A novel personalized homogenous finite element model to predict the pull-out strength of cancellous bone screws

被引:0
|
作者
Rouyin, Alireza [1 ]
Einafshar, Mohammadjavad [2 ]
Arjmand, Navid [1 ]
机构
[1] Sharif Univ Technol, Dept Mech Engn, Tehran 111559567, Iran
[2] Aalborg Univ, Dept Mat & Prod, Aalborg, Denmark
来源
关键词
Orthopedic screw; Bone screw pull-out strength; Personalized finite element analysis; Heterogeneous material mapping; Homogenous finite element; Material assignment; MISPLACED PEDICLE SCREWS; TRABECULAR BONE; DENSITY; BEHAVIOR; DESIGN; SPINE;
D O I
10.1186/s13018-024-05169-x
中图分类号
R826.8 [整形外科学]; R782.2 [口腔颌面部整形外科学]; R726.2 [小儿整形外科学]; R62 [整形外科学(修复外科学)];
学科分类号
摘要
Background Orthopedic surgeries often involve the insertion of bone screws for various fixation systems. The risk of postoperative screw loosening is usually assessed through experimental or finite element (FE) evaluations of the screw pull-out strength. FE simulations are based on either personalized complex but accurate heterogeneous modeling or non-personalized simple but relatively less accurate homogeneous modeling. This study aimed to develop and validate a novel personalized computed tomography (CT)-based homogeneous FE simulation approach to predict the pull-out force of cancellous bone screws. Methods Twenty FE simulations of L1-L5 vertebral screw pull-out tests were conducted, i.e., 10 heterogeneous and 10 homogenous models. Screws were inserted into the lower-middle region of vertebrae. In our novel homogeneous model, the region around approximately twice the diameter of the screw was used as a bone material reference volume. Subsequently, the overall material property of this region was homogeneously attributed to the entire vertebra, and pull-out simulations were conducted. Results The mean error of the predicted pull-out forces by our novel homogenous simulations was similar to 7.9% with respect to our heterogeneous model. When solely the cancellous bone was involved during the pull-out process (i.e., for L1, L2, and L3 vertebral bodies whose cortical bone in the inferior region is thin), the novel homogenous model yielded small mean error of < 6.0%. This error, however, increased to similar to 11% when the screw got involved to the cortical bone (for L4 and L5 vertebrae whose cortical bone in the inferior region is thick). Conclusion The proposed personalized CT-based homogenous model was highly accurate in estimating the pullout force especially when only the cancellous bone was involved with the screw.
引用
收藏
页数:12
相关论文
共 50 条
  • [31] Evaluation of uniaxial pull-out strength of 1.5mm monocortical screws.
    Rabelo, LRS
    Passeri, LA
    JOURNAL OF DENTAL RESEARCH, 2000, 79 (05) : 1139 - 1139
  • [32] THE INFLUENCE OF THE "PENETRATION-" AND THE "FILLING-RATIOS" ON THE PULL-OUT STRENGTH OF TRANSPEDICULAR SCREWS
    Kourkoulis, S. K.
    Chatzistergos, P.
    JOURNAL OF MECHANICS IN MEDICINE AND BIOLOGY, 2009, 9 (03) : 283 - 300
  • [33] FIXATION PULL-OUT RESPONSE MEASUREMENT OF BONE SCREWS AND POROUS-SURFACED POSTS
    SHIRAZIADL, A
    DAMMAK, M
    ZUKOR, DJ
    JOURNAL OF BIOMECHANICS, 1994, 27 (10) : 1249 - &
  • [34] Insertional torque and pull-out strengths of conical and cylindrical pedicle screws in cadaveric bone
    Kwok, AWL
    Finkelstein, JA
    Woodside, T
    Hearn, TC
    Hu, RW
    SPINE, 1996, 21 (21) : 2429 - 2434
  • [36] Insertional torque and pull-out strengths of conical and cylindrical pedicle screws in cadaveric bone
    Kwok, AWL
    Finkelstein, JA
    Woodside, T
    Hearn, TC
    Hu, RW
    SPINE, 2000, 25 (06) : 19S - 24S
  • [37] Enhanced finite element analysis of shear connector pull-out based on cohesive zone model
    Lin, Jianping
    Wang, Jinfeng
    Xu, Rongqiao
    Ling, Daosheng
    Harbin Gongye Daxue Xuebao/Journal of Harbin Institute of Technology, 2014, 46 (12): : 78 - 83
  • [38] Effects of bone materials on the screw pull-out strength in human spine
    Zhang, Qing Hang
    Tan, Soon Huat
    Chou, Siaw Meng
    MEDICAL ENGINEERING & PHYSICS, 2006, 28 (08) : 795 - 801
  • [39] Homogenized finite element models can accurately predict screw pull-out in continuum materials, but not in porous materials
    Einafshar, Mohammadjavad
    Hashemi, Ata
    van Lenthe, G. Harry
    COMPUTER METHODS AND PROGRAMS IN BIOMEDICINE, 2021, 202
  • [40] The Effect of Hubbing on the Pull-out Strength of Lateral Mass Screws in the Cervical Spine A Biomechanical Experiment
    Koo, Ki Hyoung
    Yoon, S. Tim
    Kim, Sang Bum
    Hutton, William C.
    JOURNAL OF SPINAL DISORDERS & TECHNIQUES, 2015, 28 (01): : E45 - E48