Accuracy of freehand surgery, static and dynamic computer assisted surgery on zygomatic implant placement: A systematic review and meta-analyses

被引:1
|
作者
Traboulsi-Garet, Bassel [1 ]
Jorba-Garcia, Adria [1 ,4 ]
Bara-Casaus, Javier [1 ,2 ,3 ]
Camps-Font, Octavi [1 ,4 ]
Valmaseda-Castellon, Eduard [1 ,4 ]
Figueiredo, Rui [1 ,4 ]
Sanchez-Garces, Angels [1 ,4 ]
机构
[1] Univ Barcelona, Sch Med & Hlth Sci, Barcelona, Spain
[2] Hosp Univ Sagrat Cor, Grp Quironsalud, Dent & Maxillofacial Inst, Barcelona, Spain
[3] Univ Barcelona, Hosp Univ Mutua Terrassa, Dept Oral & Maxillofacial Surg, Terrassa, Spain
[4] IDIBELL Inst, Barcelona, Spain
关键词
Accuracy; Zygomatic implants; Arch; /; Zygomatic; Image guided surgery; Static guided surgery; TIME SURGICAL NAVIGATION; DENTAL IMPLANTS; GUIDED SURGERY; PATIENT; REHABILITATION; COMPLICATIONS; SURVIVAL;
D O I
10.1016/j.jcms.2024.12.002
中图分类号
R78 [口腔科学];
学科分类号
1003 ;
摘要
Real-time surgical navigation systems (dynamic computer-aided surgery, d-CAIS) and static guided surgery (static computer-aided surgery, s-CAIS) have been shown to enhance the accuracy of zygomatic implant (ZI) placement. The objective of this systematic review was to evaluate and compare the accuracy and risk of complications associated with d-CAIS and s-CAIS in ZI placement. A systematic review of published studies involving more than 4 patients was conducted to assess and compare the accuracy of d-CAIS and s-CAIS in zygomatic implant placement. Only one study included freehand ZI placement as a control. The primary outcomes measured were the accuracy of implant placement relative to preoperative planning, with a secondary focus on evaluating any potential complications. Out of 903 screened studies, 14 met the inclusion criteria. Freehand zygomatic implant placement was used as a control in only 1 study. The results revealed a mean apex deviation of 2.07 mm (95% CI: 2.01 to 2.13; I2 = 83.14%) for d-CAIS, 1.29 mm (95% CI: 1.15 to 1.43; I2 = 94.5%) for s-CAIS, and 4.98 mm (95% CI: 3.59 to 6.37; I2 = not assessable) for freehand placement. Reported complications included mucositis, reversible bilateral sinusitis, oroantral fistula, unspecified reversible postoperative complications, and fracture of the anterior wall of the zygoma. Both CAIS systems demonstrated high accuracy and safety in ZI placement, with a nearly 99% success rate at 6 months of follow-up. These findings suggest that both d-CAIS and s-CAIS are reliable methods for improving the precision and reducing the risks associated with ZI procedures.
引用
收藏
页码:301 / 311
页数:11
相关论文
共 50 条
  • [1] The Accuracy of Zygomatic Implant Placement Assisted by Dynamic Computer-Aided Surgery: A Systematic Review and Meta-Analysis
    Fan, Shengchi
    Saenz-Ravello, Gustavo
    Diaz, Leonardo
    Wu, Yiqun
    Davo, Ruben
    Wang, Feng
    Magic, Marko
    Al-Nawas, Bilal
    Kaemmerer, Peer W.
    JOURNAL OF CLINICAL MEDICINE, 2023, 12 (16)
  • [2] Comparison of accuracy in freehand versus computer-assisted (dynamic and static) dental implant placement: A systematic review and meta-analysis
    Banerjee, Saurav
    Debnath, Anasua
    Paul, Priyanjali
    Banerjee, Tridib Nath
    JOURNAL OF INDIAN PROSTHODONTIC SOCIETY, 2025, 25 (01): : 22 - 29
  • [3] Accuracy of freehand versus dynamic computer-assisted zygomatic implant placement: An in-vitro study
    Traboulsi-Garet, Bassel
    Jorba-Garcia, Adria
    Bara-Casaus, Javier
    Camps-Font, Octavi
    Valmaseda-Castellon, Eduard
    Figueiredo, Rui
    Sanchez-Garces, M. Angeles
    JOURNAL OF DENTISTRY, 2025, 155
  • [4] Accuracy of implant placement with computer-aided static, dynamic, and robot-assisted surgery: a systematic review and meta-analysis of clinical trials
    Khaohoen, Angkoon
    Powcharoen, Warit
    Sornsuwan, Tanapon
    Chaijareenont, Pisaisit
    Rungsiyakull, Chaiy
    Rungsiyakull, Pimduen
    BMC ORAL HEALTH, 2024, 24 (01)
  • [5] Accuracy of implant placement with computer-aided static, dynamic, and robot-assisted surgery: a systematic review and meta-analysis of clinical trials
    Angkoon Khaohoen
    Warit Powcharoen
    Tanapon Sornsuwan
    Pisaisit Chaijareenont
    Chaiy Rungsiyakull
    Pimduen Rungsiyakull
    BMC Oral Health, 24
  • [6] Accuracy Comparison between Robot-Assisted Dental Implant Placement and Static/Dynamic Computer-Assisted Implant Surgery: A Systematic Review and Meta-Analysis of In Vitro Studies
    Jain, Saurabh
    Sayed, Mohammed E.
    Ibraheem, Wael I.
    Ageeli, Abrar A.
    Gandhi, Sumir
    Jokhadar, Hossam F.
    Alresayes, Saad Saleh
    Alqarni, Hatem
    Alshehri, Abdullah Hasan
    Huthan, Halah Mohammed
    Alami, Atheer
    Al Wadei, Mohammed Hussain Dafer
    Aljabri, Yahya
    MEDICINA-LITHUANIA, 2024, 60 (01):
  • [7] Combined static and dynamic computer-guided surgery for prosthetically driven zygomatic implant placement
    Du, Chenlin
    Peng, Pei
    Guo, Xiaodong
    Wu, Yunshu
    Zhang, Zeyu
    Hao, Linke
    Zhang, Zongjiu
    Xiong, Jingyu
    JOURNAL OF DENTISTRY, 2025, 152
  • [8] Accuracy of dental implant placement using static versus dynamic computer-assisted implant surgery: An in vitro study
    Otaghsara, Seyedeh Sahar Taheri
    Joda, Tim
    Thieringer, Florian Markus
    JOURNAL OF DENTISTRY, 2023, 132
  • [9] The accuracy of static computer-aided implant surgery: A systematic review and meta-analysis
    Tahmaseb, Ali
    Wu, Viviane
    Wismeijer, Daniel
    Coucke, Wim
    Evans, Christopher
    CLINICAL ORAL IMPLANTS RESEARCH, 2018, 29 : 416 - 435
  • [10] Minimal invasiveness at dental implant placement: A systematic review with meta-analyses on flapless fully guided surgery
    Romandini, Mario
    Ruales-Carrera, Edwin
    Sadilina, Sofya
    Haemmerle, Christoph H. F.
    Sanz, Mariano
    PERIODONTOLOGY 2000, 2023, 91 (01) : 89 - 112