Multidimensional RF pulse design with consideration of concomitant field effects

被引:0
|
作者
Zhao, Ziwei [1 ]
Lee, Nam G. [2 ]
Nayak, Krishna S. [1 ,2 ]
机构
[1] Univ Southern Calif, Viterbi Sch Engn, Ming Hsieh Dept Elect & Comp Engn, 3740 McClintock Ave,EEB 400, Los Angeles, CA 90089 USA
[2] Univ Southern Calif, Viterbi Sch Engn, Alfred Mann Dept Biomed Engn, Los Angeles, CA USA
基金
美国国家科学基金会;
关键词
concomitant field effects; gradient impulse response function (GIRF); low-field; multi-dimensional RF pulse design; RATE SELECTIVE EXCITATION; QUANTITATIVE CHARACTERIZATION; MAGNETIC-RESONANCE; GRADIENT; MR; INHOMOGENEITY; OPTIMIZATION; ARTIFACTS; SAR;
D O I
暂无
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Purpose: To develop a small-tip multidimensional RF pulse design procedure that incorporates linear time-invariant gradient imperfections and concomitant field effects. This could be particularly important for contemporary low-field MRI systems with high-performance gradients. Theory and Methods: We developed an extension of the small-tip excitation k-space formalism, where concomitant fields were approximated as a Bloch-Siegert shift in the rotating frame. This was evaluated using realistic simulations of 2D selective excitation at various field strengths (0.2T, 0.55T, 1.5T, 3T, and 7T) with single and parallel transmit. Simulated excitation profiles from the original and extended k-space formalisms were compared. Experimental validations were performed at 0.55T with a single-channel transmit. Results: The extended formalism provides improved 2D excitation profiles in all scenarios simulated, compared against the original formalism. The proposed method corrects the concomitant field effects on 2D selective excitations for B-0 > 0.2T when the magnitude of the B-0 is far larger than that of nonrotating concomitant fields. Simulation and phantom experiments at 0.55T match well for both original and proposed methods, with the proposed method providing sharper and more accurate excitation profiles at off-isocenter distances up to 15cm. The impact of the proposed method is greatest in scenarios where concomitant fields are substantial, such as low field strengths and off-isocenter. Conclusion: Concomitant fields can be modeled as a Bloch-Siegert shift in the rotating frame during multidimensional RF pulse design, resulting in improved excitation profiles with sharp edges. This is important to consider for off-isocenter excitations and imaging at low field strengths with strong gradients.
引用
收藏
页码:718 / 729
页数:12
相关论文
共 50 条
  • [41] Possible effects of RF field near ICRF antenna on density control during long pulse discharge in LHD
    Saito, K
    Kumazawa, R
    Mutoh, T
    Seki, T
    Watari, T
    Nakamura, Y
    Sakamoto, M
    Noda, N
    Watanabe, T
    Shoji, M
    Masuzaki, S
    Morita, S
    Goto, M
    Torii, Y
    Takeuchi, N
    Shimpo, F
    Nomura, G
    Yokota, M
    Kato, A
    Zhao, Y
    JOURNAL OF NUCLEAR MATERIALS, 2005, 337 (1-3) : 995 - 999
  • [42] Magnetization reversal of a single cobalt cluster using a RF field pulse
    Tamion, A.
    Raufast, C.
    Bonet, E.
    Dupuis, V.
    Fournier, T.
    Crozes, T.
    Bernstein, E.
    Wernsdorfer, W.
    JOURNAL OF MAGNETISM AND MAGNETIC MATERIALS, 2010, 322 (9-12) : 1315 - 1318
  • [43] Deconvolution of medical ultrasound data with consideration of the pressure field, the excitation pulse and focussing
    Hewener, Holger J.
    Lemor, Robert
    ADVANCES IN MEDICAL ENGINEERING, 2007, 114 : 157 - +
  • [44] RF design of an S-band spherical cavity pulse compressor
    Lei J.
    He X.
    Hou M.
    Shu G.
    Pei G.-X.
    Li X.-P.
    Wang H.
    Radiation Detection Technology and Methods, 2017, 1 (2)
  • [45] Novel open cavity design for rotating mode rf pulse compressors
    Wu, Xiaowei
    Grudiev, Alexej
    PHYSICAL REVIEW ACCELERATORS AND BEAMS, 2021, 24 (11)
  • [46] Magnetic Resonance RF Pulse Design by Optimal Control With Physical Constraints
    Rund, Armin
    Aigner, Christoph Stefan
    Kunisch, Karl
    Stollberger, Rudolf
    IEEE TRANSACTIONS ON MEDICAL IMAGING, 2018, 37 (02) : 461 - 472
  • [47] VERSE-guided numerical RF pulse design: A fast method for peak RF power control
    Lee, Daeho
    Grissom, William A.
    Lustig, Michael
    Kerr, Adam B.
    Stang, Pascal P.
    Pauly, John M.
    MAGNETIC RESONANCE IN MEDICINE, 2012, 67 (02) : 353 - 362
  • [48] Transabdominal Fetal Pulse Oximeter Using LEDs and Photodiode: A Design Consideration Study
    Jumadi, Nur Anida
    Beng, Gan Kok
    Ali, Mohd Alauddin Mohd
    Zahedi, Edmond
    PROCEEDINGS 2015 2ND INTERNATIONAL CONFERENCE ON BIOMEDICAL ENGINEERING (ICOBE 2015), 2015,
  • [49] Design Consideration on Converged Rx SAW Duplexer Module for Multiband RF Front End
    Iwaki, Masafumi
    Tanaka, Tabito
    Ueda, Masanori
    Satoh, Yoshio
    IEEE TRANSACTIONS ON MICROWAVE THEORY AND TECHNIQUES, 2017, 65 (11) : 4629 - 4635
  • [50] Work function consideration in vacuum field emission transistor design
    Kim, Jiwon
    Oh, Hyeongwan
    Kim, Jungsik
    Baek, Rock-Hyun
    Han, Jin-Woo
    Meyyappan, M.
    Lee, Jeong-Soo
    JOURNAL OF VACUUM SCIENCE & TECHNOLOGY B, 2017, 35 (06):