop Civils Colmar, Dept Neurosurg, Colmar, France
关键词
Lumbar disc herniation; Intradiscal spacer; Disc repair; Cadaver study; Reherniation; DISC HERNIATION; SURGERY; INSTRUMENTATION; REHERNIATION; GUIDELINE; DISEASE; HEIGHT; FUSION; COHORT; RISK;
D O I
10.1016/j.clinbiomech.2025.106491
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
Background: Discectomy plus implantation of a strutted intradiscal spacer has been shown to reduce reoperations and reherniations versus discectomy alone following lumbar disc herniation. This study explored the underlying biomechanics of the intradiscal spacer. Methods: Six fresh-frozen cadaveric lumbar spine specimens (L2 to L5) from three donors were used. Following box annulotomy to simulate disc herniation, a discectomy was performed. One segment from each donor was randomly assigned to either an untreated control group or the test group where an intradiscal spacer was implanted. A six degree of freedom universal spine tester assessed range of motion (RoM) in flexion/extension, lateral bending and axial rotation in the native state in load controlled [f7.5 Nm] and intervals up to 60,000 cycles. Disc height was measured on fluoroscopy for multiple timepoints. The segments were also analyzed to detect possible reherniation during the cycling loading. Findings: Following 60,000 cycles, the mean percentage RoM increase versus the intact state was less for discectomy plus the intradiscal spacer versus discectomy alone for lateral bending (170.7 f 10.0 vs. 222.5 f 33.3 %), flexion/extension (178.5 f 6.1 vs. 204.6 f 44.3 %) and axial rotation (284.4 f 127.2 vs. 362.3 % f 240.4 %). Mean overall disc height loss versus the annulotomy state was also less with the intradiscal spacer versus discectomy alone (-19.3 f 3.7 vs. -29.1 f 6.1 %). There was no evidence of device subsidence or migration. Interpretation: This study helps to explain the clinical observation that insertion of a strutted intradiscal spacer following discectomy reduces reherniation rate by mechanically limiting the increase in RoM and disc height loss following lumbar discectomy.
引用
收藏
页数:6
相关论文
共 50 条
  • [31] In vitro biomechanical evaluation of a monocoque plate-spacer construct for cervical open-door laminoplasty
    Ohara, Yukoh
    Hara, Takeshi
    Orias, Alejandro A. Espinoza
    Tani, Satoshi
    Inoue, Nozomu
    Mizuno, Junichi
    PLOS ONE, 2018, 13 (10):
  • [32] Percutaneous endoscopic lumbar discectomy and interbody fusion with B-Twin expandable spinal spacer
    Yao, Nuzhao
    Wang, Wenjun
    Liu, Yonghua
    ARCHIVES OF ORTHOPAEDIC AND TRAUMA SURGERY, 2011, 131 (06) : 791 - 796
  • [33] Biomechanical evaluation of posterior thoracic transpedicular discectomy Laboratory investigation
    Deniz, Fatih Ersay
    Brasiliense, Leonardo B. C.
    Lazaro, Bruno C. R.
    Reyes, Phillip M.
    Sawa, Anna G. U.
    Sonntag, Volker K. H.
    Crawford, Neil R.
    JOURNAL OF NEUROSURGERY-SPINE, 2010, 13 (02) : 253 - 259
  • [34] Percutaneous endoscopic lumbar discectomy and interbody fusion with B-Twin expandable spinal spacer
    Nuzhao Yao
    Wenjun Wang
    Yonghua Liu
    Archives of Orthopaedic and Trauma Surgery, 2011, 131 : 791 - 796
  • [35] An in vitro study examining a novel suction curette device for lumbar discectomy compared with standard manual discectomy
    Lavelle, William F.
    Ordway, Nathaniel R.
    Araghi, Ali
    Buckley, Rudolph A.
    Fayyazi, Amir H.
    JOURNAL OF NEUROSURGERY-SPINE, 2017, 26 (04) : 454 - 458
  • [36] Lateral translation of the lumbar spine: In vitro biomechanical study
    Okushima, Yuichiro
    Yamazaki, Nobutoshi
    Matsumoto, Morio
    Chiba, Kazuhiro
    Nagura, Takeo
    Toyama, Yoshiaki
    JOURNAL OF APPLIED BIOMECHANICS, 2006, 22 (02) : 83 - 92
  • [37] Biomechanical evaluation of human lumbar spine in spondylolisthesis
    Joszko, Kamil
    Gzik, Marek
    Wolanski, Wojciech
    Gzik-Zroska, Bozena
    Kawlewska, Edyta
    JOURNAL OF APPLIED BIOMEDICINE, 2018, 16 (01) : 51 - 58
  • [38] SACRAL AND LUMBAR PEDICLE FIXATION - A BIOMECHANICAL EVALUATION
    JACOBS, RR
    JAUCH, EC
    JACOBS, CR
    MONTESANO, PX
    JOURNAL OF BIOMECHANICS, 1987, 20 (09) : 905 - 905
  • [39] Biomechanical Analysis of Allograft Spacer Failure as a Function of Cortical-Cancellous Ratio in Anterior Cervical Discectomy/Fusion: Allograft Spacer Alone Model
    Kwon, Ji-Won
    Lee, Hwan-Mo
    Park, Tae-Hyun
    Lee, Sung Jae
    Kwon, Young-Woo
    Moon, Seong-Hwan
    Lee, Byung Ho
    APPLIED SCIENCES-BASEL, 2020, 10 (18):
  • [40] The Results of Percutaneous Intradiscal High-Pressure Injection of Saline in Patients with Extruded Lumbar Herniated Disc: Comparison with Microendoscopic Discectomy
    Fukui, Sei
    Iwashita, Narihito
    Nitta, Kazuhito
    Tomie, Hisashi
    Nosaka, Shuichi
    PAIN MEDICINE, 2012, 13 (06) : 762 - 768