Improving image reconstruction to quantify dynamic whole-body PET/CT: Q.Clear versus OSEM

被引:0
|
作者
Springer, Sam [1 ]
Basset-Sagarminaga, Jeremy [2 ,4 ]
van de Weijer, Tineke [1 ,2 ,4 ]
Schrauwen-Hinderling, Vera B. [1 ,4 ]
Backes, Walter H. [1 ,3 ]
Wierts, Roel [1 ]
机构
[1] Maastricht Univ Med Ctr, Dept Radiol & Nucl Med, Maastricht, Netherlands
[2] Maastricht Univ, Dept Nutr & Movement Sci, Maastricht, Netherlands
[3] Maastricht Univ, Res Inst Mental Hlth & Neurosci & Cardiovasc Dis, Maastricht, Netherlands
[4] Maastricht Univ, Inst Nutr & Translat Res Metab, Maastricht, Netherlands
来源
EJNMMI PHYSICS | 2025年 / 12卷 / 01期
关键词
PET/CT; F-18-FDG; Reconstruction; Bayesian; Dynamic; Whole-body;
D O I
10.1186/s40658-025-00736-5
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Background: The introduction of PET systems featuring increased count rate sensitivity has resulted in the development of dynamic whole-body PET acquisition protocols to assess F-18-FDG uptake rate (Ki) using F-18-FDG PET/CT. However, in short-axis field-of-view (SAFOV) PET/CT systems, multiple bed positions are required per time frame to achieve whole-body coverage. This results in high noise levels, requiring higher F-18-FDG activity administration and, consequently, increased patient radiation dose. Bayesian penalized-likelihood PET reconstruction (e.g. Q.Clear, GE Healthcare) has been shown to effectively suppress image noise compared to standard reconstruction techniques. This study investigated the impact of Bayesian penalized-likelihood reconstruction on dynamic whole-body F-18-FDG PET quantification. Methods: Dynamic whole-body F-18-FDG PET/CT data (SAFOV PET Discovery MI 5R, GE Healthcare) of healthy volunteers and one lung cancer patient, consisting of a ten-minute dynamic scan of the thoracic region followed by six whole-body passes, were reconstructed with Q.Clear and Ordered Subset Expectation Maximization (OSEM) according to EARL 2 standards. Image noise in the measured time-activity-curves (TAC) was determined for the myocardium, hamstring, liver, subcutaneous adipose tissue and lung lesion for both reconstruction methods. Ki values were calculated using Patlak analysis. Finally, bootstrapping was used to investigate the effect of image noise levels on Ki values (bias and precision) as a function of magnitude of Ki and volume-of-interest (VOI) size for both computationally simulated TACs (Ki = 1.0-50.0<middle dot>10(-3)<middle dot>ml<middle dot>cm(-3)<middle dot>min(-1)) and the measured TACs. Results: Compared to OSEM, Q.Clear showed 40-55% lower noise levels for all tissue types (p < 0.05). For the measured TACs no systematic bias in Ki with either reconstruction method was observed. Ki precision decreased with decreasing VOI size, with that of Q.Clear being superior compared to OSEM for small VOIs of 0.56 cm(3) in all tissues (p < 0.05), with the largest difference in relative precision for small values of Ki. The simulated TACs corroborated these results, with Q.Clear providing the best precision for small values of Ki and small VOIs in all tissues. Conclusion: Q.Clear reconstruction of dynamic whole-body PET/CT data yields more precise Ki values, especially for small values of Ki and smaller VOIs, compared to standard OSEM. This precision improvement shows Q.Clear's potential to better detect and characterize small lesion metabolic activity in oncology and allows for lower administered activity dosage.
引用
收藏
页数:20
相关论文
共 50 条
  • [21] Phantom and Clinical Evaluation of the Bayesian Penalized Likelihood Reconstruction Algorithm Q.Clear on an LYSO PET/CT System
    Teoh, Eugene J.
