In vitro biomechanical evaluation of a strutted intradiscal spacer for lumbar discectomy

被引:0
|
作者
Schmoelz, Werner [1 ]
Spicher, Anna [1 ]
Lindtner, Richard [1 ]
Hoermann, Romed [2 ]
Srour, Robin [3 ]
机构
[1] Med Univ Innsbruck, Dept Orthoped & Traumatol, Anichstr 35, A-6020 Innsbruck, Austria
[2] Med Univ Innsbruck, Dept Anat Histol & Embryol, Div Clin & Funct Anat, Mullerstr 59, A-6020 Innsbruck, Austria
[3] Hop Civils Colmar, Dept Neurosurg, Colmar, France
关键词
Lumbar disc herniation; Intradiscal spacer; Disc repair; Cadaver study; Reherniation; DISC HERNIATION; SURGERY; INSTRUMENTATION; REHERNIATION; GUIDELINE; DISEASE; HEIGHT; FUSION; COHORT; RISK;
D O I
10.1016/j.clinbiomech.2025.106491
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
Background: Discectomy plus implantation of a strutted intradiscal spacer has been shown to reduce reoperations and reherniations versus discectomy alone following lumbar disc herniation. This study explored the underlying biomechanics of the intradiscal spacer. Methods: Six fresh-frozen cadaveric lumbar spine specimens (L2 to L5) from three donors were used. Following box annulotomy to simulate disc herniation, a discectomy was performed. One segment from each donor was randomly assigned to either an untreated control group or the test group where an intradiscal spacer was implanted. A six degree of freedom universal spine tester assessed range of motion (RoM) in flexion/extension, lateral bending and axial rotation in the native state in load controlled [f7.5 Nm] and intervals up to 60,000 cycles. Disc height was measured on fluoroscopy for multiple timepoints. The segments were also analyzed to detect possible reherniation during the cycling loading. Findings: Following 60,000 cycles, the mean percentage RoM increase versus the intact state was less for discectomy plus the intradiscal spacer versus discectomy alone for lateral bending (170.7 f 10.0 vs. 222.5 f 33.3 %), flexion/extension (178.5 f 6.1 vs. 204.6 f 44.3 %) and axial rotation (284.4 f 127.2 vs. 362.3 % f 240.4 %). Mean overall disc height loss versus the annulotomy state was also less with the intradiscal spacer versus discectomy alone (-19.3 f 3.7 vs. -29.1 f 6.1 %). There was no evidence of device subsidence or migration. Interpretation: This study helps to explain the clinical observation that insertion of a strutted intradiscal spacer following discectomy reduces reherniation rate by mechanically limiting the increase in RoM and disc height loss following lumbar discectomy.
引用
收藏
页数:6
相关论文
共 50 条
  • [21] Evaluation of transforaminal endoscopic lumbar discectomy in the treatment of lumbar disc herniation
    Kun Wang
    Xin Hong
    Bao-Yi Zhou
    Jun-Ping Bao
    Xin-Hui Xie
    Feng Wang
    Xiao-Tao Wu
    International Orthopaedics, 2015, 39 : 1599 - 1604
  • [22] Biomechanical analysis of the lumbar spine on facet joint force and intradiscal pressure - a finite element study
    Ching-Sung Kuo
    Hsuan-Teh Hu
    Ruey-Mo Lin
    Kuo-Yuan Huang
    Po-Chun Lin
    Zheng-Cheng Zhong
    Mu-Lin Hseih
    BMC Musculoskeletal Disorders, 11
  • [23] Biomechanical analysis of the lumbar spine on facet joint force and intradiscal pressure - a finite element study
    Kuo, Ching-Sung
    Hu, Hsuan-Teh
    Lin, Ruey-Mo
    Huang, Kuo-Yuan
    Lin, Po-Chun
    Zhong, Zheng-Cheng
    Hseih, Mu-Lin
    BMC MUSCULOSKELETAL DISORDERS, 2010, 11
  • [24] THE INFLUENCE OF VERTEBRAL BODY FRACTURE, INTRADISCAL INJECTION AND PARTIAL DISCECTOMY ON THE RADIAL BULGE AND HEIGHT OF HUMAN LUMBAR DISKS
    BRINCKMANN, P
    HORST, M
    JOURNAL OF BIOMECHANICS, 1985, 18 (07) : 546 - 546
  • [25] THE INFLUENCE OF VERTEBRAL BODY FRACTURE, INTRADISCAL INJECTION, AND PARTIAL DISCECTOMY ON THE RADIAL BULGE AND HEIGHT OF HUMAN LUMBAR DISKS
    BRINCKMANN, P
    HORST, M
    SPINE, 1985, 10 (02) : 138 - 145
  • [26] Biomechanical Evaluation of a Posterolateral Lumbar Disc Arthroplasty Device An In Vitro Human Cadaveric Model
    Kikkawa, Jun
    Cunningham, Bryan W.
    Shirado, Osamu
    Hu, Nianbin
    McAfee, Paul C.
    Oda, Hiromi
    SPINE, 2010, 35 (19) : 1760 - 1768
  • [27] The Effects of Physiological Biomechanical Loading on Intradiscal Pressure and Annulus Stress in Lumbar Spine: A Finite Element Analysis
    Zahari, Siti Nurfaezah
    Latif, Mohd Juzaila Abd
    Rahim, Nor Raihanah Abdull
    Kadir, Mohammed Rafiq Abdul
    Kamarul, Tunku
    JOURNAL OF HEALTHCARE ENGINEERING, 2017, 2017
  • [28] Biomechanical Effect of Hybrid Dynamic Stabilization Implant on the Segmental Motion and Intradiscal Pressure in Human Lumbar Spine
    Hsiao, Chih-Kun
    Tsai, Yi-Jung
    Yen, Cheng-Yo
    Li, Yi-Chen
    Hsiao, Hao-Yuan
    Tu, Yuan-Kun
    BIOENGINEERING-BASEL, 2023, 10 (01):
  • [29] Biomechanical Finite Element Analysis of Percutaneous Endoscopic Lumbar Discectomy via a Transforaminal Approach
    Wu, Zhongxuan
    Sun, Hongzhou
    Zhang, Yu
    Xiao, Liang
    Zhao, Quanlai
    WORLD NEUROSURGERY, 2024, 185 : e291 - e298
  • [30] CHANGE OF DISK HEIGHT, RADIAL DISK BULGE, AND INTRADISCAL PRESSURE FROM DISCECTOMY - AN INVITRO INVESTIGATION ON HUMAN LUMBAR DISKS
    BRINCKMANN, P
    GROOTENBOER, H
    SPINE, 1991, 16 (06) : 641 - 646