A novel personalized homogenous finite element model to predict the pull-out strength of cancellous bone screws

被引:0
|
作者
Rouyin, Alireza [1 ]
Einafshar, Mohammadjavad [2 ]
Arjmand, Navid [1 ]
机构
[1] Sharif Univ Technol, Dept Mech Engn, Tehran 111559567, Iran
[2] Aalborg Univ, Dept Mat & Prod, Aalborg, Denmark
来源
关键词
Orthopedic screw; Bone screw pull-out strength; Personalized finite element analysis; Heterogeneous material mapping; Homogenous finite element; Material assignment; MISPLACED PEDICLE SCREWS; TRABECULAR BONE; DENSITY; BEHAVIOR; DESIGN; SPINE;
D O I
10.1186/s13018-024-05169-x
中图分类号
R826.8 [整形外科学]; R782.2 [口腔颌面部整形外科学]; R726.2 [小儿整形外科学]; R62 [整形外科学(修复外科学)];
学科分类号
摘要
Background Orthopedic surgeries often involve the insertion of bone screws for various fixation systems. The risk of postoperative screw loosening is usually assessed through experimental or finite element (FE) evaluations of the screw pull-out strength. FE simulations are based on either personalized complex but accurate heterogeneous modeling or non-personalized simple but relatively less accurate homogeneous modeling. This study aimed to develop and validate a novel personalized computed tomography (CT)-based homogeneous FE simulation approach to predict the pull-out force of cancellous bone screws. Methods Twenty FE simulations of L1-L5 vertebral screw pull-out tests were conducted, i.e., 10 heterogeneous and 10 homogenous models. Screws were inserted into the lower-middle region of vertebrae. In our novel homogeneous model, the region around approximately twice the diameter of the screw was used as a bone material reference volume. Subsequently, the overall material property of this region was homogeneously attributed to the entire vertebra, and pull-out simulations were conducted. Results The mean error of the predicted pull-out forces by our novel homogenous simulations was similar to 7.9% with respect to our heterogeneous model. When solely the cancellous bone was involved during the pull-out process (i.e., for L1, L2, and L3 vertebral bodies whose cortical bone in the inferior region is thin), the novel homogenous model yielded small mean error of < 6.0%. This error, however, increased to similar to 11% when the screw got involved to the cortical bone (for L4 and L5 vertebrae whose cortical bone in the inferior region is thick). Conclusion The proposed personalized CT-based homogenous model was highly accurate in estimating the pullout force especially when only the cancellous bone was involved with the screw.
引用
收藏
页数:12
相关论文
共 50 条
  • [1] PULL-OUT STRENGTH OF CASPAR CERVICAL SCREWS
    MAIMAN, DJ
    PINTAR, FA
    YOGANANDAN, N
    REINARTZ, J
    TOSELLI, R
    WOODWARD, E
    HAID, R
    NEUROSURGERY, 1992, 31 (06) : 1097 - 1101
  • [2] Effect of pedicle screw misplacement on the pull-out strength using personalized finite element modeling
    Rouyin, A.
    Nazemi, H.
    Arjmand, N.
    Einafshar, M.J.
    Computers in Biology and Medicine, 2024, 183
  • [3] Proximal half angle of the screw thread is a critical design variable affecting the pull-out strength of cancellous bone screws
    Wang, Yingxu
    Mori, Ryuji
    Ozoe, Nobuaki
    Nakai, Takahisa
    Uchio, Yuji
    CLINICAL BIOMECHANICS, 2009, 24 (09) : 781 - 785
  • [4] A Computational Approximation to Model Variation in Cancellous Bone Screw Pull-Out
    Piper, A.
    Brown, C. J.
    JOURNAL OF MEDICAL DEVICES-TRANSACTIONS OF THE ASME, 2016, 10 (02):
  • [5] PULL-OUT STRENGTH OF CASPAR CERVICAL SCREWS - REPLY
    MAIMAN, DJ
    NEUROSURGERY, 1993, 33 (01) : 172 - 172
  • [6] Pull-out strength of screws in long bones at different insertion angles: finite element analysis and experimental investigations
    Iyibilgin, Osman
    Gepek, Engin
    Bayam, Levent
    Drampalos, Efstathios
    Shoaib, Amer
    MATERIALS TESTING, 2024, 66 (03) : 380 - 388
  • [7] PULL-OUT STRENGTH OF SCREWS FROM POLYMETHYLMETHACRYLATE CEMENT
    MOTZKIN, NE
    CHAO, EYS
    AN, KN
    WIKENHEISER, MA
    LEWALLEN, DG
    JOURNAL OF BONE AND JOINT SURGERY-BRITISH VOLUME, 1994, 76B (02): : 320 - 323
  • [8] Influence of novel design alteration of pedicle screw on pull-out strength: A finite element study
    Takenaka, Shota
    Kaito, Takashi
    Ishii, Ken
    Watanabe, Kota
    Watanabe, Kei
    Shinohara, Akira
    Harada, Tomohisa
    Nakada, Fumiya
    Majima, Yusuke
    Matsumoto, Morio
    JOURNAL OF ORTHOPAEDIC SCIENCE, 2020, 25 (01) : 66 - 72
  • [9] Pull-out strength of screws from cortical bone in the maxillo-facial region
    Shelton, JC
    Loukota, RA
    JOURNAL OF MATERIALS SCIENCE-MATERIALS IN MEDICINE, 1996, 7 (04) : 231 - 235
  • [10] Clinical relevance of pull-out strength testing of pedicle screws
    Dawson, JM
    Boschert, P
    Macenski, MM
    Rand, N
    SPINAL IMPLANTS: ARE WE EVALUATING THEM APPROPRIATELY, 2003, 1431 : 68 - 77