Performance of a point-of-care ultrasound platform for artificial intelligence-enabled assessment of pulmonary B-lines

被引:0
|
作者
Labaf, Ashkan [1 ]
ahman-Persson, Linda [2 ]
Husu, Leo Silven [2 ]
Smith, J. Gustav [1 ,3 ]
Ingvarsson, Annika [1 ]
Evaldsson, Anna Werther [1 ]
机构
[1] Lund Univ, Skane Univ Hosp, Dept Clin Sci Lund, Sect Heart Failure & Valvular Dis,Cardiol, Klinikgatan 15, S-22185 Lund, Sweden
[2] Skane Univ Hosp, Dept Internal & Emergency Med, Malmo, Sweden
[3] Univ Gothenburg, Inst Med, Dept Mol & Clin Med, Gothenburg, Sweden
关键词
Lung ultrasound; B-lines; Artificial intelligence; POCUS; LUNG ULTRASOUND; HEART-FAILURE; AGREEMENT;
D O I
10.1186/s12947-025-00338-2
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
Background The incorporation of artificial intelligence (AI) into point-of-care ultrasound (POCUS) platforms has rapidly increased. The number of B-lines present on lung ultrasound (LUS) serve as a useful tool for the assessment of pulmonary congestion. Interpretation, however, requires experience and therefore AI automation has been pursued. This study aimed to test the agreement between the AI software embedded in a major vendor POCUS system and visual expert assessment. Methods This single-center prospective study included 55 patients hospitalized for various respiratory symptoms, predominantly acutely decompensated heart failure. A 12-zone protocol was used. Two experts in LUS independently categorized B-lines into 0, 1-2, 3-4, and >= 5. The intraclass correlation coefficient (ICC) was used to determine agreement. Results A total of 672 LUS zones were obtained, with 584 (87%) eligible for analysis. Compared with expert reviewers, the AI significantly overcounted number of B-lines per patient (23.5 vs. 2.8, p < 0.001). A greater proportion of zones with > 5 B-lines was found by the AI than by the reviewers (38% vs. 4%, p < 0.001). The ICC between the AI and reviewers was 0.28 for the total sum of B-lines and 0.37 for the zone-by-zone method. The interreviewer agreement was excellent, with ICCs of 0.92 and 0.91, respectively. Conclusion This study demonstrated excellent interrater reliability of B-line counts from experts but poor agreement with the AI software embedded in a major vendor system, primarily due to overcounting. Our findings indicate that further development is needed to increase the accuracy of AI tools in LUS.
引用
收藏
页数:8
相关论文
共 50 条
  • [21] Overview of artificial intelligence in point-of-care ultrasound. New horizons for respiratory system diagnoses
    Mika, Slawomir
    Gola, Wojciech
    Gil-Mika, Monika
    Wilk, Mateusz
    Misiolek, Hanna
    ANAESTHESIOLOGY INTENSIVE THERAPY, 2024, 56 (01) : 1 - 8
  • [22] POINT-OF-CARE ULTRASOUND-A NEW OPTION FOR EARLY QUANTITATIVE ASSESSMENT OF PULMONARY EDEMA
    Zhu, Zhixing
    Lian, Xihua
    Zeng, Yiming
    Wu, Weijing
    Xu, Zhirong
    Chen, Yongjian
    Li, Jingyun
    Su, Xiaoshan
    Zeng, Liqing
    Lv, Guorong
    ULTRASOUND IN MEDICINE AND BIOLOGY, 2020, 46 (01): : 1 - 10
  • [23] Artificial Intelligence Assistance in Point-of-Care Ultrasound Skill Retention for Novice Users in Space Medicine Scenarios
    Hurd, Victoria S.
    Valle, Michael Del
    Kravets, Victoria G.
    Anderson, Arian L.
    Le, Nhu-Nguyen
    Kendall, John L.
    Hayman, Allison P.
    Riscinti, Matthew
    WILDERNESS & ENVIRONMENTAL MEDICINE, 2025,
  • [24] Diagnostic Accuracy of Point-of-Care Ultrasound Performed by Pulmonary Critical Care Physicians for the Assessment of the Right Ventricle
    Filopei, Jason
    Acquah, Samuel
    Bondarsky, Eric
    Ramesh, Navitha
    Patrawalla, Paru
    CHEST, 2016, 150 (04) : 657A - 657A
  • [25] Assessment of extravascular lung water by measuring the number of pulmonary ultrasound B-lines before and after CBP in patients with MODS
    Cao, Guangke
    Wu, Yu
    Zhao, Yuliang
    Wang, Ling
    Zhang, Yang
    MEDICINE, 2021, 100 (01) : E24181
  • [26] Point-of-care lung ultrasound optimization scheme for the assessment of cardiogenic pulmonary edema: A promising tool
    Zhu, Li
    Zhong, Huohu
    Liu, Piaopiao
    Xu, Zhenhong
    JOURNAL OF RADIATION RESEARCH AND APPLIED SCIENCES, 2024, 17 (02)
  • [27] From Seeing to Knowing with Artificial Intelligence: A Scoping Review of Point-of-Care Ultrasound in Low-Resource Settings
    Venkatayogi, Nethra
    Gupta, Maanas
    Gupta, Alaukik
    Nallaparaju, Shreya
    Cheemalamarri, Nithya
    Gilari, Krithika
    Pathak, Shireen
    Vishwanath, Krithik
    Soney, Carel
    Bhattacharya, Tanisha
    Maleki, Nirvana
    Purkayastha, Saptarshi
    Gichoya, Judy Wawira
    APPLIED SCIENCES-BASEL, 2023, 13 (14):
  • [28] USE OF ARTIFICIAL INTELLIGENCE BY MEDICAL STUDENTS TO ENABLE ACCURATE POINT-OF-CARE ECHOCARDIOGRAPHIC ASSESSMENT OF EJECTION FRACTION
    Dadon, Ziv
    Rosenmann, David
    Butnaru, Adi
    Glikson, Michael
    Alper-Suissa, Liat
    Alpert, Evan Avraham
    JOURNAL OF THE AMERICAN COLLEGE OF CARDIOLOGY, 2020, 75 (11) : 1568 - 1568
  • [29] Feasibility of machine integrated point of care lung ultrasound automatic B-lines tool in the Corona-virus 2019 critical care unit
    Gal Tsaban
    Ori Galante
    Yaniv Almog
    Yuval Ullman
    Lior Fuchs
    Critical Care, 25
  • [30] Feasibility of machine integrated point of care lung ultrasound automatic B-lines tool in the Corona-virus 2019 critical care unit
    Tsaban, Gal
    Galante, Ori
    Almog, Yaniv
    Ullman, Yuval
    Fuchs, Lior
    CRITICAL CARE, 2021, 25 (01)