    McGowan, Daniel R.
    Macpherson, Ruth E.
    Bradley, Kevin M.
    Gleeson, Fergus V.
    JOURNAL OF NUCLEAR MEDICINE, 2015, 56 (09) : 1447 - 1452
  • [22] Performance evaluation of the Q.Clear reconstruction framework versus conventional reconstruction algorithms for quantitative brain PET-MR studies
    Ribeiro, Daniela
    Hallett, William
    Tavares, Adriana A. S.
    EJNMMI PHYSICS, 2021, 8 (01)
  • [23] Can Q.Clear reconstruction be used to improve [68 Ga]Ga-DOTANOC PET/CT image quality in overweight NEN patients?
    Zanoni, Lucia
    Argalia, Giulia
    Fortunati, Emilia
    Malizia, Claudio
    Allegri, Vincenzo
    Calabro, Diletta
    Civollani, Simona
    Campana, Davide
    Fanti, Stefano
    Ambrosini, Valentina
    EUROPEAN JOURNAL OF NUCLEAR MEDICINE AND MOLECULAR IMAGING, 2022, 49 (05) : 1607 - 1612
  • [24] Can Q.Clear reconstruction be used to improve [68 Ga]Ga-DOTANOC PET/CT image quality in overweight NEN patients?
    Lucia Zanoni
    Giulia Argalia
    Emilia Fortunati
    Claudio Malizia
    Vincenzo Allegri
    Diletta Calabrò
    Simona Civollani
    Davide Campana
    Stefano Fanti
    Valentina Ambrosini
    European Journal of Nuclear Medicine and Molecular Imaging, 2022, 49 : 1607 - 1612
  • [25] β1600 Q.Clear Digital Reconstruction of [68Ga]Ga-DOTANOC PET/CT Improves Image Quality in NET Patients
    Di Franco, Martina
    Fortunati, Emilia
    Zanoni, Lucia
    Bonazzi, Norma
    Mosconi, Cristina
    Malizia, Claudio
    Civollani, Simona
    Campana, Davide
    Andrini, Elisa
    Lamberti, Giuseppe
    Allegri, Vincenzo
    Fanti, Stefano
    Ambrosini, Valentina
    JOURNAL OF CLINICAL MEDICINE, 2024, 13 (13)
  • [26] Phantom, clinical, and texture indices evaluation and optimization of a penalized-likelihood image reconstruction method (Q.Clear) on a BGO PET/CT scanner
    Reynes-Llompart, Gabriel
    Gamez-Cenzano, Cristina
    Luis Vercher-Conejero, Jose
    Sabate-Llobera, Aida
    Calvo-Malvar, Nahum
    Marti-Climent, Josep M.
    MEDICAL PHYSICS, 2018, 45 (07) : 3214 - 3222
  • [27] Performance evaluation of the Q.Clear reconstruction framework versus conventional reconstruction algorithms for quantitative brain PET-MR studies
    Daniela Ribeiro
    William Hallett
    Adriana A. S. Tavares
    EJNMMI Physics, 8
  • [28] Whole-body dynamic imaging with continuous bed motion PET/CT
    Osborne, Dustin R.
    Acuff, Shelley
    NUCLEAR MEDICINE COMMUNICATIONS, 2016, 37 (04) : 428 - 431
  • [29] Dynamic whole-body 68Ga-DOTATOC PET/CT
    Lodge, Martin
    Rahmim, Arman
    Antoniou, Alexander
    Solnes, Lilja
    Wahl, Richard
    JOURNAL OF NUCLEAR MEDICINE, 2015, 56 (03)
  • [30] Image Reconstruction for the New Simultaneous Whole-body OpenPET/CT Geometry
    Yin, Yannan
    Tashima, Hideaki
    Obi, Takashi
    Yamaya, Taiga
    MEDICAL IMAGING 2014: PHYSICS OF MEDICAL IMAGING, 2014, 9033